Elecsys HBsAg Confirmatory Test 11820648122Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04015630911110), received between Feb 2, 2023 and Sep 17, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result3
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2025False Positive Result; False Positive Result
THE INITIAL REPORTER ALLEGED DISCREPANT REACTIVE RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ELECSYS HBSAG II ASSAY AND THE ELECSYS HBSAG CONFIRMATORY TEST ON A COBAS E 801 ANALYTICAL UNIT. ON (B)(6) 2022, THE ELECSYS HBSAG II ASSAY GENERATED A RESULT OF 1.51 COI (REACTIVE), AND THE ELECSYS HBSAG CONFIRMATORY TEST ALSO GENERATED A REACTIVE RESULT. ADDITIONAL TESTING ON THE SAME SAMPLE USING THE ELECSYS ANTI-HBS (AHBS) ASSAY PRODUCED A RESULT OF 2.0 IU/L (NON-REACTIVE). TESTING OF DIFFERENT BLOOD DRAWS FROM THE SAME PATIENT WAS CONDUCTED ON (B)(6) 2022 USING COMPETITOR SYSTEMS (ABBOTT AND SIEMENS). BOTH SYSTEMS REPORTED NON-REACTIVE RESULTS FOR HBSAG AND AHBS. NO FURTHER RESULT DETAILS WERE PROVIDED FOR THESE TESTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23073717 ↗MalfunctionSep 16, 2025False Positive Result; False Positive Result
THE INITIAL REPORTER ALLEGED INACCURACY IN RESULTS OBTAINED WITH THE HBSAG CONFIRMATORY TEST ELECSYS ON THE COBAS E 801 MODULE. THE SAMPLE WAS INITIALLY TESTED FOR HBSAG WITH RESULTS OF 1.93 COI AND 1.37 COI UPON REPETITION. THE SAMPLE WAS SUBSEQUENTLY TESTED USING THE HBSAG CONFIRMATORY TEST IN TWO SEPARATE SESSIONS. IN THE FIRST SESSION, THE SAMPLE WITH CONTROL REAGENT YIELDED A RESULT OF 0.888 COI, AND THE SAMPLE WITH CONFIRMATORY REAGENT YIELDED A RESULT OF 0.328 COI. CONTROL 2 WITH CONTROL REAGENT YIELDED A RESULT OF 3.96 COI, AND CONTROL 2 WITH CONFIRMATORY REAGENT YIELDED A RESULT OF 0.530 COI. IN THE SECOND SESSION, THE SAMPLE WITH CONTROL REAGENT YIELDED A RESULT OF 0.870 COI, AND THE SAMPLE WITH CONFIRMATORY REAGENT YIELDED A RESULT OF 0.359 COI. CONTROL 2 WITH CONTROL REAGENT YIELDED A RESULT OF 4.13 COI, AND CONTROL 2 WITH CONFIRMATORY REAGENT YIELDED A RESULT OF 0.526 COI. THE SAMPLE WAS REPORTED AS REACTIVE. TESTING OF THE SAME PATIENT SAMPLE AT A DIFFERENT LABORATORY YIELDED A NEGATIVE HBSAG RESULT, WHICH WAS CONFIRMED BY MOLECULAR TESTING. THE CUSTOMER INDICATED THAT THE INITIAL DILUTION OF THE SAMPLE WITH CONFIRMATORY AND CONTROL REAGENTS WAS PERFORMED CORRECTLY AT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23066832 ↗MalfunctionMay 14, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER QUESTIONED NEGATIVE RESULTS FOR 4 PATIENT SAMPLES TESTED FOR ELECSYS HBSAG CONFIRMATORY TESTS (HBSAG CONFIRMATORY) ON A COBAS E 801 ANALYTICAL UNIT WHERE THE INITIAL HBSAG RESULTS WERE HIGH (POSITIVE). OF THE DATA PROVIDED, DISCREPANT RESULTS WERE IDENTIFIED FOR 1 PATIENT SAMPLE BETWEEN 2 DIFFERENT HBSAG CONFIRMATORY REAGENTS. THIS MEDWATCH WILL COVER HBSAG CONFIRMATORY REAGENT WITH MATERIAL NUMBER 11820648122. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE HBSAG CONFIRMATORY REAGENT WITH MATERIAL NUMBER 09127127190. REFER TO THE ATTACHED DATA FOR THE PATIENT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19311832 ↗MalfunctionFeb 2, 2023False Positive Result; False Positive Result
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HBSAG CONFIRMATORY TEST RESULTS FOR 4 PATIENT SAMPLES ON A COBAS E 801 ANALYTICAL UNIT. THE CUSTOMER SUSPECTS THAT THE HBSAG CONFIRMATORY RESULTS ARE FALSE POSITIVE IN RELATION TO THE ANTI-HEPATITIS B SURFACE ANTIBODY AND ANTI-HEPATITIS B CORE ANTIBODY RESULTS. THEY WERE QUESTIONING WHETHER THIS PATTERN OF RESULTS WAS POSSIBLE. ON (B)(6) 2022, PATIENT B'S SAMPLE WAS TESTED ON LINE C AND HAD TWO REACTIVE HBSAG TEST RESULTS. THE FIRST RESULT WAS 2.04 COI AND THE SECOND RESULT AFTER RE-CENTRIFUGATION WAS 1.82 COI. THE HBSAG CONFIRMATORY TEST WAS PERFORMED AND THE RESULT WAS REACTIVE WITH A QUANTITATIVE RESULT OF 0.436. THE ANTI-HEPATITIS B SURFACE ANTIBODY RESULT WAS NON-REACTIVE WITH A QUANTITATIVE RESULT OF <2 MIU/ML. THE ANTI-HEPATITIS B CORE ANTIBODY RESULT WAS NON-REACTIVE WITH A QUANTITATIVE RESULT OF 2.02 COI. PATIENT C'S SAMPLE WAS TESTED ON LINE C AND HAD THREE REACTIVE HBSAG TEST RESULTS. THE FIRST RESULT WAS 1.5 COI, THE SECOND RESULT WAS 1.56 COI, AND THE THIRD RESULT AFTER RE-CENTRIFUGATION WAS 1.5 COI. THE HBSAG CONFIRMATORY TEST WAS PERFORMED AND THE RESULT WAS REACTIVE WITH A QUANTITATIVE RESULT OF 0.474. THE ANTI-HEPATIT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16285810 ↗
Other Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices
- Elecsys HBsAg II 08814848190Roche Diagnostics GmbH · Test, hepatitis b (b core, be antigen, be antibody, b core igm)93 reports
- ARCHITECT 4P53-25Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall16 reports
- VITROS 6801925Ortho Clinical Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall15 reports
- ARCHITECT 1L82-27Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall4 reports
- VITROS 6801926Ortho Clinical Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall0 reports
- ARCHITECT 4P54-25Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)27 reports
- ADVIA Centaur® HBsII 10492138Siemens Healthcare Diagnostics Inc. · Test, Hepatitis B (B Core, Be Antigen, Be Antibody, B Core Igm)18 reports
- Atellica™ IM HBsII 10995604Siemens Healthcare Diagnostics Inc. · Test, hepatitis b (b core, be antigen, be antibody, b core igm)12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page