Skip to content

Extension Set 471960Recalls & reported problems

B. Braun Medical Inc. · Small bore extension set with no injection site · Class 2

All models of Extension Set (41)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04022495770127), received between Dec 14, 2018 and Mar 24, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash2
  • Adverse Event Without Identified Device or Use Problem1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1
  • Pressure Sore/Ulcer1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2025Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: LEAKING EXT SET. DETAILED INQUIRY DESCRIPTION: WHEN FLUSHED OR INFUSING, LEAKS WHERE CONNECTED TO T-CONNECTOR. UPON GIVING BEDSIDE REPORT ON (B)(6) 2025 AT 07:30 IT WAS NOTED THAT THERE WAS A SMALL DAMP SPOT ON PATIENT BLANKET. A SMALL AMOUNT OF BLOOD WAS NOTED IN THE T-CONNECTOR. BLANKET WAS NOT STICKY, BUT DECISION MADE BY NIGHT AND DAY SHIFT NURSE TO FLUSH IV. PATIENT DID NOT APPEAR TO HAVE ANY OTHER SOURCE FOR WETNESS AT THE TIME (WRONG LOCATION FOR DROOL OR EMESIS OR OG TUBE ASPIRATE). UPON FLUSHING IV, FLUID SPURTED OUT OF A TINY CRACK IN THE GREEN HUB COLORED T-CONNECTOR. IV WAS TAKEN DOWN TO THE YELLOW HUB. ASEPTIC TECHNIQUE USED. NEW T-CONNECTOR ATTACHED. IV SECURED WITH NEW TEGADERM AND TAPE. PIV FLUSHED WITHOUT DIFFICULTY. PIV INFUSION OF D10% CONTINUED WITHOUT DELAY. IMMEDIATELY PLACED FAULTY T-CONNECTOR IN TO A BAG WITH FLUSH ATTACHED. PIV HAD BEEN PLACED THE DAY BEFORE AND PACKAGING FOR T-CONNECTOR WAS UNABLE TO BE LOCATED. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678997 ↗
  • MalfunctionApr 11, 2023Leak/Splash; Leak/Splash

    AS REPORTED BY THE USER FACILITY: DESCRIPTION: SET IS LEAKING BLOOD. NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16712258 ↗
  • InjuryApr 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS PER MEDWATCH (B)(4): EVENT DESCRIPTION: PERIPHERAL INTRAVENOUS (PIV) HUB WAS TAPED TO PATIENT PER STANDARDS OF CARE. AN UNSTAGEABLE INJURY WAS NOTED UNDER THE GAUZE WHICH WAS PADDING THE IV HUB WHICH WAS CAUSED BY THE FLANGES ON THE PIV HUB.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 14140375 ↗
  • MalfunctionDec 14, 2018Fluid/Blood Leak; Fluid/Blood Leak

    AS REPORTED BY THE USER FACILITY: THE EXTENSION SET WAS CRACKED AT THE FEMALE LUER WHICH CAUSED LEAKAGE. PACK RED BLOOD CELLS (PRBC) WERE INFUSING WHEN THE LEAK WAS OBSERVED. NO INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8166563 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page