Skip to content

Tina-quant D-Dimer 3001245322Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Tina-quant D-Dimer (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630914593), received between Aug 26, 2021 and May 11, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 3, 2023Open, Classified

    Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.

    Z-0352-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE LOW TINA-QUANT D-DIMER RESULTS FROM THE COBAS 4000 C311 STAND ALONE SYSTEM WHEN COMPARED TO A REFERENCE LABORATORY. OF THE DATA PROVIDED, THE RESULT FOR ONE SAMPLE WERE A REPORTABLE MALFUNCTION. THE INITIAL RESULT WAS 0.35 (<0.5) G/ML. THE RESULT FROM A REFERENCE LABORATORY USING A STA COMPACT MAX ANALYZER WAS 2.00 G/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25134911 ↗
  • MalfunctionAug 26, 2021High Test Results; High Test Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE D-DI TINA-QUANT D-DIMER RESULTS FOR ONE PATIENT SAMPLE WITH THE COBAS 6000 C501 MODULE SERIAL NUMBER (B)(4). THE INITIAL RESULT WAS 4677 NG/ML. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED BY THE PHYSICIAN. THE REPEATED RESULT WAS 4550 NG/ML. THE PHYSICIAN QUESTIONED THIS RESULT AS WELL. THE CUSTOMER SENT THE SAMPLE TO THREE DIFFERENT LABORATORIES FOR REPEAT TESTING. THE FIRST LABORATORY TESTED THE SAMPLE WITH A COBAS C501 USING A REAGENT FROM A THIRD PARTY AND RECEIVED A RESULT OF 474 NG/ML (CUTOFF < 250 DDU NG/ML). THE SECOND LABORATORY TESTED THE SAMPLE WITH A COBAS C502 USING A ROCHE REAGENT AND RECEIVED A RESULT OF 5.1 MG/L (CUTOFF 0.5 FEU MG/L). ON (B)(6) 2021, THE THIRD LABORATORY TESTED THE SAMPLE WITH A SIEMENS ANALYZER AND RECEIVED A RESULT OF 0.9 FEU MG/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12369463 ↗

Other FIBRIN SPLIT PRODUCTS devices

All FIBRIN SPLIT PRODUCTS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page