Tina-quant D-DimerRecalls & reported problems across all models
Roche Diagnostics GmbH · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Tina-quant D-Dimer model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 2 models of Tina-quant D-Dimer by Roche Diagnostics GmbH in the FDA's device registry (GUDID). Together they are named in 2 adverse event reports in the FDA's MAUDE database, received between Aug 26, 2021 and May 11, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Tina-quant D-Dimer recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 3, 2023Open, Classified
Elevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.
Z-0352-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗
Most reported Tina-quant D-Dimer problems
Device problems
- High Test Results1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Tina-quant D-Dimer models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Tina-quant D-Dimer 3001245322FIBRIN SPLIT PRODUCTS1 recall2 reports
- Tina-quant D-Dimer 08105618190FIBRIN SPLIT PRODUCTS1 recall0 reports
Recent Tina-quant D-Dimer adverse event reports
The 2 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Low Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE LOW TINA-QUANT D-DIMER RESULTS FROM THE COBAS 4000 C311 STAND ALONE SYSTEM WHEN COMPARED TO A REFERENCE LABORATORY. OF THE DATA PROVIDED, THE RESULT FOR ONE SAMPLE WERE A REPORTABLE MALFUNCTION. THE INITIAL RESULT WAS 0.35 (<0.5) G/ML. THE RESULT FROM A REFERENCE LABORATORY USING A STA COMPACT MAX ANALYZER WAS 2.00 G/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25134911 ↗MalfunctionAug 26, 2021High Test Results; High Test Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE D-DI TINA-QUANT D-DIMER RESULTS FOR ONE PATIENT SAMPLE WITH THE COBAS 6000 C501 MODULE SERIAL NUMBER (B)(4). THE INITIAL RESULT WAS 4677 NG/ML. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED BY THE PHYSICIAN. THE REPEATED RESULT WAS 4550 NG/ML. THE PHYSICIAN QUESTIONED THIS RESULT AS WELL. THE CUSTOMER SENT THE SAMPLE TO THREE DIFFERENT LABORATORIES FOR REPEAT TESTING. THE FIRST LABORATORY TESTED THE SAMPLE WITH A COBAS C501 USING A REAGENT FROM A THIRD PARTY AND RECEIVED A RESULT OF 474 NG/ML (CUTOFF < 250 DDU NG/ML). THE SECOND LABORATORY TESTED THE SAMPLE WITH A COBAS C502 USING A ROCHE REAGENT AND RECEIVED A RESULT OF 5.1 MG/L (CUTOFF 0.5 FEU MG/L). ON (B)(6) 2021, THE THIRD LABORATORY TESTED THE SAMPLE WITH A SIEMENS ANALYZER AND RECEIVED A RESULT OF 0.9 FEU MG/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12369463 ↗
Other Roche Diagnostics GmbH products
- cobas e 411 Immunoassay Analyzer2 models · 1,218 reports
- cobas e 801 Immunoassay analyzer3 models · 713 reports · 1 recall
- CoaguChek XS PT strips2 models · 632 reports
- CoaguChek XS Patient Self Testing Care Kit2 models · 346 reports
- Elecsys FT3 III4 models · 283 reports
- Elecsys Estradiol III2 models · 162 reports
- Elecsys TSH4 models · 155 reports
- Sodium electrode3 models · 150 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page