Skip to content

Endo-Model Standard Knee Prosthesis System 15-0027/16Recalls & reported problems

WALDEMAR LINK GmbH & Co. KG · Tibial Plateau for Rotational Knee Prosthesis,Replacement Set · Class 2 · Implant

All models of Endo-Model Standard Knee Prosthesis System (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04026575432134), received between Jul 25, 2023 and Aug 4, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Expiration Date Error” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 6, 2026Class IIOpen, Classified

    The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.

    Z-2934-2026 · Root cause: Process change control · Recall details · FDA record ↗

  2. Jan 12, 2026Class IIOpen, Classified

    The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

    Z-1523-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Expiration Date Error1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 4, 2023Expiration Date Error; Expiration Date Error

    DISTRIBUTOR WAS NOTIFIED BY SALES REP THAT A POLY PLATEAU WITH AN EXPIRED DATE OF 03/2023 WAS IMPLANTED ON (B)(6) 2023. THE IMPLANT WAS REMOVED PRIOR TO CLOSING AND REPLACED. [CUSTOMER]

    Patient problems: Insufficient Information

    MAUDE report 17457771 ↗
  • InjuryJul 25, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    LINKBIO WAS NOTIFIED BY THE REPRESENTATIVE FROM EARHARDT MEDICAL THAT A POLY PLATEAU WITH AN EXPIRED DATE OF 03/2023 WAS IMPLANTED ON (B)(6), 2023. THE IMPLANT WAS REMOVED PRIOR TO CLOSING AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390847 ↗

Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page