Endo-Model SL Knee Prosthesis System 16-2840/02Recalls & reported problems
WALDEMAR LINK GmbH & Co. KG · Connection Components,Rotational version, neutral · Class 2 · Implant
All models of Endo-Model SL Knee Prosthesis System (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04026575359233), received between Nov 10, 2022 and Jun 24, 2026 — classified as 12 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
15 other recalls cover devices cleared under the same FDA submission (K151008); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 12, 2026Class IIOpen, Classified
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.
Z-1513-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component11
- Activation Failure3
- Fracture3
- Crack1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Naturally Worn1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Failure of Implant5
- Insufficient Information4
- Joint Laxity2
- Fall1
- Pain1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
InjuryJun 24, 2026Fracture
BREAKAGE OF THE CONNECTION COMPONENT [CUSTOMER].
Patient problems: Insufficient Information
MAUDE report 25606416 ↗MalfunctionJan 13, 2026Crack; Fracture
DURING IMPLANTATION IT WAS NOTICED THAT THE IMPLANT HAD BEEN DELIVERED WITH A PLATEAU SCREW WHOSE BUSHING HAD DETACHED FROM THE SCREW SHAFT DUE TO A LONGITUDINAL CRACK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24050694 ↗InjurySep 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
UNKNOWN EVENT DATE. NOTIFICATION FROM REPRESENTATIVE M. D. THAT THE DOCTOR WILL BE REVISING A CONPONENT PREVIOUSLY IMPLANTED ON (B)(6) 2023, BECAUSE OF DISASSOCIATION. HE WILL ONLY BE REPLACING THE HINGE AND THE BUSHINGS. REVISION WILL TAKE PLACE ON (B)(6) 2025. [CUSTOMER]
Patient problems: Insufficient Information
MAUDE report 22932054 ↗InjuryApr 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
PATIENT CAME IN FOR A CHECK-UP WITH PAIN. UNSTABLE KNEE JOINT. [CUSTOMER].
Patient problems: Pain
MAUDE report 21924098 ↗InjuryApr 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
DISASSEMBLY FAILURE OF THE IMPLANT HINGE. [CUSTOMER].
Patient problems: Insufficient Information
MAUDE report 21823493 ↗InjurySep 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NOTIFIED BY (B)(6) ON (B)(6) 2024, THAT A REVISION OF A DISASSOCIATED CONNECTION COMPONENT ORIGINALLY IMPLANTED ON (B)(6) 2018, WAS PERFORMED BY DR. (B)(6). ON (B)(6) 2024. IMAGES INCLUDED, ONE SHOWING EXPLANTED COMPONENTS WITH SET SCREW THAT IS PROUD AND NOTIFIED THAT THIS IS HOW IT LOOKED UPON REMOVAL. REVISION SURGERY WAS SUCCESSFUL. [CUSTOMER].
Patient problems: Failure of Implant
MAUDE report 20142862 ↗InjurySep 4, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NOTIFIED ON 2024-08-14, THAT A REVISION OF A DISASSOCIATED CONNECTION COMPONENT ORIGINALLY IMPLANTED ON (B)(6) 2018, WAS PERFORMED BY DR. (B)(6) ON (B)(6) 2024.
Patient problems: Joint Laxity
MAUDE report 20140136 ↗InjurySep 4, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A REVISION DUE TO DISLOCATION AT THE CONNECTION COMPONENT OCCURRED ON (B)(6) 2024. ORIGINALLY IMPLANTED ON (B)(6) 2023.
Patient problems: Joint Laxity
MAUDE report 20140043 ↗InjuryAug 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
A REVISION DUE TO DISLOCATION AT THE CONNECTION COMPONENT OCCURRED ON (B)(6)2024. ORIGINALLY IMPLANTED ON (B)(6) 2023. [CUSTOMER].
Patient problems: Failure of Implant
MAUDE report 20105139 ↗InjuryJan 19, 2024Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
NOTIFIED ON 12/19/23 BY (B)(6) REGARDING A REVISION CASE WITH DR. (B)(6) ON (B)(6) 2023 FOR WHAT WAS THOUGHT TO BE A FAILED CONNECTION COMPONENT. AT SURGERY, IT WAS FOUND THAT THE CONNECTION COMPONENT WAS INTACT BUT THE PLASTIC BEARINGS WERE BROKEN IN THE FEMUR. THE REVISION REPLACED THE CONNECTION COMPONENT AND BEARINGS AND THE OUTCOME WAS SUCCESSFUL. INITIAL SURGERY DATE WAS (B)(6) 2022 WITH SAME SURGEON. EXPLANTED DEVICES HAVE BEEN DISCARDED. [CUSTOMER].
Patient problems: Failure of Implant
MAUDE report 18542327 ↗InjuryDec 12, 2023Detachment of Device or Device Component; Loosening of Implant Not Related to Bone-Ingrowth; Detachment of Device or Device Component; Loosening of Implant Not Related to Bone-Ingrowth
AS REPORTED BY REPRESENTATIVE T.A ON (B)(6) 2023 , A PATIENT PRESENTED TO THE EMERGENCY ROOM ON WEDNESDAY ON (B)(6) 2023 AND FOUND TO HAVE A DISLOCATED ENDO SL COMPONENT. REVISION SURGERY TOOK PLACE ON (B)(6) 2023 AND WAS VERY SUCCESSFUL. REPRESENTATIVE STATED THAT THE EXPLANTED BUSHINGS WERE "DESTROYED".
Patient problems: Failure of Implant
MAUDE report 18309046 ↗InjuryOct 18, 2023Detachment of Device or Device Component; Naturally Worn; Detachment of Device or Device Component; Naturally Worn
WEAR OF THE BEARING SHELL IN THE FEMORAL COMPONENT.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 17957029 ↗MalfunctionAug 29, 2023Positioning Failure; Positioning Failure
SET SCREW OF THE CONNECTION COMPONENT COULD NOT BE SCREWED IN. THE COUPLING HAS DECOUPLED AND OPENED. THE SET SCREW COULD NOT BE UNSCREWED FROM THE DEVICE. [TRANSLATED DESCRIPTION FROM CUSTOMER].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17643282 ↗MalfunctionAug 29, 2023Activation Failure; Activation Failure
CONNECTION COMPONENT AXIS DID NOT PROPERLY DEPLOY DURING SURGERY. THE AXIS AND THE PE BEARINGS WERE NOT COMPATIBLE. BOTH THE CONNECTOR AND THE BEARINGS HAVE BEEN EXCHANGED. [TRANSLATED DESCRIPTION FROM FIELD SERVICE].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17641712 ↗MalfunctionAug 29, 2023Activation Failure; Activation Failure
CONNECTION COMPONENT AXIS DID NOT PROPERLY DEPLOY DURING SURGERY. BY MOVING THE LEG SEVERAL TIMES (EXTENSION, FLEXION, ROTATION), THE CONNECTOR REACHED THE DESIRED POSITION IN THE INTRACONDYLAR REGION, BUT THE AXIS DID NOT DEPLOY. THE CONNECTOR HAS BEEN EXCHANGED AND THE COUPLING WAS SUCCESSFUL AT THE FIRST ATTEMPT. [TRANSLATED DESCRIPTION FROM FIELD SERVICE].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17641711 ↗MalfunctionApr 5, 2023Activation Failure; Activation Failure
CONNECTION PROBLEM.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16685919 ↗InjuryNov 10, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
AS NOTIFIED BY (B)(6). TO COMPLAINTS ON (B)(6) 2022, SURGERY ON (B)(6) 2022, A REVISION OF SL HINGE DFR DISASSOCIATED OUT OF FEMUR. THROUGH TEXT, INFORMATION AS FOLLOWS: PATIENT FELL AND TWISTED KNEE BADLY. NO FAULT OF IMPLANT. X-RAY (ATTACHED) SHOWS FEMUR WAS IN A GREAT DEAL OF HYPER EXTENSION. WAITING INFORMATION ON RETURN OF EXPLANTS. UPDATE (B)(6) 2022: NOTIFIED BY THE REPRESENTATIVE THAT THE EXPLANTED PARTS WERE DISCARDED BY THE HOSPITAL STAFF ALTHOUGH THEY WERE INTENDED TO BE RETURNED.
Patient problems: Fall
MAUDE report 15769860 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K151008. One submission often covers many models, so check the product and lot details in each record.
- Z-1517-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8030/12;
- Z-1516-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-2836/11;
- Z-1522-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/12;
- Z-1515-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-2835/12;
- Z-1523-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/16;
- Z-1520-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/15;
- Z-1511-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/05;
- Z-1519-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/11;
- Z-1514-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/07;
- Z-1512-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/15;
- Z-1521-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-0027/11;
- Z-1518-2026 · Jan 12, 2026 · Open, ClassifiedEndo-Model Replacement Plateau; Item Number: 15-8521/09;
- Z-0890-2026 · Nov 10, 2025 · Open, ClassifiedEndo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.
- Z-2410-2024 · May 22, 2024 · Open, ClassifiedEndo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02, 15-2814/03, 15-2814/04, 15-2815/11, 15-2815/12, 15-2816/11, 15-2816/12, 15-281
- Z-2411-2024 · May 22, 2024 · Open, ClassifiedEndo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75m
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented devices
- LFIT 6260-9-136Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented166 reports
- Trident 623-00-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented112 reports
- Accolade 6021-0435Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented88 reports
- EXETER 0580-1-044Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall22 reports
- Endo-Model SL Knee Prosthesis System 16-2840/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall12 reports
- Trident 623-10-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented57 reports
- LFIT 6260-9-336Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented56 reports
- Avenir® 01.06010.001Zimmer GmbH · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page