Baxter Healthcare Corporation · EXACTAMIX Empty EVA Bag · Class 2

All models of EXACTAMIX (28)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Baxter ExactaMix is an empty plastic bag used to transfer intravenous fluids. Of the 603 reports naming it, most describe fluid leaking from the bag, and some describe contamination with chemicals or other material. The reports describe no patient involvement, no injuries and no deaths. One recall, now closed, started in September 2019 and concerned a possible leak after the bag is used for compounding.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
597
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 597 adverse event reports that list this exact device (DI 00085412477213), received between Dec 29, 2017 and Sep 15, 2026 — classified as 597 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (421 reports).

1 FDA recall record name this device's identifier.

36 other recalls cover devices cleared under the same FDA submission (K900585); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 14 reportsJan 2024: 4 reportsFeb 2024: 4 reportsMar 2024: 7 reportsApr 2024: 1 reportsMay 2024: 5 reportsJun 2024: 5 reportsJul 2024: 2 reportsAug 2024: 8 reportsSep 2024: 2 reportsOct 2024: 4 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 5 reportsFeb 2025: 5 reportsMar 2025: 7 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 4 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 4 reportsOct 2025: 3 reportsNov 2025: 1 reportsDec 2025: 6 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 18, 2019Class IITerminated

    Potential ability to leak once used for compounding.

    Z-0167-2020 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak421
  • Device Contamination with Chemical or Other Material153
  • Break10
  • Disconnection5
  • Difficult to Open or Close4
  • Material Perforation4
  • Material Puncture/Hole4
  • Detachment of Device or Device Component2

Patient problems

  • No Patient Involvement282
  • No Clinical Signs, Symptoms or Conditions273
  • No Consequences Or Impact To Patient40
  • No Known Impact Or Consequence To Patient2
  • Premature Labor1

Recent adverse event reports

The 25 most recent of 597 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50781560 ↗
  • MalfunctionMay 26, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT THREE (3) 250ML EXACTAMIX EMPTY EVA (ETHYLENE VINYL ACETATE) BAGS LEAKED AT THE MEMBRANE PORT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25289545 ↗
  • MalfunctionMay 6, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT SEVEN (07) EXACTAMIX 250 ML EVA TPN BAGS LEAKED FROM AN UNSPECIFIED LOCATION. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096608 ↗
  • MalfunctionMay 6, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA TPN BAG LEAKED FROM AN UNSPECIFIED LOCATION. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096607 ↗
  • MalfunctionMar 30, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT FOUR (4) 250ML EXACTAMIX EVA (ETHYLENE VINYL ACETATE) BAGS WERE LEAKING AT THE ADMINISTRATION PORT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24717376 ↗
  • MalfunctionMar 25, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) 250ML EXACTAMIX EVA (ETHYLENE VINYL ACETATE) BAGS WERE LEAKING FROM THE ADMINISTRATION PORT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691812 ↗
  • MalfunctionMar 9, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX® 250 ML EVA TPN BAG LEAKED AT THE MEMBRANE PORT. THE ISSUE WAS NOTICED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24553515 ↗
  • MalfunctionFeb 24, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT THREE (3) EXACTAMIX 250 ML EVA (ETHYLENE VINYL ACETATE) TPN (TOTAL PARENTERAL NUTRITION) BAGS LEAKED AT THE SPIKE PORT. THIS WAS OBSERVED DURING SETUP PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434204 ↗
  • MalfunctionFeb 17, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) 250ML EXACTAMIX EVA (ETHYLENE VINYL ACETATE) BAGS WERE LEAKING FROM THE MEMBRANE PORT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24368697 ↗
  • MalfunctionFeb 13, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) EXACTAMIX 250 ML EVA TPN BAGS LEAKED AT THE ADMINISTRATION PORT. THE ISSUE OCCURRED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24332133 ↗
  • MalfunctionFeb 11, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT A 250ML EXACTAMIX EVA (ETHYLENE VINYL ACETATE) BAG WAS LEAKING FROM THE ADMINISTRATION PORT AREA. THE LEAK WAS IDENTIFIED DURING MIXING PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24318003 ↗
  • MalfunctionJan 27, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT A 250ML EXACTAMIX EVA BAG WAS LEAKING FROM THE ADMINISTRATION PORT. THE LEAK WAS OBSERVED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24195908 ↗
  • MalfunctionJan 13, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA (ETHYLENE-VINYL ACETATE) TPN (TOTAL PARENTERAL NUTRITION) BAG LEAKED FROM THE ADMIN PORT. THIS WAS OBSERVED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24058865 ↗
  • MalfunctionJan 9, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA (ETHYLENE-VINYL ACETATE) TPN (TOTAL PARENTERAL NUTRITION) BAG LEAKED FROM AN UNSPECIFIED LOCATION. THIS WAS OBSERVED DURING SETUP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24030447 ↗
  • MalfunctionJan 2, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) 250ML EVA (ETHYLENE-VINYL ACETATE) TPN (TOTAL PARENTERAL NUTRITION) BAGS LEAKED THROUGH THE MIDDLE PORT. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23943099 ↗
  • MalfunctionDec 29, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA TPN BAG LEAKED FROM AN UNSPECIFIED LOCATION. THE ISSUE OCCURRED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23904605 ↗
  • MalfunctionDec 29, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT FIVE (5) EXACTAMIX 250 ML EVA TPN BAGS LEAKED FROM UNSPECIFIED LOCATION. THE ISSUE OCCURRED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23904600 ↗
  • MalfunctionDec 16, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (02) EXACTAMIX 250 ML EVA TPN BAGS LEAKED FROM THE TUBING. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23816410 ↗
  • MalfunctionDec 4, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA (ETHYLENE VINYL ACETATE) TPN (TOTAL PARENTERAL NUTRITION) BAG LEAKED AT THE MEMBRANE PORT. THIS WAS OBSERVED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23721284 ↗
  • MalfunctionDec 4, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA TPN BAG HAD PARTICULATE MATTER OR FOREIGN MATTER. THE BAG CONTAINED 125 ML OF 0.9% SODIUM CHLORIDE. THIS ISSUE WAS DISCOVERED DURING SETUP AND PREPARATION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23716087 ↗
  • MalfunctionDec 4, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 250ML EXACTAMIX EVA (ETHYLENE VINYL ACETATE) BAG WAS LEAKING FROM THE MEMBRANE PORT. THE LEAK WAS OBSERVED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23716086 ↗
  • MalfunctionNov 5, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA (ETHYLENE-VINYL ACETATE) TPN (TOTAL PARENTERAL NUTRITION) BAG LEAKED FROM AN UNSPECIFIED LOCATION. THIS WAS OBSERVED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23478736 ↗
  • MalfunctionOct 31, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA (ETHYLENE-VINYL ACETATE) TPN (TOTAL PARENTERAL NUTRITION) BAGS LEAKED FROM THE PORT. THIS WAS OBSERVED DURING SETUP PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441535 ↗
  • MalfunctionOct 30, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT EIGHT (08) EXACTAMIX 250 ML EVA TPN BAGS HAD HAIR ON THE PORT. THIS WAS OBSERVED AFTER FILLING. HOWEVER, IT WAS UNSPECIFIED IF PATIENT WAS INVOLVED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23422845 ↗
  • MalfunctionOct 29, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN EXACTAMIX 250 ML EVA TPN BAG HAD HAIR AT THE PORT. THIS ISSUE WAS NOTICED AFTER FILLING. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23416738 ↗

FDA clearance and classification

Cleared through 510(k) K900585 on Apr 18, 1990 (decision: Substantially Equivalent). Submitted by Baxa Corp., Sub. of Cook Group, Inc. as “Tubing Sets For Baxa Exacta-Med Pharmacy Pump”. The FDA review took 70 days. FDA record ↗

The FDA regulates product code LHI (set, i.v. fluid transfer) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K900585. One submission often covers many models, so check the product and lot details in each record.

  • Z-0364-2019 · Sep 12, 2018 · Terminated
    EXACTAMIX Empty EVA Bags - 500 mL - Product Usage: A plastic container used to hold large-volume sterile solutions to be administered to a patient through an intravascular administ
  • Z-0365-2019 · Sep 12, 2018 · Terminated
    EXACTAMIX Empty EVA Bags - 1000 mL Product Usage: A plastic container used to hold large-volume sterile solutions to be administered to a patient through an intravascular administr
  • Z-0367-2019 · Sep 12, 2018 · Terminated
    EXACTAMIX Empty EVA Bags - 3000 mL Product Usage: A plastic container used to hold large-volume sterile solutions to be administered to a patient through an intravascular administr
  • Z-0366-2019 · Sep 12, 2018 · Terminated
    EXACTAMIX Empty EVA Bags - 2000 mL Product Usage: A plastic container used to hold large-volume sterile solutions to be administered to a patient through an intravascular administr
  • Z-0363-2019 · Sep 12, 2018 · Terminated
    EXACTAMIX Empty EVA Bags - 250 mL Product Usage: A plastic container used to hold large-volume sterile solutions to be administered to a patient through an intravascular administra
  • Z-0007-2017 · Aug 24, 2016 · Terminated
    0.2 Micron Filter, 50 mm Product Usage: The 0.2 Micron Filter is a bacteria and particulate filter for aqueous solutions. It attaches to a standard Baxter Pharmacy Pump Tube Set.
  • Z-2165-2013 · Aug 22, 2013 · Terminated
    Baxa Vial Adapter, 14mm - Luer Lock, Order No. REF: 90, Quantity: 50 each, Single Use. Product Code H93890. Used as a needleless access device for reconstitution and drug transfer
  • Z-1962-2011 · Mar 8, 2011 · Terminated
    ExactaMix EVA Container, Calibration Bag, REF: 735, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to hold large-volume sterile solutions to be adminis
  • Z-1965-2011 · Mar 8, 2011 · Terminated
    ExactaMix EVA Container, 1000mL, REF: 739, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to hold large-volume sterile solutions to be administered to
  • Z-1963-2011 · Mar 8, 2011 · Terminated
    ExactaMix EVA Container, 250mL, REF: 737, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to hold large-volume sterile solutions to be administered to
  • Z-1964-2011 · Mar 8, 2011 · Terminated
    ExactaMix EVA Container, 500mL, REF: 738, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to hold large-volume sterile solutions to be administered to
  • Z-1967-2011 · Mar 8, 2011 · Terminated
    ExactaMix EVA Container, 3000mL, REF: 741, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to hold large-volume sterile solutions to be administered to
  • Z-1966-2011 · Mar 8, 2011 · Terminated
    ExactaMix EVA Container, 2000mL, REF: 740, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to hold large-volume sterile solutions to be administered to
  • Z-2147-2010 · May 26, 2010 · Terminated
    Exacta-Mix Valve Set, Six Inlet, Non-Vented, Assembled, Tubing Kit w/Calibration Bag, Sterile R, non-pyrogenic fluid path, 250 mL, Order No. REF 706, Baxa Corporation.
  • Z-2140-2010 · May 26, 2010 · Terminated
    Exacta-Mix EVA Container, Sterile R, 2000 mL, Order No. REF: 140, Baxa Corporation.

…and 21 more.

Common questions about the EXACTAMIX H938737

Has the EXACTAMIX H938737 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0167-2020 (Class II), began on Sep 18, 2019: Potential ability to leak once used for compounding. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the EXACTAMIX H938737?

The FDA's MAUDE database lists 597 adverse event reports for this device, received between Dec 29, 2017 and Sep 15, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (421), “Device Contamination with Chemical or Other Material” (153) and “Break” (10); the most common patient problem is “Premature Labor” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EXACTAMIX H938737?

None of the reports are classified as deaths, 0 as injuries and 597 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the EXACTAMIX H938737 FDA approved?

It was cleared through 510(k) K900585 on Apr 18, 1990. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the EXACTAMIX H938737, and is it still sold?

The FDA's device database lists Baxter Healthcare Corporation as the company responsible for it (brand EXACTAMIX, model H938737). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Set, i.V. Fluid transfer devices