Potassium Electrode 21029355001Recalls & reported problems
Roche Diagnostics GmbH · Class 2
All models of Potassium Electrode (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630918737), received between Nov 16, 2023 and Feb 20, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2024High Test Results; High Test Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE POTASSIUM ELECTRODE ON A COBAS INTEGRA 400 PLUS. THE SAMPLE INITIALLY RESULTED IN A POTASSIUM VALUE OF 7.4 MMOL/L, WITH A DATA FLAG INDICATING THE ISE WAS UNSTABLE. THE SAMPLE WAS AUTOMATICALLY REPEATED, RESULTING IN A POTASSIUM VALUE OF 8.5 MMOL/L WITH A DATA FLAG. THE CUSTOMER RE-CALIBRATED THE ISE TESTS, BUT CALIBRATION FAILED FOR POTASSIUM. THE CUSTOMER TRIED REPLACING THE ISE MIX TOWER, PRIMED THE ISE CALIBRATORS, AND ACTIVATED THE ELECTRODES, BUT THE POTASSIUM CALIBRATION STILL FAILED. DUE TO THE CALIBRATION FAILURE, THE SAMPLE WAS REPEATED AT ANOTHER LABORATORY. THE PATIENT SAMPLE WAS SENT TO ANOTHER LABORATORY FOR TESTING ON AN UNKNOWN ANALYZER, RESULTING IN A POTASSIUM VALUE OF 4.4 MMOL/L. THE REPEAT VALUE WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18738232 ↗MalfunctionNov 16, 2023High Test Results; High Test Results
THERE WAS AN ALLEGATION OF ISE NOISE ALARMS AND QUESTIONABLE ISE INDIRECT GEN 2 POTASSIUM RESULTS FROM THE COBAS PRO ISE ANALYTICAL UNIT. ONE EXAMPLE WAS PROVIDED OF AN INITIAL RESULT WAS 7.16 MMOL/L AND A REPEAT RESULT OF 4.27 MMOL/L. THE QUESTIONABLE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE REPEAT RESULT WAS BELIEVED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18153889 ↗
FDA clearance and classification
Cleared through 510(k) K963627 on Oct 24, 1996 (decision: Substantially Equivalent). Submitted by Roche Diagnostic Systems, Inc. as “(Modified Version) Integra Ion Selective Electrode (Ise) Module”. The FDA review took 43 days. FDA record ↗
The FDA regulates product code CEM (electrode, ion specific, potassium) under 21 CFR 862.1600 as a Class II device. 21 CFR 862.1600 ↗
Common questions about the Potassium Electrode 21029355001
Has the Potassium Electrode 21029355001 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04015630918737), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Potassium Electrode 21029355001?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Nov 16, 2023 and Feb 20, 2024. In those reports, the most common specific device problem is “High Test Results” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Potassium Electrode 21029355001?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Potassium Electrode 21029355001 FDA approved?
It was cleared through 510(k) K963627 on Oct 24, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Potassium Electrode 21029355001, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand Potassium Electrode, model 21029355001). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ELECTRODE, ION SPECIFIC, POTASSIUM devices
- FlexLab FLXInpeco SA · Electrode, Ion Specific, Potassium8 recalls16 reports
- Aptio Automation AP2Inpeco SA · Electrode, Ion Specific, Potassium5 recalls6 deaths16 reports
- Accelerator a3600 ACPInpeco SA · Electrode, Ion Specific, Potassium5 recalls13 reports
- FlexLab X FXXInpeco SA · Electrode, Ion Specific, Potassium2 recalls2 reports
- EPOC BGEM CREA TEST CARD CT-1006-00-00Epocal Inc · Electrode, Ion Specific, Potassium2 recalls1 death1 report
- cobas b 221 <6> system 3337154001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM1 recall43 reports
- EPOC READER HR-1002-00-00Epocal Inc · Electrode, Ion Specific, Potassium11 deaths87 reports
- cobas b 221 <4> system 3337138001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM1 recall7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page