COAPTIUM® Connect NCOCO3010Recalls & reported problems
Tissium · Sutureless Nerve Coaptation System COAPTIUM® CONNECT is a bioabsorbable coaptation system for sutureless peripheral nerve repair of peripheral nerves not in continuity. The system includes a single-use syringe pre-filled with a photoactive COAPTIUM® polymer which is used to secure an implantable three-dimensional coaptation chamber to the nerve segments that are distal and proximal to a peripheral nerve injury. · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03760279380117), received between Mar 26, 2026 and Mar 26, 2026 — classified as 1 injury.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 15, 2026Open, Classified
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Z-2905-2026 · Root cause: Incorrect or no expiration date · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Wound Dehiscence1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 26, 2026Adverse Event Without Identified Device or Use Problem
ON 26 JANUARY 2026, A COAPTIUM® CONNECT DEVICE WAS SUCCESSFULLY IMPLANTED TO REPAIR THE RADIAL DIGITAL NERVE OF THE THUMB FOLLOWING A TRAUMATIC INJURY. APPROXIMATELY FOUR WEEKS POST-OPERATIVELY, THE PATIENT PRESENTED WITH WOUND DEHISCENCE. THE DEVICE REMAINED IN SITU, WITH NO REPORTED EXTRUSION. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2026 TO EXPLORE THE REPAIR SITE AND ADDRESS THE WOUND COMPLICATION. INTRAOPERATIVELY, NO DEVICE MALFUNCTION OR ABNORMAL TISSUE RESPONSE WAS IDENTIFIED. THE DEVICE WAS EXPLANTED WITHOUT DAMAGE TO THE NATIVE NERVE TISSUE. SEPARATELY, DURING THE PROCEDURE, FAILURE OF THE TENDON REPAIR WAS OBSERVED, SUGGESTING SUBOPTIMAL POST-OPERATIVE COMPLIANCE. THE WOUND WAS DESCRIBED AS COLONIZED BUT WITHOUT SIGNS OF ACTIVE INFECTION. THERE IS NO EVIDENCE OF DEVICE MALFUNCTION, AND THE EVENT IS CONSIDERED PRIMARILY ATTRIBUTABLE TO PATIENT, PROCEDURAL, AND POST-OPERATIVE FACTORS.
Patient problems: Wound Dehiscence
MAUDE report 24700750 ↗
Other In Situ Polymerizing Peripheral Nerve Repair Device devices
- COAPTIUM® Connect NCOCO6010Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall1 report
- COAPTIUM® Connect NCOCO1510Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall0 reports
- COAPTIUM® Connect NCOCO1575Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall0 reports
- COAPTIUM® Connect NCOCO2075Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall0 reports
- COAPTIUM® Connect NCOCO4010Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall0 reports
- COAPTIUM® Connect NCOCO5010Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall0 reports
- COAPTIUM® Connect NCOCO2010Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.