RUSCH IPN044686Recalls & reported problems
Teleflex Incorporated · FLEXI-SET CUFFED STERILE · Class 2
All models of RUSCH (220)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04026704547623), received between Jan 4, 2023 and Nov 20, 2025 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (10 reports).
1 FDA recall record name this device's identifier, 1 still open.
35 other recalls cover devices cleared under the same FDA submission (K961837); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 25, 2023Open, Classified
Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.
Z-1850-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component10
- Material Twisted/Bent2
- Connection Problem1
- Leak/Splash1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Airway Obstruction2
- Insufficient Information2
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT: "HOSPITAL STATED PRODUCT WITH SEVER KINKING IN OROPHARYNX RESULTING IN SEVERE ELEVATION IN AIRWAY PRESSURE AND HEMODYNAMIC CHANGES IN THE PATIENT. THE ETT HAD TO BE EXCHANGED IN THE MIDDLE OF AN ONGOING SURGICAL PROCEDURE, SINCE THE KIND WAS NOT REVERSIBLE".
Patient problems: Airway Obstruction
MAUDE report 23602534 ↗MalfunctionOct 9, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT: "PATIENT WAS INTUBATED WITH A 7.5 ETT RUSCH (504575), THEN FLIPPED PRONE. MIDWAY THROUGH THE CASE, ANESTHESIA EXPERIENCED DIFFICULTIES WITH THE PATIENT'S AIRWAY FROM MONITORING DEVICES. THE DECISION WAS MADE TO FLIP THE PATIENT BACK TO SUPINE AND STILL EXPERIENCED DIFFICULTY WITH PATIENTS' AIRWAY. DECISION WAS MADE TO RE-INTUBATED PATIENT WITH A REINFORCED ETT. THE TUBE THAT HAD PREVIOUSLY PLACED WAS REMOVED: A VISIBLE KINK IN THE TUBE WAS DETERMINED TO HAVE OBSTRUCTED THE FLOW OF GASES THROUGH THE CIRCUIT. UNFORTUNATELY, WE DO NOT HAVE THE TUBE THAT WAS INITIALLY USED BUT DO HAVE LOT#. A GOOSE NECK ADAPTER HAD BEEN ATTACHED BETWEEN THE ETT AND THE ELBOW OF THE CIRCUIT. THERE WAS NO PATIENT INJURY OR CONSEQUENCE."
Patient problems: Airway Obstruction
MAUDE report 23252544 ↗MalfunctionJul 16, 2025Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
IT WAS REPORTED THAT "AFTER INSERTING IN THE PATIENT, THEY TRIED TO INFLATE AND THE TUBE FAILED. IT APPEARED TO HAVE A LEAK. THERE WAS A HOLE IN IT. PATIENT HAD TO BE EXTUBATED AND RE INTUBATED."
Patient problems: Insufficient Information
MAUDE report 22518093 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00061, 8040412-2024-00062, 8040412-2024-00063, 8040412-20 24-00065, 8040412-2024-00066, 8040412-2024-00067, 8040412-2024-00068 AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18711704 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709815 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00061, 8040412-2024-00062, 8040412-2024-00063, 8040412-2024-00064, 8040412-20 24-00065, 8040412-2024-00066, 8040412-2024-00067, AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709792 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00061, 8040412-2024-00062, 8040412-2024-00063, 8040412-2024-00064, 8040412-2024-00065, 8040412-2024-00066, 8040412-2024-00068 AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709776 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00061, 8040412-2024-00062, 8040412-2024-00063, 8040412-2024-00064, 8040412-20 24-00065, 8040412-2024-00067, 8040412-2024-00068 AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709767 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00061, 8040412-2024-00062, 8040412-2024-00063, 8040412-2024-00064, 8040412-2024-00066, 8040412-2024-00067, 8040412-2024-00068 AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709418 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00061, 8040412-2024-00062, 8040412-2024-00063, 8040412-2024-00064, 8040412-20 24-00065, 8040412-2024-00066, 8040412-2024-00067, 8040412-2024-00068 AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709404 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. THE ISSUE WAS IDENTIFIED DURING USE ON PATIENT BUT THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00061 , 8040412-2024-00063, 8040412-2024-00064, 8040412-2024-00065, 8040412-2024-00066, 8040412-2024-00067, 8040412-2024-00068 AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709387 ↗MalfunctionFeb 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT: THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. THIS DOES NOT HAPPEN IMMEDIATELY BUT NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST. THE END USER THEN REMOVES THE ORIGINAL CONNECTOR AND REPLACES WITH A 1/2 SIZE LARGER ETT NEW CONNECTOR WHICH SEEMS TO ALLEVIATE THE CONNECTION SEPARATION. ASSOCIATED COMPLAINTS 8040412-2024-00026, 8040412-2024-00062, 8040412-2024-00063, 8040412-2024-00064, 8040412-20 24-00065, 8040412-2024-00066, 8040412-2024-00067, 8040412-2024-00068 AND 8040412-2024-00069.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709341 ↗MalfunctionFeb 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED "THE END CONNECTOR ON THE ETT BECOMES SEPARATED FROM THE ETT TUBE ITSELF WHEN ATTACHED TO A VENT CIRCUIT. NOTICED IT OCCURS WHEN THE TUBE IS WARM AND MOIST." THE ORIGINAL CONNECTOR WAS REMOVED AND REPLACED WITH A LARGER CONNECTOR. NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18658436 ↗MalfunctionJan 4, 2023Connection Problem; Connection Problem
REPORTED ISSUE: PILOT BALLOON FOUND TO BE OUT OF ETT. IT IS NOT CLEAR IF IT WAS PULLED OR BIT OFF. NO ALARMS ON THE VENTILATOR WERE ACTIVATED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16099108 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K961837. One submission often covers many models, so check the product and lot details in each record.
- Z-1890-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082080
- Z-1961-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080070
- Z-1899-2023 · May 25, 2023 · Open, ClassifiedRUSCHELIT Safety Clear Tracheal Tube, oral/nasal, Magill, REF 112480045
- Z-1881-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080055
- Z-1886-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082060
- Z-1965-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080090
- Z-1892-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082090
- Z-1891-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082085
- Z-1894-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082100
- Z-1845-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504550
- Z-1853-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504590
- Z-1885-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082055
- Z-1888-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082070
- Z-1847-2023 · May 25, 2023 · Open, ClassifiedFlexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504560
- Z-1893-2023 · May 25, 2023 · Open, ClassifiedEndotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082095
…and 20 more.
Other Tube, tracheal (w/wo connector) devices
- Shiley 18875Covidien LP · Tube, tracheal (w/wo connector)4 deaths570 reports
- RUSCH IPN041476Teleflex Incorporated · Tube, tracheal (w/wo connector)10 recalls0 reports
- Shiley 18775Covidien LP · Tube, tracheal (w/wo connector)489 reports
- Shiley 18770Covidien LP · Tube, tracheal (w/wo connector)2 deaths434 reports
- Shiley 18880Covidien LP · Tube, tracheal (w/wo connector)2 deaths370 reports
- Shiley 18780Covidien LP · Tube, tracheal (w/wo connector)3 deaths286 reports
- Shiley 18770SCovidien LP · Tube, tracheal (w/wo connector)273 reports
- Shiley 18780SCovidien LP · Tube, tracheal (w/wo connector)251 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page