Skip to content

Iron Gen.2 5169291190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

All models of Iron Gen.2 (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630925049), received between Jan 2, 2025 and Aug 28, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Non Reproducible Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Non Reproducible Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2025Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR IRON GEN.2 (IRON2) ON A COBAS 8000 C702 MODULE. THE INITIAL RESULT FROM THE C702 MODULE WAS 545 UG/DL. THE DOCTOR QUESTIONED THIS RESULT. ON (B)(6) 2025, THE SAMPLE WAS TESTED BY THE BECKMAN METHOD WITH A RESULT OF 267 UG/DL. THE RESULT FROM THE BECKMAN METHOD WAS BELIEVED TO BE CORRECT. ON (B)(6) 2025, THE CUSTOMER DILUTED THE SAMPLE TO TEST FOR INTERFERENCE, AND THE FOLLOWING RESULTS WERE OBTAINED: 1:3 DILUTION = 611 UG/DL, 1:5 DILUTION = 620 UG/DL, 1:10 DILUTION = 564 UG/DL, 1:50 DILUTION = 419 UG/DL. THE CUSTOMER SUSPECTS AN INTERFERENCE AFFECTING THE ROCHE RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22914263 ↗
  • MalfunctionJan 2, 2025Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR IRON2 ON A COBAS 8000 C 702 MODULE. THE INITIAL RESULT WAS 951 UG/DL. THE REPEAT RESULT WAS 58 UG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21054420 ↗

Other Photometric method, iron (non-heme) devices

All Photometric method, iron (non-heme) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page