Potassium Electrode 3111628180Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Potassium Electrode (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04015630014293), received between Aug 22, 2018 and Aug 6, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings3
  • Non Reproducible Results2
  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results; Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE POTASSIUM RESULTS FROM THE COBAS B 221 4 ROCHE OMNI S4 SYSTEM. THE SAMPLES WERE REPEATED AFTER THE CUSTOMER CLEANED THE ANALYZER. PATIENT 1 INITIAL RESULT WAS 10.84 MMOL/L, AND THE REPEAT RESULT WAS 4.63 MMOL/L. PATIENT 2 INITIAL RESULT WAS 6.82 MMOL/L, AND THE REPEAT RESULT WAS 3.8 MMOL/L. THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22726604 ↗
  • MalfunctionJun 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 2 PATIENTS' SAMPLES TESTED FOR POTASSIUM (K) ON A COBAS B221 6 ROCHE OMNI S6 SYSTEM. PATIENT 1 AND PATIENT 2 WERE INITIALLY TESTED (TUBE 1) WITH AN ARTERIAL BLOOD GAS (ABG) TEST ON A COBAS B221, WHILE SIMULTANEOUSLY, VENOUS BLOOD SAMPLES (TUBE 2) WERE DRAWN FROM PATIENT 1 AND PATIENT 2 AND TESTED ON DIFFERENT ANALYZERS. PATIENT 1: TUBE 1 RESULT: 2.61 MMOL/L. TUBE 2 RESULT: 4.2 MMOL/L (TESTED ON A BECKMAN COULTER AU5800 INSTRUMENT). PATIENT 2 WAS TESTED ON (B)(6) 2025: TUBE 1 RESULT: 3.74 MMOL/L. TUBE 2 RESULT: 4.7 MMOL/L (TESTED ON A C503 ANALYZER).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22206167 ↗
  • MalfunctionMar 6, 2023Low Test Results; Low Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED A QUESTIONABLE RESULT FOR ONE PATIENT SAMPLE TESTED WITH THE POTASSIUM ELECTRODE ON A COBAS B 221 ANALYZER (SERIAL NUMBER (B)(4). THE WHOLE BLOOD SAMPLE WAS TESTED ON THE COBAS B 221 ANALYZER, RESULTING IN A POTASSIUM VALUE OF 2.85 MMOL/L.THE A SERUM SAMPLE COLLECTED AT THE SAME TIME AS THE COMPLAINED SAMPLE WAS TESTED ON A COBAS C 501 ANALYZER, RESULTING IN A POTASSIUM VALUE OF 4.1 MEQ/L. THE SERUM SAMPLE WAS REPEATED AGAIN ON THE C 501 ANALYZER ON (B)(6) 2023, RESULTING IN A VALUE OF 3.9 MMOL/L. THE RESULT WAS CORRECTED AFTER REPEAT TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16489549 ↗
  • MalfunctionJan 18, 2023High Test Results; High Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO PATIENT SAMPLES TESTED WITH THE POTASSIUM (K) AND SODIUM (NA) ELECTRODES ON A COBAS B 221 <6> SYSTEM. PATIENT SAMPLE 1: THE SAMPLE INITIALLY RESULTED IN A K VALUE OF 7.81 MMOL/L AND REPEATED AS 3.4 MMOL/L ON A SECOND COBAS B 221 SYSTEM. PATIENT SAMPLE 2: THE SAMPLE INITIALLY RESULTED IN A NA VALUE OF 124.8 MMOL/L AND REPEATED AS 135.6 MMOL/L ON A SECOND COBAS B 221 SYSTEM. THE SAMPLE INITIALLY RESULTED IN A K VALUE OF 16.74 MMOL/L AND REPEATED AS 5.38 MMOL/L ON A SECOND COBAS B 221 SYSTEM. THE REPEAT RESULTS WERE DEEMED CORRECT. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE K ELECTRODE LOT NUMBER WAS 21512951 WITH AN EXPIRATION DATE OF 21-JAN-2022. THE NA ELECTRODE LOT NUMBER AND EXPIRATION DATE WERE REQUESTED BUT NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16199621 ↗
  • MalfunctionJan 19, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO SAMPLES FROM THE SAME PATIENT TESTED WITH THE ROCHE OMNI ELECTRODE POTASSIUM ON A COBAS B 221 ROCHE OMNI S6 SYSTEM. THE SAMPLE VALUES WERE REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED BY A PHYSICIAN AS THE AVAILABLE OUTPUTS FROM PATIENT MONITORING DID NOT SUGGEST THAT THE PATIENT WAS HYPOKALEMIC. THE REPORTER ALSO MENTIONED THEY HAVE BEEN GETTING "SODIUM SIGNAL SENSOR UNSTABLE" ERRORS ON THE COBAS B 221 ANALYZER FOR SEVERAL MONTHS. THE REPORTER STATED THAT THE K ELECTRODE WAS LAST CHANGED ON THIS ANALYZER ON (B)(6) 2021. PER PRODUCT LABELING, THE K ELECTRODE IS STABLE FOR 26 WEEKS. THE FIRST SAMPLE FROM THE PATIENT RESULTED IN A K VALUE OF 3.14 MMOL/L WHEN TESTED ON THE COBAS B 221. THE SECOND SAMPLE FROM THE PATIENT RESULTED IN A K VALUE OF 3.14 MMOL/L WHEN TESTED ON THE COBAS B 221. A THIRD SAMPLE WAS COLLECTED FROM THE PATIENT AND TESTED ON A COBAS 6000 C (501) MODULE, RESULTING IN A K VALUE OF 4.2 MMOL/L. THIS VALUE WAS DEEMED CORRECT BY THE PHYSICIAN AS IT WAS CONSISTENT WITH THE CLINICAL PRESENTATION OF THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13290645 ↗
  • MalfunctionAug 22, 2018Non Reproducible Results; Non Reproducible Results

    THE CUSTOMER QUESTIONED RESULTS FOR 3 PATIENTS TESTED FOR POTASSIUM (K) ON A COBAS B 221 SYSTEM. BASED ON THE DATA PROVIDED, THE RESULTS FOR 1 PATIENT ARE A REPORTABLE MALFUNCTION. THE INITIAL RESULT AT 5:59 A.M. WAS 6.34 MMOL/L. THE SAME SAMPLE WAS REPEATED AT 6:08 A.M. AND THE RESULT WAS 7.21 MMOL/L. THESE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY BUT THE PHYSICIAN DISMISSED THEM AND NO CLINICAL DECISIONS WERE MADE. THERE WAS NO ALLEGATION THAT AN ADVERSE EVENT OCCURRED. THE B 221 INSTRUMENT SERIAL NUMBER WAS (B)(4). THE K ELECTRODE WAS REQUESTED FOR INVESTIGATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7804341 ↗

Other ELECTRODE, ION SPECIFIC, POTASSIUM devices

All ELECTRODE, ION SPECIFIC, POTASSIUM devices →