Skip to content

Fusion Nail 15-0028/08Recalls & reported problems

WALDEMAR LINK GmbH & Co. KG · Fusion Nail, modular · Class 2 · Implant

All models of Fusion Nail (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04026575175222), received between Oct 30, 2025 and Oct 30, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Component Misassembled” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

15 other recalls cover devices cleared under the same FDA submission (K151008); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 10, 2025Class IIOpen, Classified

    The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.

    Z-0890-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Component Misassembled1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2025Component Misassembled; Component Misassembled

    "NOTIFIED THAT DURING A FUSION NAIL CASE, THE RESIDENT WAS UNABLE TO ADVANCE THE SET SCREW. ANOTHER BOX WAS OPENED AND USED WITH NO ISSUE. THE REPRESENTATIVE STATED THAT WHILE THE CEMENT WAS HARDENING IN A FUSION NAIL CASE, THE MORSE TAPER WAS IMPACTED. THEN THEY WENT TO ENGAGE THE SET SCREW BUT THEY WERE NOT ABLE TO ADVANCE IT. IN ORDER TO SAVE TIME, A NEW IMPLANT WAS OPENED AND USED WITHOUT ISSUE. THE DELAY WAS LESS THAN 4 MINUTES. UPON INVESTIGATING THE WASTED IMPLANT LATER, THE REPRESENTATIVE FOUND THE SCREW IN THE MORSE TAPER WAS ASSEMBLED BACKWARDS. THE PART WILL BE RETURNED FOR THE INVESTIGATION." [CUSTOMER].

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23424135 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K151008. One submission often covers many models, so check the product and lot details in each record.

  • Z-1523-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-0027/16;
  • Z-1516-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-2836/11;
  • Z-1522-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-0027/12;
  • Z-1512-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-0027/15;
  • Z-1514-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/07;
  • Z-1513-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;
  • Z-1520-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-8521/15;
  • Z-1515-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-2835/12;
  • Z-1519-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-8521/11;
  • Z-1521-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-0027/11;
  • Z-1518-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-8521/09;
  • Z-1511-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/05;
  • Z-1517-2026 · Jan 12, 2026 · Open, Classified
    Endo-Model Replacement Plateau; Item Number: 15-8030/12;
  • Z-2410-2024 · May 22, 2024 · Open, Classified
    Endo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02, 15-2814/03, 15-2814/04, 15-2815/11, 15-2815/12, 15-2816/11, 15-2816/12, 15-281
  • Z-2411-2024 · May 22, 2024 · Open, Classified
    Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75m

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page