Skip to content

alpha-Amylase EPS ver.2 5167027190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630924875), received between Mar 23, 2026 and Mar 23, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

1 other recall cover devices cleared under the same FDA submission (K972250); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2026Low Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE AMYLASE EPS VER.2 ASSAY RESULTS FOR A PATIENT SAMPLE TESTED ON A COBAS 8000 CORE UNIT. THE INITIAL RESULT WAS -1 U/L (ACCOMPANIED BY A DATA FLAG). THE REPEAT RESULT WAS 199 U/L WITH A 5-FOLD DILUTION. THE SAMPLE WAS REPEATED NEAT AND THE RESULT WAS 0 (ACCOMPANIED BY A DATA FLAG), AND WITH A 1:5 DILUTION, THE RESULT WAS 198 U/L. NO ERRONEOUS RESULT WAS RELEASED OUTSIDE OF THE LABORATORY. AN INTERFERENCE IS SUSPECTED IN THE SAMPLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24660351 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K972250. One submission often covers many models, so check the product and lot details in each record.

  • Z-0688-2008 · Dec 3, 2007 · Terminated
    Roche LIPC Lipase Colorimetric Reagent; Catalog Number: 03029590322. Roche Diagnostics Corp., Indianapolis, IN 46250

Other CATALYTIC METHODS, AMYLASE devices

All CATALYTIC METHODS, AMYLASE devices →