Elecsys Beta-CrossLaps/serum 09005781190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336171561), received between Jun 14, 2023 and Jun 23, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Low Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS B-CROSSLAPS ASSAY ON A COBAS E 801 ANALYTICAL UNIT. INITIAL RESULT: >6 NG/ML (REFERENCE RANGE 0.21-0.44 NG/ML). THE INITIAL RESULT DID NOT MATCH THE PATIENT'S CLINICAL DIAGNOSIS, AND THE SAMPLE WAS MANUALLY DILUTED WITH A LOW-CONCENTRATION SERUM AND REPEATED. THE REPEAT RESULTS WERE AS FOLLOWS: 1ST REPEAT RESULT: >6 NG/ML (TESTED WITH A DILUTION FACTOR OF 1:2). 2ND REPEAT RESULT: 1.94 NG/ML (TESTED WITH A DILUTION FACTOR OF 1:4). 3RD REPEAT RESULT: 0.872 NG/ML (TESTED WITH A DILUTION FACTOR OF 1:8). THE CUSTOMER SUSPECTED AN INTERFERENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22309715 ↗
  • MalfunctionJun 14, 2023Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS -CROSSLAPS/SERUM IMMUNOASSAY RESULTS FOR MULTIPLE PATIENT SAMPLES ON A COBAS E 801 ANALYTICAL UNIT. THE CUSTOMER PROVIDED AN EXAMPLE OF DISCREPANT RESULTS FOR 4 PATIENT SAMPLES. ON (B)(6) 2023, PATIENT 1 HAD AN INITIAL -CROSSLAPS RESULT OF 70.4 PG/ML. THE REPEAT RESULT WAS 521 PG/ML. ON (B)(6) 2023, PATIENT 2 HAD AN INITIAL -CROSSLAPS RESULT OF 63.3 PG/ML. THE REPEAT RESULT WAS 529 PG/ML. ON (B)(6) 2023, PATIENT 3 HAD AN INITIAL -CROSSLAPS RESULT OF 60.6 PG/ML. THE REPEAT RESULT WAS 548 PG/ML. ON (B)(6) 2023, PATIENT 4 HAD AN INITIAL -CROSSLAPS RESULT OF 60.1 PG/ML. THE REPEAT RESULT WAS 624 PG/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17129878 ↗

Other COLUMN CHROMATOGRAPHY & COLOR DEVELOPMENT, HYDROXYPROLINE devices

All COLUMN CHROMATOGRAPHY & COLOR DEVELOPMENT, HYDROXYPROLINE devices →