Atellica CH Trig 11097591Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Triglycerides (Trig) Atellica CH - RGT 2000T · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00630414596495), received between Nov 7, 2021 and Nov 7, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K133067); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jan 27, 2023Class IIOpen, Classified
Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results
Z-1270-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 7, 2021Low Test Results; Low Test Results
AN ATELLICA CH 930 INSTRUMENT GENERATED A LIPEMIC INDEX VALUE OF 0 ON A PATIENT SAMPLE. THE SAMPLE WAS VISUALLY INSPECTED BY THE OPERATOR AND WAS RATED AS HAVING A LIPEMIC INDEX OF 4. THE SAME SAMPLE WAS REPORTEDLY TESTED FOR TRIGLYCERIDE, SODIUM AND POTASSIUM. THE SAMPLE WAS REPEAT TESTED FOR SODIUM AND POTASSIUM USING AN ALTERNATE INSTRUMENT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE LIPEMIC INDEX.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12763124 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133067. One submission often covers many models, so check the product and lot details in each record.
- Z-1657-2016 · Mar 17, 2016 · TerminatedADVIA CHEMISTRY TRIGLYCERIDES_2 (TRIG_2) REAGENT & ADVIA CHEMISTRY TRIGLYCERIDES_2 Concentrated (TRIG_c) REAGENT
- Z-1670-2016 · Mar 7, 2016 · TerminatedADVIA Chemistry Triglyceride_2, concentrated; TRIG_c; Catalog # SMN # 10697575. For in vitro diagnostic use in the quantitative measurement of triglycerides in human serum and pla
Other LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES devices
- Atellica CH Trig_2 11537222Siemens Healthcare Diagnostics Inc. · LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES1 recall0 reports
- Triglycerides 08058687190Roche Diagnostics GmbH · LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES21 reports
- Triglycerides 20767107322Roche Diagnostics GmbH · LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES13 reports
- Triglycerides 05171407190Roche Diagnostics GmbH · LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES11 reports
- Triglycerides 04657594190Roche Diagnostics GmbH · LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES1 report
- Triglycerides/Glycerol Blanked 05976006190Roche Diagnostics GmbH · LIPASE HYDROLYSIS/GLYCEROL KINASE ENZYME, TRIGLYCERIDES1 report
- AU OSR6130Beckman Coulter, Inc. · Lipase Hydrolysis/Glycerol Kinase Enzyme, Triglycerides1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page