Potassium Electrode 3111628180Recalls & reported problems
Roche Diagnostics GmbH · Class 2
All models of Potassium Electrode (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04015630014293), received between Aug 22, 2018 and Aug 6, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings3
- Non Reproducible Results2
- High Test Results1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results; Incorrect, Inadequate or Imprecise Result or Readings; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE POTASSIUM RESULTS FROM THE COBAS B 221 4 ROCHE OMNI S4 SYSTEM. THE SAMPLES WERE REPEATED AFTER THE CUSTOMER CLEANED THE ANALYZER. PATIENT 1 INITIAL RESULT WAS 10.84 MMOL/L, AND THE REPEAT RESULT WAS 4.63 MMOL/L. PATIENT 2 INITIAL RESULT WAS 6.82 MMOL/L, AND THE REPEAT RESULT WAS 3.8 MMOL/L. THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22726604 ↗MalfunctionJun 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 2 PATIENTS' SAMPLES TESTED FOR POTASSIUM (K) ON A COBAS B221 6 ROCHE OMNI S6 SYSTEM. PATIENT 1 AND PATIENT 2 WERE INITIALLY TESTED (TUBE 1) WITH AN ARTERIAL BLOOD GAS (ABG) TEST ON A COBAS B221, WHILE SIMULTANEOUSLY, VENOUS BLOOD SAMPLES (TUBE 2) WERE DRAWN FROM PATIENT 1 AND PATIENT 2 AND TESTED ON DIFFERENT ANALYZERS. PATIENT 1: TUBE 1 RESULT: 2.61 MMOL/L. TUBE 2 RESULT: 4.2 MMOL/L (TESTED ON A BECKMAN COULTER AU5800 INSTRUMENT). PATIENT 2 WAS TESTED ON (B)(6) 2025: TUBE 1 RESULT: 3.74 MMOL/L. TUBE 2 RESULT: 4.7 MMOL/L (TESTED ON A C503 ANALYZER).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22206167 ↗MalfunctionMar 6, 2023Low Test Results; Low Test Results
THE INITIAL REPORTER STATED THEY RECEIVED A QUESTIONABLE RESULT FOR ONE PATIENT SAMPLE TESTED WITH THE POTASSIUM ELECTRODE ON A COBAS B 221 ANALYZER (SERIAL NUMBER (B)(4). THE WHOLE BLOOD SAMPLE WAS TESTED ON THE COBAS B 221 ANALYZER, RESULTING IN A POTASSIUM VALUE OF 2.85 MMOL/L.THE A SERUM SAMPLE COLLECTED AT THE SAME TIME AS THE COMPLAINED SAMPLE WAS TESTED ON A COBAS C 501 ANALYZER, RESULTING IN A POTASSIUM VALUE OF 4.1 MEQ/L. THE SERUM SAMPLE WAS REPEATED AGAIN ON THE C 501 ANALYZER ON (B)(6) 2023, RESULTING IN A VALUE OF 3.9 MMOL/L. THE RESULT WAS CORRECTED AFTER REPEAT TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16489549 ↗MalfunctionJan 18, 2023High Test Results; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO PATIENT SAMPLES TESTED WITH THE POTASSIUM (K) AND SODIUM (NA) ELECTRODES ON A COBAS B 221 <6> SYSTEM. PATIENT SAMPLE 1: THE SAMPLE INITIALLY RESULTED IN A K VALUE OF 7.81 MMOL/L AND REPEATED AS 3.4 MMOL/L ON A SECOND COBAS B 221 SYSTEM. PATIENT SAMPLE 2: THE SAMPLE INITIALLY RESULTED IN A NA VALUE OF 124.8 MMOL/L AND REPEATED AS 135.6 MMOL/L ON A SECOND COBAS B 221 SYSTEM. THE SAMPLE INITIALLY RESULTED IN A K VALUE OF 16.74 MMOL/L AND REPEATED AS 5.38 MMOL/L ON A SECOND COBAS B 221 SYSTEM. THE REPEAT RESULTS WERE DEEMED CORRECT. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE K ELECTRODE LOT NUMBER WAS 21512951 WITH AN EXPIRATION DATE OF 21-JAN-2022. THE NA ELECTRODE LOT NUMBER AND EXPIRATION DATE WERE REQUESTED BUT NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16199621 ↗MalfunctionJan 19, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO SAMPLES FROM THE SAME PATIENT TESTED WITH THE ROCHE OMNI ELECTRODE POTASSIUM ON A COBAS B 221 ROCHE OMNI S6 SYSTEM. THE SAMPLE VALUES WERE REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED BY A PHYSICIAN AS THE AVAILABLE OUTPUTS FROM PATIENT MONITORING DID NOT SUGGEST THAT THE PATIENT WAS HYPOKALEMIC. THE REPORTER ALSO MENTIONED THEY HAVE BEEN GETTING "SODIUM SIGNAL SENSOR UNSTABLE" ERRORS ON THE COBAS B 221 ANALYZER FOR SEVERAL MONTHS. THE REPORTER STATED THAT THE K ELECTRODE WAS LAST CHANGED ON THIS ANALYZER ON (B)(6) 2021. PER PRODUCT LABELING, THE K ELECTRODE IS STABLE FOR 26 WEEKS. THE FIRST SAMPLE FROM THE PATIENT RESULTED IN A K VALUE OF 3.14 MMOL/L WHEN TESTED ON THE COBAS B 221. THE SECOND SAMPLE FROM THE PATIENT RESULTED IN A K VALUE OF 3.14 MMOL/L WHEN TESTED ON THE COBAS B 221. A THIRD SAMPLE WAS COLLECTED FROM THE PATIENT AND TESTED ON A COBAS 6000 C (501) MODULE, RESULTING IN A K VALUE OF 4.2 MMOL/L. THIS VALUE WAS DEEMED CORRECT BY THE PHYSICIAN AS IT WAS CONSISTENT WITH THE CLINICAL PRESENTATION OF THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13290645 ↗MalfunctionAug 22, 2018Non Reproducible Results; Non Reproducible Results
THE CUSTOMER QUESTIONED RESULTS FOR 3 PATIENTS TESTED FOR POTASSIUM (K) ON A COBAS B 221 SYSTEM. BASED ON THE DATA PROVIDED, THE RESULTS FOR 1 PATIENT ARE A REPORTABLE MALFUNCTION. THE INITIAL RESULT AT 5:59 A.M. WAS 6.34 MMOL/L. THE SAME SAMPLE WAS REPEATED AT 6:08 A.M. AND THE RESULT WAS 7.21 MMOL/L. THESE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY BUT THE PHYSICIAN DISMISSED THEM AND NO CLINICAL DECISIONS WERE MADE. THERE WAS NO ALLEGATION THAT AN ADVERSE EVENT OCCURRED. THE B 221 INSTRUMENT SERIAL NUMBER WAS (B)(4). THE K ELECTRODE WAS REQUESTED FOR INVESTIGATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7804341 ↗
Other ELECTRODE, ION SPECIFIC, POTASSIUM devices
- FlexLab FLXInpeco SA · Electrode, Ion Specific, Potassium8 recalls17 reports
- Aptio Automation AP2Inpeco SA · Electrode, Ion Specific, Potassium5 recalls6 deaths16 reports
- Accelerator a3600 ACPInpeco SA · Electrode, Ion Specific, Potassium5 recalls13 reports
- FlexLab X FXXInpeco SA · Electrode, Ion Specific, Potassium2 recalls2 reports
- EPOC BGEM CREA TEST CARD CT-1006-00-00Epocal Inc · Electrode, Ion Specific, Potassium2 recalls1 death1 report
- cobas b 221 <6> system 3337154001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM1 recall43 reports
- EPOC READER HR-1002-00-00Epocal Inc · Electrode, Ion Specific, Potassium11 deaths88 reports
- cobas b 221 <4> system 3337138001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM1 recall7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page