Skip to content

Endo-Model Standard Knee Prosthesis System recallZ-2934-2026 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Jul 6, 2026

Class II Open, Classified All recalls posted in August 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.

FDA root cause category: Process change control

What the company is doing

Foreign manufacturer, Waldemar Link GmbH & Co KG, notified importer LinkBio Corp. on 7/6/2026 via email. LinkBio then notified the single US consignee by email/Docusign on 7/7/2026. Customer was instructed to return the affected product to LinkBio.

Recall details

Recall number
Z-2934-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jul 6, 2026
Posted by the FDA
Aug 11, 2026
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Quantity
1 US; 9 OUS
Product code
KRO
FDA submissions
K143179
Distribution
Domestic US distribution to Indiana. OUS distribution to Belarus and Germany.

Product description

Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16

Code information (models, lots, serial numbers)

UDI-DI 04026575432134 Lot for Implant REF 15-0027/16: 2604069

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates