Fusion Nail 15-0028/08Recalls & reported problems
WALDEMAR LINK GmbH & Co. KG · Fusion Nail, modular · Class 2 · Implant
All models of Fusion Nail (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04026575175222), received between Oct 30, 2025 and Oct 30, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Component Misassembled” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
15 other recalls cover devices cleared under the same FDA submission (K151008); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 10, 2025Class IIOpen, Classified
The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.
Z-0890-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Component Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2025Component Misassembled; Component Misassembled
"NOTIFIED THAT DURING A FUSION NAIL CASE, THE RESIDENT WAS UNABLE TO ADVANCE THE SET SCREW. ANOTHER BOX WAS OPENED AND USED WITH NO ISSUE. THE REPRESENTATIVE STATED THAT WHILE THE CEMENT WAS HARDENING IN A FUSION NAIL CASE, THE MORSE TAPER WAS IMPACTED. THEN THEY WENT TO ENGAGE THE SET SCREW BUT THEY WERE NOT ABLE TO ADVANCE IT. IN ORDER TO SAVE TIME, A NEW IMPLANT WAS OPENED AND USED WITHOUT ISSUE. THE DELAY WAS LESS THAN 4 MINUTES. UPON INVESTIGATING THE WASTED IMPLANT LATER, THE REPRESENTATIVE FOUND THE SCREW IN THE MORSE TAPER WAS ASSEMBLED BACKWARDS. THE PART WILL BE RETURNED FOR THE INVESTIGATION." [CUSTOMER].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23424135 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K151008. One submission often covers many models, so check the product and lot details in each record.
- Z-1523-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/16;
- Z-1516-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-2836/11;
- Z-1522-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/12;
- Z-1512-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/15;
- Z-1514-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/07;
- Z-1513-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;
- Z-1520-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/15;
- Z-1515-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-2835/12;
- Z-1519-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/11;
- Z-1521-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/11;
- Z-1518-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/09;
- Z-1511-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/05;
- Z-1517-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8030/12;
- Z-2410-2024 · May 22, 2024 · Open, ClassifiedEndo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02, 15-2814/03, 15-2814/04, 15-2815/11, 15-2815/12, 15-2816/11, 15-2816/12, 15-281
- Z-2411-2024 · May 22, 2024 · Open, ClassifiedEndo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75m
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented devices
- LFIT 6260-9-136Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented166 reports
- Trident 623-00-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented112 reports
- Accolade 6021-0435Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented88 reports
- EXETER 0580-1-044Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall22 reports
- Endo-Model SL Knee Prosthesis System 16-2840/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall17 reports
- Endo-Model SL Knee Prosthesis System 16-2840/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall12 reports
- Trident 623-10-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented57 reports
- LFIT 6260-9-336Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page