Skip to content

COAPTIUM® Connect NCOCO6010Recalls & reported problems

Tissium · Sutureless Nerve Coaptation System COAPTIUM® CONNECT is a bioabsorbable coaptation system for sutureless peripheral nerve repair of peripheral nerves not in continuity. The system includes a single-use syringe pre-filled with a photoactive COAPTIUM® polymer which is used to secure an implantable three-dimensional coaptation chamber to the nerve segments that are distal and proximal to a peripheral nerve injury. · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03760279380209), received between Feb 24, 2026 and Feb 24, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 15, 2026Class IIOpen, Classified

    Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

    Z-2908-2026 · Root cause: Incorrect or no expiration date · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Fracture; Material Integrity Problem

    DURING A LEFT ULNAR NERVE RECONSTRUCTION USING TWO COAPTIUM® CONNECT DEVICES, AN APPROXIMATELY 3 MM FRACTURE LINE WAS OBSERVED NEAR THE DISTAL RIDGE REGION OF THE PROXIMAL COAPTATION CHAMBER WHILE ACCOMMODATING A NERVE MEASURING APPROXIMATELY 7 MM INTO A 6 MM CHAMBER. NO FRAGMENT DETACHMENT WAS OBSERVED AND NO MATERIAL WAS RETAINED; THE DEVICE WAS NOT REPLACED. AT THE DISTAL REPAIR SITE, FIXATION TO THE POSTERIOR EPINEURIUM WAS INCOMPLETE AND TWO SUTURES WERE PLACED THROUGH THE CHAMBER AND EPINEURIUM TO COMPLETE THE REPAIR. THE PROCEDURE WAS PROLONGED BY APPROXIMATELY 30 MINUTES. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24430041 ↗

FDA clearance and classification

The FDA regulates product code SFD (in situ polymerizing peripheral nerve repair device) under 21 CFR 882.5270 as a Class II device. 21 CFR 882.5270 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An in situ polymerizing peripheral nerve repair device is intended for the sutureless repair of peripheral nerve injuries.

Common questions about the COAPTIUM® Connect NCOCO6010

Has the COAPTIUM® Connect NCOCO6010 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2908-2026 (Class II), began on Jun 15, 2026: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the COAPTIUM® Connect NCOCO6010?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 24, 2026. That report lists device problems “Fracture” (1) and “Material Integrity Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the COAPTIUM® Connect NCOCO6010?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the COAPTIUM® Connect NCOCO6010 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the COAPTIUM® Connect NCOCO6010, and is it still sold?

The FDA's device database lists Tissium as the company responsible for it (brand COAPTIUM® Connect, model NCOCO6010). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other In Situ Polymerizing Peripheral Nerve Repair Device devices

All In Situ Polymerizing Peripheral Nerve Repair Device devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page