Skip to content

ROSA One 3.1.0.5Recalls & reported problems

Medtech · Standard ROSA® ONE Surgical Assistant device( including robot arm and stand , touch screen, support arm) 120 V - 60 Hz - Plug B · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03760244030948), received between May 5, 2023 and Nov 11, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

1 FDA recall record name this device's identifier.

2 other recalls cover devices cleared under the same FDA submission (K182417); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 6, 2020Terminated

    Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.

    Z-1151-2020 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect Measurement1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2024Incorrect Measurement; Incorrect Measurement

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20658051 ↗
  • MalfunctionMay 5, 2023Output Problem; Output Problem

    IT WAS REPORTED THAT DURING SURGERY WHILE USING A ROSA, REGISTRATION AND VERIFICATION WERE SUCCESSFUL. AFTER THE FIRST DBS LEAD WAS PLACED, AN INTRAOPERATIVE IMAGE WAS TAKEN VIA O-ARM. AFTER MERGING THE IMAGES, THE LEAD WAS NOTED TO BE OVER 3MM OFF FROM THE TARGET. SURGEON REQUESTED TO CREATE A TRAJECTORY TO ONE OF THE BONE FIDUCIALS USED FOR REGISTRATION AFTER BEING CONCERNED THE HEAD HAD MOVED. WHEN DRIVING TO THE FIDUCIAL TRAJECTORY, IT WAS NOTICED THE TRAJECTORY WAS OFF FROM THE FIDUCIAL. SURGEON UNDRAPED AND REREGISTERED. REGISTRATION AND VERIFICATION WERE SUCCESSFUL. SURGEON USED TWO FIDUCIAL TRAJECTORIES TO CHECK THAT WE WERE ACCURATE. AFTER DRILLING THE SECOND BURR HOLE, THE SURGEON DROVE BACK TO ONE OF THE FIDUCIAL TRAJECTORIES TO CHECK WE HAD NOT MOVED. THE TRAJECTORY WAS OFF AGAIN. SURGEON HAD TO UNDRAPE AND REREGISTER A THIRD TIME. THE SECOND LEAD WAS PLACED AND INCISIONS WERE CLOSED. IT IS SUSPECTED THE RADIOLUCENT HEADFRAME ALLOWED THE HEAD TO MOVE. OVER 90 MINUTES WAS ADDED TO THE SURGERY DUE TO THE REREGISTRATION TIME. NO ADDITIONAL PATIENT IMPACT WAS REPORTED. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16877045 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K182417. One submission often covers many models, so check the product and lot details in each record.

  • Z-0118-2022 · Sep 22, 2021 · Open, Classified
    ROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained surgeons to guide s
  • Z-1150-2020 · Jan 6, 2020 · Terminated
    ROSA Brain 3.0 Application-Brain

Other Neurological Stereotaxic Instrument devices

All Neurological Stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page