Creatine Kinase 5168546190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Creatine Kinase (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630924974), received between Dec 8, 2023 and May 17, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Incorrect Measurement1
  • Non Reproducible Results1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Renal Failure1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 17, 2024Incorrect Measurement; Incorrect Measurement

    THERE WAS AN ALLEGATION OF QUESTIONABLE CK CREATINE KINASE RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 8000 C702 MODULE. ON (B)(6) 2024, THE INITIAL CK RESULT WAS 8238 U/L. THE SAMPLE WAS REPEATED WITH AN INCREASED DILUTION AND THE RESULT WAS 3958 U/L. THE SAMPLE WAS REPEATED AGAIN AND THE RESULT WAS 10250 U/L. ON (B)(6) 2024, THE SAMPLE WAS REPEATED A THIRD TIME AND THE RESULT WAS 8256 U/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19345754 ↗
  • InjuryDec 11, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE INITIAL REPORTER ALLEGED THERE WAS A DISCREPANT RESULT FOR ONE PATIENT SAMPLE TESTED WITH CK ON A COBAS 8000 C702 MODULE WHICH LED TO A DELAY IN THE PATIENT RECEIVING THE CORRECT DIAGNOSIS AND THE APPROPRIATE TREATMENT. THE PATIENT REPORTEDLY PRESENTED TO THE HOSPITAL WITH WHOLE-BODY ACHES, SPECIFICALLY IN THE ABDOMEN, REDUCED SENSATION AND STRENGTH IN ONE LEG, AND VOMITING. THE PATIENT REPORTEDLY FELL A FEW DAYS BEFORE ADMISSION. THE PATIENT REPORTEDLY HAD A CREATININE VALUE OF APPROXIMATELY 400 UMOL/L AND WAS ALLEGEDLY IN RENAL FAILURE AT THE TIME OF ADMISSION. THE SAMPLE WAS REPORTEDLY COLLECTED FROM THE PATIENT AT THE TIME THE PATIENT WAS ADMITTED TO THE HOSPITAL. THE SAMPLE REPORTEDLY HAD LOW VOLUME, SO THE ANALYSIS WAS PERFORMED WITH THE SAMPLE PLACED INTO A MICRO-CUP ON TOP OF A SAMPLE TUBE. THE SAMPLE INITIALLY RESULTED IN A CK VALUE OF 414.2 U/L. THE PATIENT WAS ALLEGEDLY INCORRECTLY DIAGNOSED WITH COLD SEPSIS DUE TO THE VALUE. A SECOND SAMPLE COLLECTED FROM THE PATIENT ON (B)(6) 2023 RESULTED IN A CK VALUE OF 99843 U/L, SO THE LOW VALUE FROM THE FIRST SAMPLE WAS QUESTIONED BY THE PHYSICIAN. THE SAMPLE WAS REPEATED ON (B)(6) 2023, RESULTING IN A CK VALUE OF - 305.3 U/L

    Patient problems: Renal Failure

    MAUDE report 18304186 ↗
  • MalfunctionDec 8, 2023Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Non Reproducible Results; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Non Reproducible Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH CK ON A COBAS 8000 C702 MODULE. THE CUSTOMER ALLEGES THAT THE RESULTS MAY INDICATE THERE IS AN INTERFERENT IN THE SAMPLE THAT REACTS WITH THE ASSAY. THE SAMPLE INITIALLY RESULTED IN A CK VALUE OF 1026 IU/L ACCOMPANIED BY A DATA FLAG INDICATING THE REACTION LIMIT WAS EXCEEDED. THE SAMPLE REPEATED WITH A CK VALUE OF > 34041 IU/L ACCOMPANIED BY A DATA FLAG. THE SAMPLE WAS DILUTED X20 AND REPEATED TWICE ON A SECOND ANALYZER, RESULTING IN CK VALUES OF 22 IU/L AND 30 IU/L. THE SAMPLE WAS TESTED AGAIN WITHOUT DILUTION ON THE COMPLAINED ANALYZER, RESULTING IN A CK VALUE OF 30 IU/L. THE SAMPLE WAS DILUTED AGAIN USING AN UNKNOWN DILUTION FACTOR AND REPEATED ON AN UNKNOWN ANALYZER, RESULTING IN A CK VALUE OF 30 IU/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18293647 ↗

Other ENZYMATIC METHOD, CREATININE devices

All ENZYMATIC METHOD, CREATININE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page