Quick summary

This B. Braun non-vented dispensing pin is used to transfer IV fluids. Most of the 63 reports describe contamination concerns: users mentioning white particles, matter, or tiny flakes found in the pin cap, on its tips, or left behind in mixing bags and final drug product, plus a few leaks. There are no deaths and no injuries. No recalls name this device. Reports rose from 5 in the earlier 12 months to 14 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
63
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 04022495769756), received between Dec 21, 2016 and May 20, 2026 — classified as 63 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (58 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 5 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 5 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 10 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem58
  • Fluid/Blood Leak2
  • Device Contamination with Chemical or Other Material1
  • Device-Device Incompatibility1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • No Known Impact Or Consequence To Patient25
  • No Consequences Or Impact To Patient9
  • Hemorrhage/Blood Loss/Bleeding1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2026Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: A SMALL WHITE PARTICLE WAS OBSERVED IN ONE OF OUR FINAL DRUG PRODUCT BAGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25231193 ↗
  • MalfunctionMay 3, 2026Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: THE DISPENSING PIN IS LEAVING A PARTICULATE IN THE MIXING BAG. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25062849 ↗
  • MalfunctionMay 1, 2026Leak/Splash

    AS REPORTED BY THE USER FACILITY: DURING THE PREPARATION OF UPSTREAM PROCESS, A SIGNIFICANT LEAK WAS OBSERVED OUTSIDE THE CLASS D BIOSAFETY CABINET. THE SAMPLING SYSTEM SEPARATED INTO TWO PIECES. THERE WAS A SUBSTANTIAL LOSS OF PRODUCT AND CELLS, AND A HIGH RISK OF CONTAMINATION.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 25058228 ↗
  • MalfunctionFeb 4, 2026Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: DISPENSING PIN HAD PARTICULATE MATTER IN THE CAP OF THE DISPENSING PIN. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24258737 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 9: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563871 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 8: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563870 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 10: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563869 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 6: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563868 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 7: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563866 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 5: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563863 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 1: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563862 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 3: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563861 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 4: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563860 ↗
  • MalfunctionNov 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: EVENT 2: END USER OBSERVED TINY FLAKES ON MULTIPLE TIPS OF THE NON-VENTED DISPENSING PINS. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23563859 ↗
  • MalfunctionMay 7, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: END USER REPORTING DISPENSING PINS HAVING PLASTIC PARTICULATE IN THEM. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21975671 ↗
  • MalfunctionMay 5, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: WE CONTINUE TO SEE EXCESSIVE PLASTIC IN USE OF VARIOUS LOT #S FOR THE SPIKE 413501.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21961624 ↗
  • MalfunctionMay 5, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: WE CONTINUE TO SEE EXCESSIVE PLASTIC IN USE OF VARIOUS LOT #S FOR THE SPIKE 413501.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21961623 ↗
  • MalfunctionMay 1, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: WE CONTINUE TO SEE EXCESSIVE PLASTIC IN USE OF VARIOUS LOT #S FOR THE SPIKE 413501.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21942484 ↗
  • MalfunctionMay 1, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: WE CONTINUE TO SEE EXCESSIVE PLASTIC IN USE OF VARIOUS LOT #S FOR THE SPIKE 413501.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21941908 ↗
  • MalfunctionOct 21, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY, WE HAVE A "NON-VENTED DISPENSING PIN" REF (B)(4) THAT LOOKS LIKE IT HAS MOLD OR SOMETHING SIMILAR ON THE TIP OF THE DISPENSING PIN UNDER THE CAP. WAS SOME BLACK, GREY DARK MATERIAL AT THE END OF THE SPIKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20493070 ↗
  • MalfunctionOct 17, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY, A LITTLE PIECE OF THE SPIKE, LEAVES A LITTLE PIECE OF PLASTIC IN THE SOLUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20474620 ↗
  • MalfunctionOct 11, 2024Device-Device Incompatibility; Device-Device Incompatibility

    AS REPORTED WE CONTINUE TO HAVE CORING ISSUES WITH THE B. BRAUN NON VENTED DISPENSING PIN # 413501 - WE HAVE HAD THE ISSUE WITH MULTIPLE LOT#'S.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20428913 ↗
  • MalfunctionOct 10, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY USER FACILITY, A LITTLE PIECE OF THE SPIKE, LEAVES A LITTLE PIECE OF PLASTIC IN THE SOLUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20417729 ↗
  • MalfunctionOct 2, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED, WE CONTINUE TO HAVE CORING ISSUES WITH THE B. BRAUN NON VENTED DISPENSING PIN # 413501 - WE HAVE HAD THE ISSUE WITH MULTIPLE LOT#'S. THIS IS FOR LOT # 0061932546.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20352272 ↗
  • MalfunctionApr 19, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    AS REPORTED BY THE USER FACILITY: "WE FOUND A FIBER INSIDE THE TIP OF THE PIN ON PRODUCT 413501. IT WAS REMOVED FROM THE PACKAGE BUT NOT YET USED." NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14140373 ↗

Other Set, i.V. Fluid transfer devices