cobas b 221 <2> system 3337111001Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630018307), received between Aug 30, 2021 and Aug 30, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 22, 2018Class IITerminated

    The software responsible for starting scheduled AutoQC measurements (scheduler) will not activate.

    Z-1610-2018 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation1
  • Suction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2021Suction Problem; Material Deformation; Suction Problem; Material Deformation

    THE INITIAL REPORTER COMPLAINED OF AUTO QC (AQC) ERRORS ON A COBAS B221 SYSTEM. THE CUSTOMER OPENED THE AQC DRAWER AND OBSERVED THAT THE MOTOR WAS NOT IN THE CORRECT POSITION AND WOULD NOT MOVE WHEN PROGRAMMED TO MOVE. THE QC MECHANISM WAS STUCK AND WOULD NOT RELEASE. THE FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER SITE AND FOUND A BURNT TUBE. NO ONE WAS INJURED BY THE BURNT TUBE AND NO QUESTIONABLE RESULTS WERE GENERATED DUE TO THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12382784 ↗