Skip to content

Elecsys ?-CrossLaps/serum 07026960190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04015630940141), received between May 14, 2024 and May 14, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2024Incorrect Measurement; Incorrect Measurement

    THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS B-CROSSLAPS/SERUM (B-CROSSLAPS) RESULTS FROM ONE PATIENT SAMPLE TESTED ON THE COBAS 8000 CORE UNIT. ON (B)(6) 2024, THE INITIAL RESULT FROM THE ANALYZER WAS <10 PG/ML. THE FIRST REPEAT RESULT WAS CONSISTENT. THE REPORTER NOTED THAT THE PATIENT HAS OSTEOPOROSIS AND SECONDARY HYPERPARATHYROIDISM AND THAT HER OTHER BONE MARKERS (N-MID AND TP1NP) WERE BOTH WITHIN THE REFERENCE RANGE. THE REPORTER QUESTIONED THE INITIAL RESULT AND SENT THE PATIENT SAMPLE FOR FURTHER TESTING USING A DOMESTIC PLATFORM WITH CHEMILUMINESCENCE IMMUNOASSAY (CLIA) AS THE LABORATORY METHOD. ON (B)(6) 2024, THE SECOND REPEAT RESULT FROM THE DOMESTIC PLATFORM USING THE CLIA METHOD WAS 34 PG/ML (REFERENCE RANGE <1017 PG/ML).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19312618 ↗

Other COLUMN CHROMATOGRAPHY & COLOR DEVELOPMENT, HYDROXYPROLINE devices

All COLUMN CHROMATOGRAPHY & COLOR DEVELOPMENT, HYDROXYPROLINE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page