Elecsys ACTH 07026684190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630939893), received between Jul 25, 2022 and Dec 29, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Non Reproducible Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2025High Test Results; High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE HIGH RESULTS FOR 1 PATIENT TESTED FOR ELECSYS ACTH (ACTH) ON A COBAS E 801 ANALYTICAL UNIT. THE ACTH RESULTS ARE ALLEGEDLY HIGH WITH NO KNOWN CAUSE, AS THE PATIENT IS ASYMPTOMATIC (PREVIOUSLY CUSHING'S DISEASE). THE INITIAL RESULT WAS 118,000 PG/ML. THE SAMPLE WAS REPEATED WITH VARIOUS DILUTIONS: X5: > 2000 PG/ML WITH A DATA FLAG, X10: >2000 PG/ML WITH A DATA FLAG, X20: >2000 PG/ML WITH A DATA FLAG, X40: >2000 PG/ML WITH A DATA FLAG, X75: 118,350 PG/ML, X100: 126,400 PG/ML. ON (B)(6) 2025, A NEW SAMPLE WAS OBTAINED, AND THE RESULT WAS APPROXIMATELY 25,000 PG/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23905893 ↗
  • MalfunctionJun 5, 2024Non Reproducible Results; Non Reproducible Results

    THERE IS A SUSPECTED INTERFERENCE WITH THE ELECSYS ACTH ASSAY DUE TO QUESTIONABLE RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT. ON 15-JUN-2022 THE INITIAL RESULT WAS > 2000 PG/ML. THE REPEATED RESULT WAS 39480 PG/ML FOLLOWING A DILUTION. ON 15-SEP-2022 THE INITIAL RESULT WAS > 2000 PG/ML. THE REPEATED RESULT WAS 29000 PG/ML FOLLOWING A DILUTION. ON 10-MAR-2023 THE INITIAL RESULT WAS > 2000 PG/ML. THE REPEATED RESULT WAS 61050 PG/ML FOLLOWING A DILUTION. ON 28-JUL-2023 THE INITIAL RESULT WAS > 2000 PG/ML. THE REPEATED RESULT WAS 7375 PG/ML FOLLOWING A DILUTION. ON 16-JAN-2024 THE INITIAL RESULT WAS > 2000 PG/ML. THE REPEATED RESULT WAS 71850 PG/ML FOLLOWING A DILUTION. THE SAMPLE WAS TESTED USING THE SIEMENS IMMULITE METHOD WITH MULTIPLE DILUTIONS AND THE RESULTS WERE HIGH. THE NUMERIC RESULTS WERE NOT PROVIDED. THE REPEATED RESULTS WERE BELIEVED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19471990 ↗
  • MalfunctionJul 25, 2022High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ACTH ELECSYS E2G RESULTS FROM A COBAS E 801 MODULE. THE SERIAL NUMBER WAS REQUESTED BUT WAS NOT PROVIDED. THE SAMPLE FROM (B)(6) 2022 WAS SENT FOR INVESTIGATION AND TESTED ON A COBAS E 801 MODULE AND A COBAS E 602 MODULE. REFER TO THE ATTACHMENT TO THE MEDWATCH FOR PATIENT DATA. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY TO THE PHYSICIAN. THE CUSTOMER BELIEVED THE RESULTS FROM (B)(6) 2021 AND (B)(6) 2022 WERE INCORRECT BUT BELIEVED THE RESULTS FROM (B)(6) 2022 WERE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15094293 ↗