ROSA One 3.1.0.5Recalls & reported problems
Medtech · Standard ROSA® ONE Surgical Assistant device( including robot arm and stand , touch screen, support arm) 120 V - 60 Hz - Plug B · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03760244030948), received between May 5, 2023 and Nov 11, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Incorrect Measurement” (1 report).
1 FDA recall record name this device's identifier.
2 other recalls cover devices cleared under the same FDA submission (K182417); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jan 6, 2020Terminated
Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
Z-1151-2020 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 11, 2024Incorrect Measurement; Incorrect Measurement
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20658051 ↗MalfunctionMay 5, 2023Output Problem; Output Problem
IT WAS REPORTED THAT DURING SURGERY WHILE USING A ROSA, REGISTRATION AND VERIFICATION WERE SUCCESSFUL. AFTER THE FIRST DBS LEAD WAS PLACED, AN INTRAOPERATIVE IMAGE WAS TAKEN VIA O-ARM. AFTER MERGING THE IMAGES, THE LEAD WAS NOTED TO BE OVER 3MM OFF FROM THE TARGET. SURGEON REQUESTED TO CREATE A TRAJECTORY TO ONE OF THE BONE FIDUCIALS USED FOR REGISTRATION AFTER BEING CONCERNED THE HEAD HAD MOVED. WHEN DRIVING TO THE FIDUCIAL TRAJECTORY, IT WAS NOTICED THE TRAJECTORY WAS OFF FROM THE FIDUCIAL. SURGEON UNDRAPED AND REREGISTERED. REGISTRATION AND VERIFICATION WERE SUCCESSFUL. SURGEON USED TWO FIDUCIAL TRAJECTORIES TO CHECK THAT WE WERE ACCURATE. AFTER DRILLING THE SECOND BURR HOLE, THE SURGEON DROVE BACK TO ONE OF THE FIDUCIAL TRAJECTORIES TO CHECK WE HAD NOT MOVED. THE TRAJECTORY WAS OFF AGAIN. SURGEON HAD TO UNDRAPE AND REREGISTER A THIRD TIME. THE SECOND LEAD WAS PLACED AND INCISIONS WERE CLOSED. IT IS SUSPECTED THE RADIOLUCENT HEADFRAME ALLOWED THE HEAD TO MOVE. OVER 90 MINUTES WAS ADDED TO THE SURGERY DUE TO THE REREGISTRATION TIME. NO ADDITIONAL PATIENT IMPACT WAS REPORTED. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16877045 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K182417. One submission often covers many models, so check the product and lot details in each record.
- Z-0118-2022 · Sep 22, 2021 · Open, ClassifiedROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained surgeons to guide s
- Z-1150-2020 · Jan 6, 2020 · TerminatedROSA Brain 3.0 Application-Brain
Other Neurological Stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record