Heater-Cooler System 3T 208V ~ / 60HzRecalls & reported problems

LivaNova Deutschland GmbH · Class 2

All models of Heater-Cooler System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This LivaNova Heater-Cooler System 3T is a temperature controller used during cardiopulmonary bypass, which warms or cools the patient's blood as it circulates. The 33 reports mostly describe microbial contamination of the device, pumping problems, and fluid or blood leaks, along with incorrect display messages and temperature problems. No deaths and no injuries are mentioned in the reports. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 04033817900511), received between Mar 7, 2017 and Sep 21, 2026 — classified as 34 malfunctions.

The most frequently reported specific device problem is “Microbial Contamination of Device” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Microbial Contamination of Device11
  • Pumping Problem7
  • Fluid/Blood Leak4
  • Device Displays Incorrect Message3
  • Leak/Splash3
  • Temperature Problem3
  • Appropriate Device Problem Term/Code Not Available2
  • Insufficient Cooling2

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • No Patient Involvement5
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Pumping Problem

    LivaNova Deutschland received a report of a 3T Heater/Cooler for which the patient pump was found to not work properly. The issue was highlighted during a service intervention. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50835487 ↗
  • MalfunctionJul 9, 2026Pumping Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25762677 ↗
  • MalfunctionNov 21, 2025Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT THE HEATER COOLER 3T SYSTEM HAS TESTED POSITIVE FOR MNT CHIMERA. THERE IS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23618877 ↗
  • MalfunctionOct 31, 2025Temperature Problem; Temperature Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23433997 ↗
  • MalfunctionJul 29, 2025Unexpected Shutdown; Unexpected Shutdown

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22646245 ↗
  • MalfunctionJul 11, 2025Pumping Problem; Pumping Problem

    LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT OF A NOT FUNCTIONING PATIENT PUMP 1 IN A 3T HEATER COOLER. THE CUSTOMER INSTALLED A NEW PATIENT PUMP AND THE NEW PUMP IS MAKING NOISE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22477795 ↗
  • MalfunctionJul 10, 2024Temperature Problem; Temperature Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19708379 ↗
  • MalfunctionOct 12, 2023Fluid/Blood Leak; Leak/Splash; Pumping Problem; Fluid/Blood Leak; Leak/Splash; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17921219 ↗
  • MalfunctionJul 11, 2023Fluid/Blood Leak; Leak/Splash; Fluid/Blood Leak; Leak/Splash

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17297919 ↗
  • MalfunctionJun 5, 2023Pumping Problem; Pumping Problem

    SEE INITIAL PART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17058351 ↗
  • MalfunctionApr 17, 2023Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16750545 ↗
  • MalfunctionApr 11, 2023Pumping Problem; Temperature Problem; Pumping Problem; Temperature Problem

    LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT THAT A 3T HEATER COOLER PRESENTED DISPLAYED AN ERROR RELATED TO BLOCKED OR TOO SLOW PUMP (EE7) ON PATIENT SIDE DURING A PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16711473 ↗
  • MalfunctionApr 3, 2023Fluid/Blood Leak; Leak/Splash; Fluid/Blood Leak; Leak/Splash

    LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT OF WATER LEAKING FROM A 3T HEATER COOLER DURING A PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16663445 ↗
  • MalfunctionMar 29, 2023Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16636423 ↗
  • MalfunctionOct 7, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DEVICE WAS FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHIMAERA. REPORTEDLY, THE EQUIPMENT WAS NEVER PLACED IN THE OR, WEEKLY CHANGE OF WATER WAS DONE AND DISINFECTION WITH MINNCARE WAS DONE EVERY 2 WEEKS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15560537 ↗
  • MalfunctionJul 21, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA RECEIVED REPORT OF HEATER COOLER 3T SYSTEM FOUND CONTAMINATED WITH NTM. THE TYPE OF BACTERIUM WAS NO SPECIFIED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15072577 ↗
  • MalfunctionJul 21, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA RECEIVED REPORT OF HEATER COOLER 3T SYSTEM FOUND CONTAMINATED WITH NTM. THE TYPE OF BACTERIUM WAS NO SPECIFIED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15072567 ↗
  • MalfunctionJul 21, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA RECEIVED REPORT OF HEATER COOLER 3T SYSTEM FOUND CONTAMINATED WITH NTM. THE TYPE OF BACTERIUM WAS NO SPECIFIED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15072558 ↗
  • MalfunctionMay 24, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DEVICE WAS FOUND TO BE CONTAMINATED. THE TYPE OF BACTERIUM WAS NOT SPECIFIED BY THE CUSTOMER. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14486013 ↗
  • MalfunctionApr 28, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA (B)(4) RECEIVED A REPORT THAT, THE LABORATORY OF (B)(6) 2020 SHOW CONTAMINATION WITH NTM CHIMERA. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14237433 ↗
  • MalfunctionApr 20, 2022Degraded; Degraded

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T HAD A DEFECTIVE COOLING COMPRESSOR. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14155364 ↗
  • MalfunctionApr 15, 2022Insufficient Cooling; Insufficient Cooling

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T HAD A DEFECTIVE COOLING COMPRESSOR. THE ISSUE WAS IDENTIFIED BY THE CUSTOMER DURING ROUTINE MAINTENANCE.THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14117005 ↗
  • MalfunctionApr 1, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T WAS FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHELONAE IN (B)(6) 2021. THERE IS NO KNOWN PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13974385 ↗
  • MalfunctionApr 1, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T WAS FOUND TO BE CONTAMINATED WITH MOLD LIKE FUNGUS (TMTC). THERE IS NO KNOWN PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13974362 ↗
  • MalfunctionApr 1, 2022Microbial Contamination of Device; Microbial Contamination of Device

    LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T WAS FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHELONAE IN (B)(6) 2020. THERE IS NO KNOWN PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13974314 ↗

Other Controller, temperature, cardiopulmonary bypass devices

All Controller, temperature, cardiopulmonary bypass devices →