RUSCH IPN040897Recalls & reported problems
Teleflex Incorporated · Endotracheal tube for laser surgery · Class 2
All models of RUSCH (220)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04026704002641), received between Mar 30, 2021 and Mar 30, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
7 other recalls cover devices cleared under the same FDA submission (K935842); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak2
Patient problems
- Fluid Discharge2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "OUTER CUFF SUDDENLY LEAKING, WAS FILLED WITH WATER AS RECOMMENDED. SLOW LEAKAGE FROM THE CUFF PERIOPERATIVELY SOME TIME INTO THE OPERATION. CLINICAL CONSEQUENCES: THE OPERATION WAS UNNECESSARILY PROLONGED". NO PATIENT HARM OR INJURY REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: Fluid Discharge
MAUDE report 11592157 ↗MalfunctionMar 30, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "PATIENT: MALE. OUTER CUFF BEGINS LEAKING PERIOPERATIVELY. IT WAS FILLED WITH WATER AS PRESCRIBED. THE CUFF BEGINS LEAKING IN THE MIDDLE OF THE OPERATION. NO EXPLANATION WHEN THE POSITION OF THE TUBE WAS INSPECTED USING VIDEO OPTICS. IT WAS NOT POSSIBLE TO GET THIS SEALED AND IT WAS CHANGED TO A STANDARD TUBE AND THE OPERATION WAS COMPLETED WITHOUT COMPLICATION. CLINICAL CONSEQUENCES: NONE, AS THE EVENT WAS AVOIDED". NO PATIENT HARM OR INJURY REPORTED. PATIENT CONDITION REPORTED AS "FINE".
Patient problems: Fluid Discharge
MAUDE report 11592099 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K935842. One submission often covers many models, so check the product and lot details in each record.
- Z-2533-2020 · May 29, 2020 · TerminatedRUSCH LASERTUBE (Rubber), RES 102004, Sizes: a) I.D. mm 4,0, Product Code 102004-000040 b) I.D. mm 5,0, Product Code 102004-000050 c) I.D. mm 6,0, Product Code 102004-000060 d) I.D
- Z-1686-2019 · Apr 1, 2019 · TerminatedRUSCH LASERTUBE (Rubber), Order No. : 102004-000080, REF 102004, SIZE I.D. mm 8,0
- Z-1684-2019 · Apr 1, 2019 · TerminatedRUSCH LASERTUBE (Rubber), Order No. : 102004-000060, REF 102004, SIZE I.D. mm 6,0
- Z-1685-2019 · Apr 1, 2019 · TerminatedRUSCH LASERTUBE (Rubber), Order No. : 102004-000070, REF 102004, SIZE I.D. mm 7,0
- Z-1683-2019 · Apr 1, 2019 · TerminatedRUSCH LASERTUBE (Rubber), Order No. : 102004-000050, REF 102004, SIZE I.D. mm 5,0
- Z-1682-2019 · Apr 1, 2019 · TerminatedRUSCH LASERTUBE (Rubber), Order No. : 102004-000040, REF 102004, SIZE I.D. mm 4,0
- Z-1851-2014 · May 23, 2014 · TerminatedRUSCH, LaserTube (Rubber), Laser Resistant Tracheal Tube, cuffed, Teleflex Medical, Willy-Rusch, GmbH. A tracheal tube is a device inserted into a patient's trachea via the n
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page