ROBOT ROSA ONE (USA) recallZ-1151-2020 · Medtech Sas · initiated Jan 6, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtech Sas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
FDA root cause category: Software design
What the company is doing
On January 6, 2020, Zimmer Biomet issued URGENT MEDICAL DEVICE CORRECTION notices to customers via FedEx. Actions Required By Users: Your device can continue to be used, as long as the recommendations below are followed: While the current Instructions For Use do not require the images to be reviewed with ROSA Brain software prior to surgery, they adequately describe the standard workflow and indicate that the image acquisition of the patients head (MRI / CT or 2D X-Ray images) is performed prior to surgery, and is then loaded into the device and used by the surgeon in the preoperative phase to carry out the surgical planning. Due to this potential image management issue, Zimmer Biomet recommends that images be reviewed using the ROSA Brain software prior to surgery. If an issue is detected with the reconstruction and display of the images using the ROSA software, and if the artifacts are located in the Region Of Interest and prevent the surgical planning, then the surgery might have to be cancelled or converted to traditional surgery. However, if the artifacts are not located inside the Region Of Interest and if the surgery planning is successful, the affected device can continue to be used. Transmission of the Medical Device Correction: Please advise the appropriate personnel working in your department with the ROSA system of the content of the letter. Risk Manager Responsibilities: 1. Review the notification and ensure affected personnel are aware of the contents. 2. Complete the Certificate of Acknowledgement. a. Return a digital copy to [email protected]. b. Retain a copy of the Acknowledgement Form with your field action records in the event of a compliance audit of your facilities documentation. 3. If you have further questions or concerns after reviewing the notice, please call Perry Twyford at 281-389-3236 between 9:00 am and 6:00pm EST, Monday through Friday. Alternatively, your questions may be e
Recall details
- Recall number
- Z-1151-2020
- Status
- Terminated
- Initiated by the company
- Jan 6, 2020
- Recalling firm
- Medtech Sas
- Quantity
- 39 units
- Product code
- HAW
- FDA submissions
- K151359, K172444, K182417
- Distribution
- US Nationwide
Product description
ROSA One 3.1 Brain Application
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Serial Number / UDI Number: 1) BS19037 / UDI - (01)03760244030948(11)190423(21)BS19037; 2) BS19038 / UDI - (01)03760244030948(11)190415(21)BS19038; 3) BS19039 / UDI - (01)03760244030948(11)190430(21)BS19039; 4) BS19040 / UDI - (01)03760244030948(11)190513(21)BS19040; 5) BS19041 / UDI - (01)03760244030948(11)190523(21)BS19041; 6) BS19043 / UDI - (01)03760244030948(11)190513(21)BS19043; 7) BS19049 / UDI - (01)03760244032492(11)190730(21)BS19049; 8) BS19050 / UDI - (01)03760244032492(11)190726(21)BS19050; 9) BS19052 / UDI - (01)03760244032492(11)190731(21)BS19052; 10) BS19054 / UDI - (01)03760244032492(11)190912(21)BS19054; 11) BS19057 / UDI - (01)03760244032492(11)191003(21)BS19057; 12) BS19058 / UDI - (01)03760244032492(11)191011(21)BS19058.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ROBOT ROSA ONE (USA) 3.1.1.0Medtech · Orthopedic Stereotaxic Instrument1 recall22 reports
- ROSA One 3.1.0.5Medtech · Neurological Stereotaxic Instrument1 recall2 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K151359, K172444, K182417 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAYFIELD HEAD HOLDER ADAPTOR ROSA3-177DMedtech · Neurological Stereotaxic Instrument1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 84642).
- Z-1150-2020 · TerminatedROSA Brain 3.0 Application-Brain
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗