HUDSON RCI IPN044722Recalls & reported problems
Teleflex Incorporated · ET TUBE,CF,3.5 · Class 2
All models of HUDSON RCI (152)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04026704396733), received between Jan 5, 2024 and Sep 23, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (5 reports).
27 other recalls cover devices cleared under the same FDA submission (K822082); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Inflation Problem5
- Gas/Air Leak1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870974 ↗MalfunctionSep 23, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870723 ↗MalfunctionApr 16, 2026Inflation Problem
IT WAS REPORTED THAT "THE CUFF DIDN'T INFLATE DURING THE TEST BEFORE USE. THEREFORE, A NEW UNIT (LOT#: UNKNOWN) WAS USED INSTEAD". NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24904951 ↗MalfunctionAug 22, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED THAT: "THERE WERE TWO TUBES THAT FAILED PRE-OP INSPECTION. THEY WERE INFLATED BUT FAILED TO HOLD AIR. THE TUBES WERE BRAND NEW RIGHT OUT OF THE PACKAGE. THE ISSUE WAS IDENTIFIED PRIOR TO INTUBATION."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22870672 ↗MalfunctionJan 19, 2024Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE CUFF DIDN'T INFLATE DURING THE INFLATION TEST BEFORE USE. THEREFORE, A NEW KIT WAS USED INSTEAD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18542595 ↗MalfunctionJan 5, 2024Material Puncture/Hole; Gas/Air Leak; Material Puncture/Hole; Gas/Air Leak
IT WAS REPORTED THAT: THIS WAS AN INCIDENCE IN CARDIOSURGERY, INTUBATION OF SMALL VERY YOUNG PATIENT. NORMAL CONDITION. NURSE INFLATED THE CUFF, VISUAL INSPECTION, LUBRICATED THE TUBE. BUT ANESTHIOLOGIST SAW THE PATIENT DIDN'T REACH NORMAL VENTILATION VALUES. HIGHER PRESSURES AND SAW PATIENT WAS NOT VENTILATED ENOUGH. THEY HEARD AIR WAS LEAKING. SO THEY EXTUBATED THE PATIENT AND TRIED AGAIN WITH AN OTHER TUBE SUCCESSFULLY. AFTER ANOTHER VISUAL INSPECTION OF THE FIRST TUBE SHOWED THERE WAS A HOLE IN THE TUBE CLOSE TO THE CUFF. PATIENT WAS REPORTED AS FINE POST THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18451984 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K822082. One submission often covers many models, so check the product and lot details in each record.
- Z-2318-2019 · Jul 30, 2019 · TerminatedSheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management
- Z-2317-2019 · Jul 30, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
- Z-2316-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm,
- Z-2315-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF 5-10407 e) 4.0 mm, REF 5-10408 f) 4.5 mm, REF 5-10409 g)
- Z-2313-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-22515 e) 8.0 mm, REF 5-22516 f) 8.5 mm, REF 5-22517 g) 9.0
- Z-2312-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-10208 d) 4.5 mm, REF 5-10109 e) 5.0 mm, REF 5-10110, REF 5-10
- Z-2314-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF 5-22008, REF 5-22108, REF 5-22208 d) 4.5 mm, REF 5-22009,
- Z-2311-2019 · Jul 30, 2019 · TerminatedHudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
- Z-1956-2019 · Jun 7, 2019 · TerminatedSheridan Endotracheal Tube contained inside Centurion kit code TC7855.
- Z-1763-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/
- Z-1764-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm, Product Code 5-10314 4) HVT 7.5 mm, Product Code 5-10315 5) HVT 8.0
- Z-1761-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-2
- Z-1760-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3) CF 7.0 mm, Product Codes: 5-10114, 5-10214 4) CF 7.5 mm,
- Z-1765-2019 · May 24, 2019 · TerminatedSheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
- Z-1759-2019 · May 24, 2019 · TerminatedHudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
…and 12 more.
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page