Aspartate Aminotransferase acc to IFCC without pyridoxal phosphate activation 20764949322Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04015630913855), received between Jul 26, 2019 and Nov 26, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “High Test Results” (4 reports).
7 other recalls cover devices cleared under the same FDA submission (K951595); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results4
- Incorrect, Inadequate or Imprecise Result or Readings3
- Incorrect Measurement1
- Low Test Results1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ASPARTATE AMINOTRANSFERASE (AST) RESULTS FOR 2 PATIENT SAMPLES ON A COBAS 8000 COBAS C 502 MODULE. THIS MEDWATCH WILL COVER AST. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE ALP RESULTS, MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE CALCIUM RESULTS, MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE LACTATE RESULTS, AND MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE LDH RESULTS. THE DOCTOR QUESTIONED THE RESULTS AS THEY WERE HIGH AND DID NOT MATCH THE CLINICAL STATUS OF OUTPATIENTS IN "NORMAL" CONDITION. NO SPECIFIC SAMPLE RESULTS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23657156 ↗MalfunctionDec 30, 2024High Test Results; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITHOUT PYRIDOXAL PHOSPHATE ACTIVATION (AST) ON A COBAS 8000 C 502 MODULE. THE SAMPLE INITIALLY RESULTED IN AN AST VALUE OF 93.9 U/L. THE PATIENT QUESTIONED THE RESULT, SO THE SAMPLE WAS REPEATED. THE SAMPLE WAS REPEATED ON (B)(6) 2024, RESULTING IN AN AST VALUE OF 20.6 U/L. A SECOND TUBE COLLECTED AT THE SAME TIME WAS ALSO REPEATED ON A SECOND ANALYZER ON (B)(6) 2024, RESULTING IN A VALUE OF 19.9 U/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21038996 ↗MalfunctionOct 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE ASSAY RESULTS FROM MULTIPLE PATIENT SAMPLES TESTED ON THE COBAS 6000 C 501 (UL) V. THE REPORTER PROVIDED EXAMPLES OF QUESTIONABLE ASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITHOUT PYRIDOXAL PHOSPHATE ACTIVATION (ASTL) AND CREP GEN.2 (CREP2) RESULTS. THIS MEDWATCH IS FOR THE ASPARTATE AMINOTRANSFERASE ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR THE CREATININE PLUS VER.2 ASSAY. THE INITIAL RESULT FROM THE ANALYZER WAS 65 U/L. THE REPEAT RESULT FROM ANOTHER ANALYZER WAS 26 U/L. THE REPEAT RESULT WAS DEEMED CORRECT. THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE PATIENT SAMPLE WAS RERUN BECAUSE OF MULTIPLE INVALIDATING FLAGS FOR OTHER ASSAYS. THE REPORTER SAID THEY ACCIDENTALLY RAN A WHOLE BLOOD PATIENT SAMPLE FOR A COMPLETE METABOLIC PANEL (CMP) ASSAY BEFORE THE REPORTED PATIENT SAMPLE WAS RUN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20482619 ↗MalfunctionJan 23, 2024Non Reproducible Results; Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ASTL AND ALTL ON AN INTEGRA 400 PLUS SYSTEM. THE CUSTOMER NOTED THAT THE PATIENT SAMPLE WAS ICTERIC. THIS MEDWATCH WILL APPLY TO THE AST ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE ALT ASSAY. THE SAMPLE INITIALLY RESULTED IN AN AST VALUE OF 77 U/L WITH A DATA FLAG. THE SAMPLE WAS REPEATED TWICE, RESULTING IN AST VALUES OF 279 U/L WITH A DATA FLAG AND 715 U/L WITH A DATA FLAG. THE SAMPLE WAS PLACED IN A CUP AND REPEATED TWICE, RESULTING IN AST VALUES OF 731 U/L WITH A DATA FLAG AND 732 U/L WITH A DATA FLAG. THE SAMPLE INITIALLY RESULTED IN AN ALT VALUE OF 36 U/L WITH A DATA FLAG AND IT REPEATED AS 1027 U/L WITH A DATA FLAG. THE SAMPLE WAS PLACED IN A CUP AND REPEATED TWICE, RESULTING IN ALT VALUES OF 1044 U/L WITH A DATA FLAG AND 1063 U/L WITH A DATA FLAG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18564147 ↗MalfunctionAug 23, 2021Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect Measurement; Low Test Results
THE INITIAL REPORTER STATED THEY HAVE BEEN HAVING ISSUES WITH CONTROL RECOVERY AND PROFICIENCY SURVEY RECOVERY FOR ASTL ASPARTATE AMINOTRANSFERASE - PYRIDOXAL PHOSPHATE ACTIVATED ON THE COBAS INTEGRA 400 PLUS (SERIAL NUMBER (B)(4)) AND AN UNKNOWN COBAS 6000 SYSTEM (SERIAL NUMBER UNKNOWN) AT A SECOND SITE. THE REPORTER REPEATED PATIENT SAMPLES IN ORDER TO TROUBLESHOOT THE ISSUE AND FOUND DISCREPANT RESULTS FOR FOUR PATIENT SAMPLES. THE FIRST SAMPLE INITIALLY RESULTED IN AN AST VALUE OF 18 U/L WHEN TESTED ON THE INTEGRA 400 PLUS ANALYZER AND THIS VALUE WAS REPORTED OUTSIDE OF THE LABORATORY. THE ASSAY WAS RECALIBRATED AND THE SAMPLE WAS THEN REPEATED ON THE INTEGRA ON (B)(6) 2021, RESULTING IN A VALUE OF 28 U/L. THE 28 U/L VALUE WAS BELIEVED TO BE CORRECT. THE SAMPLE WAS ALSO REPEATED THREE TIMES ON A COBAS 6000 SYSTEM AT A SECOND SITE ON (B)(6) 2021, RESULTING IN VALUES OF 30 U/L, 22 U/L, AND 18 U/L. THE FIELD SERVICE ENGINEER CHECKED THE INTEGRA ANALYZER OPTICS AND CHANGED THE HALOGEN LAMP. CONTROL RECOVERY IMPROVED AFTER THIS. THE CUSTOMER THEN ENCOUNTERED ISSUES WITH THE SECOND, THIRD, AND FOURTH COMPLAINED SAMPLES. THE SECOND SAMPLE INITIALLY RESULTED IN AN AST VALUE OF 18 U/L W
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12355478 ↗MalfunctionJul 26, 2019High Test Results; High Test Results
THIS REPORT SUMMARIZES <NOE>1</NOE> MALFUNCTION EVENT. DISCREPANT LOW RESULTS WERE GENERATED BY A COBAS 6000 C (501) MODULE. THE EVENTS INVOLVED A TOTAL OF 1 PATIENT WITH THE FOLLOWING: THREE SAMPLES WITH ERRONEOUS RESULTS FOR ASTL ASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITHOUT PYRIDOXAL PHOSPHATE ACTIVATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8830157 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K951595. One submission often covers many models, so check the product and lot details in each record.
- Z-1675-2021 · Apr 26, 2021 · Open, ClassifiedIRON GEN 2: ROCHE COBAS INTEGRA ANALYZER, REAGENT CASSETTES, Catalog#: 03183696122 - Product Usage: In vitro test for the quantitative determination of iron in human serum and plas
- Z-0100-2017 · Jul 15, 2016 · TerminatedRoche COBAS INTEGRA 800 /800 CTS Analyzer Analyzer, Chemistry (Photometric, Discrete for Clinical use Product Usage: The system cobas p612 is a computer-controlled stand-alone sys
- Z-0168-2014 · Sep 16, 2013 · TerminatedCOBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed
- Z-0696-2013 · Nov 19, 2012 · TerminatedRoche COBAS INTEGRA 400 and 400 plus Analyzer, an in vitro diagnostic analyzer.
- Z-0007-2009 · Jun 11, 2008 · TerminatedRoche Diagnostics IRON2, Iron Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 03183696122. An in-vitro diagnostic reagent system intended for use on
- Z-0012-2009 · Jun 11, 2008 · TerminatedRoche Diagnostics NAPA2, N-Acetyl-Procainamide, COBAS INTEGRA, Roche Diagnostics, Indianapolis, IN.; 20737852322. An in-vitro diagnostic reagent system intended for use on COBAS I
- Z-0667-03 · Feb 13, 2003 · TerminatedCOBAS INTEGRA Creatinine Jaffe for use on the INTEGRA 400, 700 and 800 analyzers; system ID # 0764345, catalog # 20764345
Other Nadh oxidation/nad reduction, ast/sgot devices
- Aspartate Aminotransferase acc. to IFCC II 08104719190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot5 reports
- Aspartate Aminotransferase acc to IFCC without pyridoxal phosphate activation 5850819190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot4 reports
- Aspartate Aminosransferase acc. to IFCC with pyridoxal phosphate activation 08056838190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot4 reports
- Aspartate Aminotransferase acc to IFCC with pyridoxal phosphate activation 4467493190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot3 reports
- Piccolo AmLyte 13 400-0041Abaxis, Inc. · Nadh Oxidation/Nad Reduction, Ast/Sgot1 report
- ARCHITECT 7D81-22Abbott GmbH · Nadh oxidation/nad reduction, ast/sgot1 report
- Piccolo General Chemistry 13 400-0029Abaxis, Inc. · Nadh Oxidation/Nad Reduction, Ast/Sgot1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page