Skip to content

Triglycerides/Glycerol Blanked 05976006190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630928125), received between Sep 21, 2018 and Sep 21, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Non Reproducible Results” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K873049); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Non Reproducible Results1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2018Non Reproducible Results; Non Reproducible Results

    THE CUSTOMER COMPLAINED OF LOW DISCREPANT RESULTS FOR 3 PATIENT SAMPLES TESTED FOR TRIGB TRIGLYCERIDES/GLYCEROL BLANKED (TRIGB) ON A COBAS 8000 C 702 MODULE. PATIENT 1 INITIAL TRIGB RESULT WAS 319 MG/DL. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY WHERE IT WAS QUESTIONED BY THE DOCTOR. ON (B)(6) 2018 THE SAMPLE WAS REPEATED AND THE RESULT WAS 232 MG/DL WITH A DATA FLAG. THE SAMPLE WAS REPEATED WITH AUTO-DILUTION AND THE RESULT WAS 4121 MG/DL. ON (B)(6) 2018 PATIENT 2 (FEMALE, (B)(6)) INITIAL TRIGB RESULT WAS 558 MG/DL. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY WHERE IT WAS QUESTIONED BY THE DOCTOR. ON (B)(6) 2018 THE SAMPLE WAS REPEATED AND THE RESULT WAS 514 MG/DL WITH A DATA FLAG. THE SAMPLE WAS REPEATED WITH AUTO-DILUTION AND THE RESULT WAS 3511 MG/DL. ON (B)(6) 2018 PATIENT 3 (FEMALE, (B)(6)) INITIAL TRIGB RESULT WAS 681 MG/DL. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY WHERE IT WAS QUESTIONED BY THE DOCTOR. THE SAMPLE WAS REPEATED WITH AUTO-DILUTION AND THE RESULT WAS 3337 MG/DL. THE REPEAT RESULTS USING AUTO-DILUTION WERE BELIEVED TO BE CORRECT. NO ADVERSE EVENT OCCURRED. THE C702 MODULE SERIAL NUMBER WAS (B)(4). THE FIELD SERVICE ENGINEER (FSE) VISITE

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7898155 ↗

FDA clearance and classification

Cleared through 510(k) K873049 on Aug 28, 1987 (decision: Substantially Equivalent). Submitted by Boehringer Mannheim Corp. as “Triglycerides Gpo Without Free Glycerol”. The FDA review took 24 days. FDA record ↗

The FDA regulates product code CDT (lipase hydrolysis/glycerol kinase enzyme, triglycerides) under 21 CFR 862.1705 as a Class I device. 21 CFR 862.1705 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K873049. One submission often covers many models, so check the product and lot details in each record.

  • Z-1125-2019 · Dec 14, 2018 · Terminated
    cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number: 04657594190 - Product
  • Z-1930-2015 · May 7, 2015 · Terminated
    Triglycerides/Glycerol Blanked. In vitro test for the quantitative determination of triglycerides in human serum and plasma or Roche/Hitachi MODULAR systems, and cobas c systems.

Common questions about the Triglycerides/Glycerol Blanked 05976006190

Has the Triglycerides/Glycerol Blanked 05976006190 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04015630928125), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Triglycerides/Glycerol Blanked 05976006190?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 21, 2018. That report lists device problem “Non Reproducible Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Triglycerides/Glycerol Blanked 05976006190?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Triglycerides/Glycerol Blanked 05976006190 FDA approved?

It was cleared through 510(k) K873049 on Aug 28, 1987. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Triglycerides/Glycerol Blanked 05976006190, and is it still sold?

The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand Triglycerides/Glycerol Blanked, model 05976006190). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES devices

All LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page