Iron Gen.2 5169291190Recalls & reported problems
Roche Diagnostics GmbH · Class 1
All models of Iron Gen.2 (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630925049), received between Jan 2, 2025 and Aug 28, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2025Non Reproducible Results; Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR IRON GEN.2 (IRON2) ON A COBAS 8000 C702 MODULE. THE INITIAL RESULT FROM THE C702 MODULE WAS 545 UG/DL. THE DOCTOR QUESTIONED THIS RESULT. ON (B)(6) 2025, THE SAMPLE WAS TESTED BY THE BECKMAN METHOD WITH A RESULT OF 267 UG/DL. THE RESULT FROM THE BECKMAN METHOD WAS BELIEVED TO BE CORRECT. ON (B)(6) 2025, THE CUSTOMER DILUTED THE SAMPLE TO TEST FOR INTERFERENCE, AND THE FOLLOWING RESULTS WERE OBTAINED: 1:3 DILUTION = 611 UG/DL, 1:5 DILUTION = 620 UG/DL, 1:10 DILUTION = 564 UG/DL, 1:50 DILUTION = 419 UG/DL. THE CUSTOMER SUSPECTS AN INTERFERENCE AFFECTING THE ROCHE RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22914263 ↗MalfunctionJan 2, 2025Non Reproducible Results; Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR IRON2 ON A COBAS 8000 C 702 MODULE. THE INITIAL RESULT WAS 951 UG/DL. THE REPEAT RESULT WAS 58 UG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21054420 ↗
Other Photometric method, iron (non-heme) devices
- Iron Gen.2 3183696122Roche Diagnostics GmbH · Photometric method, iron (non-heme)1 recall16 reports
- ADVIA® Chemistry IRON_2 Iron 2 Reagents 10377510Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall1 report
- Atellica CH Iron_2 11097601Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall1 report
- Alinity 08P3920Abbott Laboratories · Photometric method, iron (non-heme)1 recall0 reports
- ARCHITECT 6K95-30Abbott Laboratories · Photometric method, iron (non-heme)1 recall0 reports
- ARCHITECT 6K95-41Abbott Laboratories · Photometric method, iron (non-heme)1 recall0 reports
- ADVIA® Chemistry IRON_2 Reagents 10341118Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall0 reports
- Atellica CH Iron3 (Iron3) 11537211Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page