Skip to content

COAPTIUM® Connect NCOCO3010Recalls & reported problems

Tissium · Sutureless Nerve Coaptation System COAPTIUM® CONNECT is a bioabsorbable coaptation system for sutureless peripheral nerve repair of peripheral nerves not in continuity. The system includes a single-use syringe pre-filled with a photoactive COAPTIUM® polymer which is used to secure an implantable three-dimensional coaptation chamber to the nerve segments that are distal and proximal to a peripheral nerve injury. · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03760279380117), received between Mar 26, 2026 and Mar 26, 2026 — classified as 1 injury.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 15, 2026Open, Classified

    Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

    Z-2905-2026 · Root cause: Incorrect or no expiration date · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Wound Dehiscence1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 26, 2026Adverse Event Without Identified Device or Use Problem

    ON 26 JANUARY 2026, A COAPTIUM® CONNECT DEVICE WAS SUCCESSFULLY IMPLANTED TO REPAIR THE RADIAL DIGITAL NERVE OF THE THUMB FOLLOWING A TRAUMATIC INJURY. APPROXIMATELY FOUR WEEKS POST-OPERATIVELY, THE PATIENT PRESENTED WITH WOUND DEHISCENCE. THE DEVICE REMAINED IN SITU, WITH NO REPORTED EXTRUSION. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2026 TO EXPLORE THE REPAIR SITE AND ADDRESS THE WOUND COMPLICATION. INTRAOPERATIVELY, NO DEVICE MALFUNCTION OR ABNORMAL TISSUE RESPONSE WAS IDENTIFIED. THE DEVICE WAS EXPLANTED WITHOUT DAMAGE TO THE NATIVE NERVE TISSUE. SEPARATELY, DURING THE PROCEDURE, FAILURE OF THE TENDON REPAIR WAS OBSERVED, SUGGESTING SUBOPTIMAL POST-OPERATIVE COMPLIANCE. THE WOUND WAS DESCRIBED AS COLONIZED BUT WITHOUT SIGNS OF ACTIVE INFECTION. THERE IS NO EVIDENCE OF DEVICE MALFUNCTION, AND THE EVENT IS CONSIDERED PRIMARILY ATTRIBUTABLE TO PATIENT, PROCEDURAL, AND POST-OPERATIVE FACTORS.

    Patient problems: Wound Dehiscence

    MAUDE report 24700750 ↗

Other In Situ Polymerizing Peripheral Nerve Repair Device devices

All In Situ Polymerizing Peripheral Nerve Repair Device devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.