Skip to content

RUSCH IPN044685Recalls & reported problems

Teleflex Incorporated · FLEXI-SET CUFFED STERILE · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04026704547616), received between May 18, 2021 and Jun 30, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

35 other recalls cover devices cleared under the same FDA submission (K961837); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 25, 2023Open, Classified

    Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.

    Z-1849-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Gas/Air Leak1
  • Leak/Splash1
  • Material Twisted/Bent1
  • Obstruction of Flow1

Patient problems

  • Airway Obstruction1
  • Insufficient Information1
  • Low Oxygen Saturation1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2026Gas/Air Leak

    IT WAS REPORTED THAT "THE CUFF AND THE PILOT BALLOON SEEM TO BE TRANSFERRING THE AIR BETWEEN EACH OTHER. THE PILOT BALLOON DOESN'T RETAIN ANY AIR, AND ETT CUFF STARTED TO DEFLATE WHILE IN PATIENT. CUSTOMER WENT A DIFFERENT DIRECTION TO COMPLETE THE PROCEDURE." NO PATIENT INJURY OR CONSEQUENCE REPORTED, NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S STATUS IS REPORTED AS "FINE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25669503 ↗
  • MalfunctionMar 21, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT: "END USER INDICATED KINKING IN ETT. PHYSICIAN INDICATED THERE ARE NO PATIENT HEALTH ISSUES CONNECTED TO THE COMPLAINTS. PRODUCT WAS PRE-INFLATED FOR TESTING. RE-INTUBATION AND ANY DESATURATION OF 02 ARE NOTED IN CASE EVENTS."

    Patient problems: Low Oxygen Saturation

    MAUDE report 21661421 ↗
  • MalfunctionJan 30, 2023Leak/Splash; Leak/Splash

    REPORTED EVENT: ENDOTRACHEAL TUBE CUFF LEAKED AND REQUIRED TUBE TO BE REPLACED. NO REPORTED INJURY. MEDWATCH REPORT#: (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 16265435 ↗
  • MalfunctionMay 18, 2021Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED "[ON] (B)(6) 2021 5:19PM PT INTUBATED IN THE OR WITH 7.0 TELEFLEX ETT. IN ICU - (B)(6) AM PT HAD PEAK VENTILATOR PRESSURES OF 10 - 11 CM H20 THEN THEY BEGAN READING 34-38 CM H2O, PATIENT WAS AGITATED AND HAD DIFFICULTY PASSING A SUCTION CATHETER. (B)(6) 12:29PM REMOVED THE ETT AND REINTUBATED, VENTILATOR PEAK PRESSURES RETURNED TO BASELINE OF 10-11 CM H20". PATIENT CONDITION REPORTED AS "FINE".

    Patient problems: Airway Obstruction

    MAUDE report 11843248 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K961837. One submission often covers many models, so check the product and lot details in each record.

  • Z-1846-2023 · May 25, 2023 · Open, Classified
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504555
  • Z-1853-2023 · May 25, 2023 · Open, Classified
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504590
  • Z-1883-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080100
  • Z-1891-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082085
  • Z-1889-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082075
  • Z-1851-2023 · May 25, 2023 · Open, Classified
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504580
  • Z-1962-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080075
  • Z-1899-2023 · May 25, 2023 · Open, Classified
    RUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Magill, REF 112480045
  • Z-1886-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082060
  • Z-1881-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080055
  • Z-1960-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080065
  • Z-1885-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082055
  • Z-1963-2023 · May 25, 2023 · Open, Classified
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080080
  • Z-1847-2023 · May 25, 2023 · Open, Classified
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504560
  • Z-1852-2023 · May 25, 2023 · Open, Classified
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504585

…and 20 more.

Other Tube, tracheal (w/wo connector) devices

All Tube, tracheal (w/wo connector) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.