Uric Acid ver.2 5171857190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04015630925124), received between Dec 13, 2024 and Sep 24, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Incorrect Measurement1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888731 ↗
  • MalfunctionMay 4, 2026Incorrect Measurement

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 5 PATIENTS' SAMPLES TESTED WITH URIC ACID VER.2 ASSAY ON A COBAS 8000 C702 MODULE. SAMPLE 1: INITIAL RESULT: 702.2 MMOL/L. REPEAT RESULT: 244.4 MMOL/L. SAMPLE 2: INITIAL RESULT: 683.9 MMOL/L. REPEAT RESULT: 242.9 MMOL/L. SAMPLE 3: INITIAL RESULT: 692.6 MMOL/L. REPEAT RESULT: 298.8 MMOL/L. SAMPLE 4: INITIAL RESULT: 681.7 MMOL/L. REPEAT RESULT: 229.4 MMOL/L. SAMPLE 5: INITIAL RESULT: 675.9 MMOL/L. REPEAT RESULT: 439.4 MMOL/L. THE CUSTOMER QUESTIONED THE INITIAL RESULTS AS THEY DID NOT MATCH THE PATIENTS' CLINICAL DIAGNOSIS, AND REPEATED THE SAMPLES AFTER REPLACING THE REAGENT. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE RESULTS WITH THE LOWER VALUES WERE DEEMED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25073316 ↗
  • MalfunctionSep 18, 2025High Test Results; High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE URIC ACID VER.2 RESULTS FROM THE COBAS 8000 C702 MODULE. THE CUSTOMER REPLACED THE REAGENT PACK AND PERFORMED REPEAT TESTING. SAMPLE (B)(6), INITIAL RESULT WAS 636 UMOL/L, AND THE REPEAT RESULT WAS 312 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 583 UMOL/L, AND THE REPEAT RESULT WAS 218 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 733 UMOL/L, AND THE REPEAT RESULT WAS 419 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 648 UMOL/L, AND THE REPEAT RESULT WAS 308 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 621 UMOL/L, AND THE REPEAT RESULT WAS 268 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 595 UMOL/L, AND THE REPEAT RESULT WAS 227 UMOL/L. SAMPLE (B)(6) INITIAL RESULT WAS 528 UMOL/L, AND THE REPEAT RESULT WAS 85 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 713 UMOL/L, AND THE REPEAT RESULT WAS 394 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 587 UMOL/L, AND THE REPEAT RESULT WAS 194 UMOL/L. SAMPLE (B)(6), INITIAL RESULT WAS 650 UMOL/L, AND THE REPEAT RESULT WAS 295 UMOL/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23082834 ↗
  • MalfunctionDec 13, 2024Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE URIC ACID (UA2) RESULTS FOR ONE PATIENT SAMPLE TESTED ON A COBAS 8000 CORE UNIT INSTRUMENT. TWO SAMPLES FROM THE SAME PATIENT WERE TESTED WHERE ONE RESULT WAS 120 UMOL/L AND THE OTHER RESULT WAS <12 UMOL/L. THE INITIAL RESULT OF 120 UMOL/L WAS QUESTIONED AS IT DID NOT MATCH THE PATIENT'S PREVIOUS RESULTS. THE SAMPLE WAS REPEATED TWICE AND THE RESULT WAS <12 UMOL/L BOTH TIMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20924999 ↗

Other Acid, uric, uricase (colorimetric) devices

All Acid, uric, uricase (colorimetric) devices →