Creatinine Jaff Gen2 compensated) 5401755190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04015630926824), received between Mar 13, 2024 and Mar 3, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results4
- High Test Results1
- Incorrect Measurement1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Low Test Results
THERE WAS AN ALLEGATION OF A DISCREPANT LOW RESULT FOR 1 PATIENT SAMPLE TESTED FOR CREATININE JAFFE GEN.2 (COMPENSATED) (CREAJ2) ON A COBAS C 111 ANALYZER. THE INITIAL RESULT WAS 0.67 MG/DL. THE REPEAT RESULT FROM A DIFFERENT LABORATORY WAS 1.03 MG/DL. THE REPEAT RESULT AFTER RECALIBRATION WAS 1.02 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24500355 ↗MalfunctionAug 22, 2025Non Reproducible Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE CREATININE JAFFE GEN.2 (COMPENSATED) RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS C 111 ANALYZER. THE INITIAL RESULT WAS 3.2 UMOL/L. THE FIRST REPEAT RESULT WAS 128.2 UMOL/L. THE SECOND REPEAT RESULT WAS 123.5 UMOL/L. NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22873010 ↗MalfunctionAug 11, 2025Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CREATININE JAFFÉ GEN.2 (COMPENSATED) RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 111 ANALYZER. THE INITIAL CREATININE RESULT WAS 1.13 MG/DL. THE NEXT DAY, THE SAMPLE WAS REPEATED AND THE RESULT WAS 1.15 MG/DL. THE ANALYZER WAS RECALIBRATED AND AN ALIQUOT OF THE SAMPLE WAS REPEATED AND THE RESULT WAS 1.58 MG/DL. A SEPARATE SAMPLE WAS COLLECTED ON (B)(6) 2025 AND TESTED AT ANOTHER LABORATORY AND THE CREATININE RESULT WAS 0.68 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22770707 ↗MalfunctionMay 23, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH CREATININE JAFFE GEN.2 (CREAJ2) ASSAY ON A COBAS C 111 ANALYZER. INITIAL RESULT: 2.73 MG/DL. ON (B)(6) 2025, THE SAMPLE WAS REPEATED. 1ST REPEAT RESULT: 0.50 MG/DL. 2ND REPEAT RESULT: 0.43 MG/DL (TESTED AT ANOTHER BRANCH).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22077395 ↗MalfunctionNov 4, 2024Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE CREATININE JAFFÉ GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 111 ANALYZER. THE INITIAL RESULT WAS 3.90 MG/DL. THE REPEATED RESULT WAS 0.8 MG/DL. THE REPEATED RESULT WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20602179 ↗MalfunctionMar 13, 2024Incorrect Measurement; Incorrect Measurement
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S PLASMA SAMPLE TESTED WITH CREATININE JAFFÉ GEN.2 (CREJ2) ASSAY ON A COBAS C111 SYSTEM. INITIAL RESULT: 394.8 MOL/L. THE CUSTOMER DOUBTED THE RESULT. THE SAMPLE WAS REPEATED 3 TIMES AND THE REPEAT RESULTS WERE 90.3 MOL/L, 90.5 MOL/L, AND 89.3 MOL/L. NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18896332 ↗
Other ALKALINE PICRATE, COLORIMETRY, CREATININE devices
- Dimension® Flex® reagent cartridge CRE2 10872079Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE1 recall9 reports
- ABX PENTRA CREATININE 120 CP 1220001933Horiba Abx Sas · Alkaline Picrate, Colorimetry, Creatinine1 recall0 reports
- ARCHITECT 3L81-22Abbott Laboratories · ALKALINE PICRATE, COLORIMETRY, CREATININE31 reports
- Alinity 07P9920Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE24 reports
- Atellica CH Crea_2 11097596Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE19 reports
- Creatinine Jaffe Gen.2 08057532190Roche Diagnostics GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE17 reports
- Alinity 04T9120Abbott Ireland · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
- Alinity 07P9930Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page