Aspartate Aminosransferase acc. to IFCC with pyridoxal phosphate activation 08056838190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613336121023), received between Jul 3, 2023 and Jan 10, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Non Reproducible Results2
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 10, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ASPARTATE AMINOTRANSFERASE (AST) AND ALANINE AMINOTRANSFERASE (ALT) RESULTS FOR ONE PATIENT FROM THE COBAS PURE C 303 ANALYTICAL UNIT. THE CUSTOMER ALLEGED SOMETHING WITH THE PATIENT SAMPLE CAUSED DIFFERENT RESULTS FOR THE ASSAYS WITH PYRIDOXAL PHOSPHATE ACTIVATION COMPARED TO THE ASSAYS WITHOUT AS WERE USED ON THE COBAS C311. THIS MEDWATCH WILL COVER AST. REFER TO THE MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR THE ALT ASSAY. FOR AST: THE INITIAL RESULT WAS 243 U/L. THE REPEAT RESULTS USING THE SAME ANALYZER WERE 247 U/L AND 249 U/L. THE REPEAT RESULTS USING TWO COBAS C311 ANALYZERS WERE 66.9 U/L, 70 U/L, AND 70 U/L. ON (B)(6) 2024, A NEW SAMPLE FROM THE PATIENT HAD RESULTS OF 195 U/L AND 200 U/L. THE RESULT FROM THE COBAS C311 ANALYZER WERE 64 U/L AND 65.4 U/L. THE RESULTS FROM THE COBAS C311 WERE BELIEVED CORRECT AND REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21127332 ↗MalfunctionSep 3, 2024Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITH PYRIDOXAL PHOSPHATE ACTIVATION (ASTP) RESULTS FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 35.8 UI/L WITH A DATA FLAG. THE REPEAT RESULT WITH AN AUTOMATIC 1:10 DILUTION WAS 62.4 UI/L. THE REPEAT RESULTS WITH NO DILUTION WERE 24.9 UI/L AND 25.3 UI/L. THE RESULT OF 25 UIU/L WAS BELIEVED TO BE CORRECT. THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20130302 ↗MalfunctionFeb 28, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THERE WAS AN ALLEGATION OF QUESTIONABLE ASTPM ASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITH PYRIDOXAL PHOSPHATE ACTIVATION RESULTS FOR 2 PATIENT SAMPLES ON A COBAS PURE C 303 ANALYTICAL UNIT. FOR SAMPLE 1, THE INITIAL AST RESULT WAS 70.7 U/L. THE SAMPLE WAS REPEATED ON A COBAS INTEGRA ANALYZER AND THE REPEAT RESULT WAS 45.7 U/L. FOR SAMPLE 2, THE INITIAL AST RESULT WAS 75.0 U/L. THE SAMPLE WAS REPEATED ON A COBAS INTEGRA ANALYZER AND THE REPEAT RESULT WAS 35.7 U/L. THE EXACT DATES OF TESTING FOR SAMPLE 2 WERE NOT PROVIDED. THE COBAS INTEGRA RESULTS WERE DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18798188 ↗MalfunctionJul 3, 2023High Test Results; High Test Results
THE INITIAL REPORTER COMPLAINED OF A DISCREPANT HIGH RESULT FOR 1 PATIENT SAMPLE TESTED FOR ASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITH PYRIDOXAL PHOSPHATE ACTIVATION (ASTP) ON A COBAS PRO C 503 ANALYTICAL UNIT COMPARED TO A DXC 700 AU ANALYZER. THE RESULT FROM THE C503 ANALYZER WAS 385 U/L. THE RESULT FROM A DXC 700 AU ANALYZER WAS 181.99 U/L. THE QUESTIONABLE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17250596 ↗
Other Nadh oxidation/nad reduction, ast/sgot devices
- Aspartate Aminotransferase acc to IFCC without pyridoxal phosphate activation 20764949322Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot6 reports
- Aspartate Aminotransferase acc. to IFCC II 08104719190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot5 reports
- Aspartate Aminotransferase acc to IFCC without pyridoxal phosphate activation 5850819190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot4 reports
- Aspartate Aminotransferase acc to IFCC with pyridoxal phosphate activation 4467493190Roche Diagnostics GmbH · Nadh oxidation/nad reduction, ast/sgot3 reports
- Piccolo AmLyte 13 400-0041Abaxis, Inc. · Nadh Oxidation/Nad Reduction, Ast/Sgot1 report
- ARCHITECT 7D81-22Abbott GmbH · Nadh oxidation/nad reduction, ast/sgot1 report
- Piccolo General Chemistry 13 400-0029Abaxis, Inc. · Nadh Oxidation/Nad Reduction, Ast/Sgot1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page