Elecsys ACTH Test System 3255751190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630921522), received between Jan 31, 2023 and Apr 23, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE ELECSYS CORTISOL II AND ELECSYS ACTH TEST SYSTEM (ACTH) RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 6000 E601 MODULE. THIS MEDWATCH IS FOR THE ELECSYS ACTH TEST SYSTEM (ACTH) ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR THE ELECSYS CORTISOL II ASSAY REPORT. THE INITIAL RESULT OF THE INITIAL PATIENT SAMPLE WAS 626.6 PG/ML. THE REPEAT RESULT WAS 581.0 PG/ML. THE RESULT OF A NEW PATIENT SAMPLE WAS 7.33 PG/ML. THE PHYSICIAN QUESTIONED THE INITIAL RESULT AS IT DID NOT MATCH THE PREVIOUS RESULTS, PROMPTING THE PATIENT SAMPLE TO BE RERUN. THE RESULT OF THE NEW PATIENT SAMPLE WAS DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21892557 ↗MalfunctionJan 31, 2023High Test Results; High Test Results
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS ACTH TEST SYSTEM (ACTH) RESULTS FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 8000 - COBAS E 602 MODULE WITH SERIAL NUMBER: (B)(4). THE INITIAL RESULTS WERE REPORTED TO THE PHYSICIAN WHO REQUESTED A RERUN. THE REPORTER STATED THAT THEY BELIEVE THERE WAS AN ANALYTICAL INTERFERENCE WITH THE RESULT FROM THIS PATIENT SAMPLE. THE PATIENT SAMPLE WAS ALSO RERUN ON THEIR OTHER E 602 MODULE WITH SERIAL NUMBER: (B)(4), BUT THE REPORTER WAS NOT ABLE TO PROVIDE THE RESULT FROM THIS MODULE. THE PATIENT SAMPLE WAS ALSO SENT TO ANOTHER LABORATORY THAT USES A SIEMENS ANALYZER. THE INITIAL RESULT FROM THE MODULE USING THE PRIMARY TUBE WAS 102.4 PG/ML. THE FIRST REPEAT RESULT FROM THE MODULE USING THE PRIMARY TUBE WAS 101.7 PG/ML. THE SECOND REPEAT RESULT FROM THE OTHER LABORATORY'S ANALYZER USING AN ALIQUOT SAMPLE WAS 64 PG/ML. THE REFERENCE RANGE IS 6-50 PG/ML. THE REPORTER DILUTED THE PATIENT SAMPLES AS PART OF THEIR INVESTIGATION. THE THIRD REPEAT RESULT FROM THE MODULE USING THE PATIENT SAMPLE IN A 1:2 DILUTION WAS 111.6 PG/ML. THE FOURTH REPEAT RESULT FROM THE MODULE USING THE PATIENT SAMPLE IN A 1:4 DILUTION WAS 124.4 PG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16272046 ↗
FDA clearance and classification
Cleared through 510(k) K060585 on May 3, 2006 (decision: Substantially Equivalent). Submitted by Roche Diagnostics Corp. as “Roche Elecsys- Acth,Acth Calset,Acth Calcheck,Precicontrol Acth”. The FDA review took 57 days. FDA record ↗
The FDA regulates product code CKG (radioimmunoassay, acth) under 21 CFR 862.1025 as a Class II device. 21 CFR 862.1025 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Elecsys ACTH Test System 3255751190
Has the Elecsys ACTH Test System 3255751190 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04015630921522), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Elecsys ACTH Test System 3255751190?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 31, 2023 and Apr 23, 2025. In those reports, the most common specific device problems are “High Test Results” (2) and “Non Reproducible Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Elecsys ACTH Test System 3255751190?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Elecsys ACTH Test System 3255751190 FDA approved?
It was cleared through 510(k) K060585 on May 3, 2006. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Elecsys ACTH Test System 3255751190, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand Elecsys ACTH Test System, model 3255751190). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other RADIOIMMUNOASSAY, ACTH devices
- IMMULITE®/IMMULITE® 1000 Systems ACT ACTH 10381415Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, ACTH6 reports
- Elecsys ACTH 07026684190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, ACTH3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page