Magnesium Gen.2 06407358190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04015630930821), received between Jun 28, 2022 and May 19, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Non Reproducible Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Non Reproducible Results3
  • High Test Results2
  • Low Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE MAGNESIUM GEN.2 RESULTS FROM THE COBAS 8000 C702 MODULE. SAMPLE 1'S INITIAL RESULT WAS 0.8 MG/DL ACCOMPANIED BY A DATA FLAG, AND THE REPEAT RESULT WAS 1.7 MG/DL. SAMPLE 2'S INITIAL RESULT WAS 0.9 MG/DL ACCOMPANIED BY A DATA FLAG, AND THE REPEAT RESULT WAS 1.7 MG/DL. SAMPLE 3'S INITIAL RESULT WAS 0.8 MG/DL ACCOMPANIED BY A DATA FLAG, AND THE REPEAT RESULT WAS 2.1 MG/DL. THE REPEAT RESULTS WERE DEEMED TO BE CORRECT. THE INITIAL RESULTS WERE QUESTIONED BECAUSE THEY WERE FLAGGED AS A CRITICAL RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22043383 ↗
  • MalfunctionApr 28, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE MAGNESIUM GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 8000 C702 MODULE. THE INITIAL RESULT WAS 2.67 MG/DL. THE SAMPLE WAS REPEATED ON ANOTHER MODULE, AND THE RESULTS WERE 2.11 MG/DL, 2.02 MG/DL, AND 2.07 MG/DL. THE REPEATED RESULTS WERE BELIEVED TO BE CORRECT. THE SAMPLE WAS REPEATED BECAUSE THE CUSTOMER HAS HAD ONGOING ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21917475 ↗
  • MalfunctionJul 15, 2024High Test Results; High Test Results

    THE INITIAL REPORTER COMPLAINED OF QC ISSUES AND QUESTIONED HIGH RESULTS FOR 35 PATIENT SAMPLES TESTED FOR MAGNESIUM (MG2) ON A COBAS 8000 C 702 MODULE. THE CUSTOMER REMOVED THE ONBOARD REAGENT AND RECALIBRATED USING THE STANDBY REAGENT. THE FOLLOWING ARE EXAMPLES OF DISCREPANT RESULTS FOR 2 PATIENT SAMPLES. PATIENT 1 INITIAL RESULT WAS 3.6 MG/DL. THE REPEAT RESULT WAS 2.1 MG/DL. PATIENT 2 INITIAL RESULT WAS 3.4 MG/DL. THE REPEAT RESULT WAS 2.0 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19745564 ↗
  • MalfunctionNov 7, 2023Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE MG2 MAGNESIUM GEN.2 RESULTS FROM FIVE PATIENT SAMPLES TESTED ON THE COBAS 8000 C702 MODULE. IT IS UNKNOWN IF THE INITIAL RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER SUSPECTS IMPRECISION. THE REPORTER WAS ABLE TO PROVIDE THE FOLLOWING EXAMPLES OF DISCREPANT RESULTS: (B)(6). THE PATIENT SAMPLE WAS RUN ON LINE 2 OF THE MODULE. THE INITIAL RESULT WAS 0.68 MMOL/L. THE FIRST REPEAT RESULT WAS 0.47 MMOL/L. THE SECOND REPEAT RESULT WAS 0.42 MMOL/L. (B)(6). THE PATIENT SAMPLE WAS RUN ON LINE 2 OF THE MODULE. THE INITIAL RESULT WAS 1.15 MMOL/L. THE FIRST REPEAT RESULT WAS 0.79 MMOL/L. THE SECOND REPEAT RESULT WAS 0.81 MMOL/L. (B)(6). THE INITIAL RESULT FROM LINE 2 WAS 0.28 MMOL/L. THE FIRST REPEAT RESULT FROM LINE 2 WAS 0.55 MMOL/L. THE SECOND REPEAT RESULT FROM LINE 1 WAS 0.31 MMOL/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18085267 ↗
  • MalfunctionJun 28, 2022Low Test Results; Low Test Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE MG2 MAGNESIUM GEN.2 RESULTS FOR NINE PATIENTS TESTED ON A COBAS 8000 C 702 MODULE SERIAL NUMBER (B)(4). PRIOR TO PATIENT TESTING, THE CUSTOMER CONFIRMED CALIBRATION AND QC WERE WITHIN THE ACCEPTABLE RANGES. THE INITIAL MAGNESIUM RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER PERFORMED QC AND HAD RESULTS OUT OF THE ACCEPTABLE RANGES. THE CUSTOMER ATTEMPTED TO CALIBRATE THE REAGENT BUT HAD FAILING RESULTS. THE CUSTOMER REMOVED THE REAGENT PACK AND PERFORMED REPEAT TESTING WITH THE SAMPLES ON A DIFFERENT COBAS 8000 C 702 MODULE. PATIENT 1'S INITIAL MAGNESIUM RESULT WAS 0.8 MG/DL. THE PATIENT'S REPEAT MAGNESIUM RESULT WAS 2.0 MG/DL. PATIENT 2'S INITIAL MAGNESIUM RESULT WAS 0.7 MG/DL. THE PATIENT'S REPEAT MAGNESIUM RESULT WAS 1.8 MG/DL. PATIENT 3'S INITIAL MAGNESIUM RESULT WAS 0.8 MG/DL. THE PATIENT'S REPEAT MAGNESIUM RESULT WAS 2.1 MG/DL. PATIENT 4'S INITIAL MAGNESIUM RESULT WAS 1.2 MG/DL. THE PATIENT'S REPEAT MAGNESIUM RESULT WAS 3.2 MG/DL. PATIENT 5'S INITIAL MAGNESIUM RESULT WAS 0.8 MG/DL. THE PATIENT'S REPEAT MAGNESIUM RESULT WAS 2.0 MG/DL. PATIENT 6'S INITIAL MAGNESIUM RESULT WAS 0.8 MG/DL. THE PATIENT'S REPEAT MAGNESIUM RESUL

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14844059 ↗

Other PHOTOMETRIC METHOD, MAGNESIUM devices

All PHOTOMETRIC METHOD, MAGNESIUM devices →