Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336121009), received between Aug 7, 2024 and Aug 7, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Non Reproducible Results” (1 report).

1 other recall cover devices cleared under the same FDA submission (K972250); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2024Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER QUESTIONED HIGH RESULTS NOT CORRESPONDING TO THE CLINICAL PICTURE FOR 1 PATIENT TESTED FOR LIPASE COLORIMETRIC ASSAY (LIPC) AND A-AMYLASE EPS VER. 2 (AMYL2) ON 2 COBAS C 503 ANALYTICAL UNITS. THIS MEDWATCH WILL COVER AMYL2. REFER TO MEDWATCH WITH A1: PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE LIPC RESULTS. ON (B)(6) 2024 THE INITIAL LIPC RESULT FROM C503 ANALYZER A WAS 292 U/L. THE AMYL2 RESULT WAS 161 U/L. ON (B)(6) 2024 THE PATIENT WAS TESTED IN A DIFFERENT LABORATORY USING A COBAS 6000 C501 MODULE WHERE THE LIPC RESULT WAS 38 U/L AND THE AMYL2 RESULT WAS 86 U/L. THESE RESULTS WERE BELIEVED TO BE CORRECT. ON (B)(6) 2024 THE ORIGINAL SAMPLE WAS REPEATED ON THE CUSTOMER'S BACKUP C503 ANALYZER B AND THE LIPC RESULT WAS 279 U/L AND THE AMYL2 RESULT WAS 164 U/L. ON (B)(6) 2024 A NEW SAMPLE WAS OBTAINED AND RUN ON C503 ANALYZER A WITH A LIPC RESULT OF 102 U/L AND AN AMYL2 RESULT OF 124 U/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19929735 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K972250. One submission often covers many models, so check the product and lot details in each record.

  • Z-0688-2008 · Dec 3, 2007 · Terminated
    Roche LIPC Lipase Colorimetric Reagent; Catalog Number: 03029590322. Roche Diagnostics Corp., Indianapolis, IN 46250

Other CATALYTIC METHODS, AMYLASE devices

All CATALYTIC METHODS, AMYLASE devices →