Skip to content

CardioQuip Modular Cooler-Heater (MCH) MCH-1000(i)Recalls & reported problems

Cardioquip, LLC · The CardioQuip Modular Cooler-Heater (MCH) is indicated to supply temperature-controlled water to heat exchange devices (e.g.; CPB Heat Exchangers, Cardioplegia Heat Exchangers, and thermal regulating system blankets) to help control a patient's temperature during cardiopulmonary bypass procedures lasting not longer than six hours. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This CardioQuip Modular Cooler-Heater controls the water temperature used during heart bypass surgery. Its 334 reports mostly describe microbial contamination of the device. The reports mention no deaths and no injuries. 2 recalls are open, over possible bacterial contamination that could infect patients and a heating chamber that could melt or burn its parts.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
334
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 334 adverse event reports that list this exact device (DI 00860000846103), received between Jul 15, 2023 and Sep 21, 2026 — classified as 334 malfunctions.

The most frequently reported specific device problem is “Microbial Contamination of Device” (332 reports).

2 FDA recall records name this device's identifier, 2 still open.

7 other recalls cover devices cleared under the same FDA submission (K102147); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 20 reportsSep 2024: 15 reportsOct 2024: 15 reportsNov 2024: 10 reportsDec 2024: 21 reportsJan 2025: 11 reportsFeb 2025: 6 reportsMar 2025: 17 reportsApr 2025: 6 reportsMay 2025: 19 reportsJun 2025: 28 reportsJul 2025: 7 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 55 reportsDec 2025: 4 reportsJan 2026: 11 reportsFeb 2026: 7 reportsMar 2026: 22 reportsApr 2026: 6 reportsMay 2026: 20 reportsJun 2026: 8 reportsJul 2026: 5 reportsAug 2026: 0 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202655
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 24, 2023Class IIOpen, Classified

    There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.

    Z-0565-2024 · Root cause: Process control · Recall details · FDA record ↗

  2. Oct 19, 2023Class IIOpen, Classified

    There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.

    Z-0439-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Microbial Contamination of Device332
  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions333
  • Intraoperative Cardiac Valve Injury1

Recent adverse event reports

The 25 most recent of 334 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848531 ↗
  • MalfunctionSep 21, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848530 ↗
  • MalfunctionSep 21, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848529 ↗
  • MalfunctionSep 21, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848527 ↗
  • MalfunctionSep 21, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848526 ↗
  • MalfunctionSep 21, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848525 ↗
  • MalfunctionSep 17, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809888 ↗
  • MalfunctionSep 17, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809883 ↗
  • MalfunctionSep 8, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50675039 ↗
  • MalfunctionSep 8, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50675037 ↗
  • MalfunctionSep 8, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50674954 ↗
  • MalfunctionJul 16, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25841735 ↗
  • MalfunctionJul 16, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25841734 ↗
  • MalfunctionJul 7, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON (B)(6) 2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25733070 ↗
  • MalfunctionJul 7, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 06/12/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25733062 ↗
  • MalfunctionJul 6, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON (B)(6) 2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25725828 ↗
  • MalfunctionJun 29, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED HPC TESTING TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 06/09/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25660645 ↗
  • MalfunctionJun 29, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 06/08/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25652132 ↗
  • MalfunctionJun 29, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED HPC TESTING TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 06/09/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25652130 ↗
  • MalfunctionJun 29, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 06/08/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25652121 ↗
  • MalfunctionJun 29, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 06/08/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25652120 ↗
  • MalfunctionJun 19, 2026Contamination /Decontamination Problem

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 05/27/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25561824 ↗
  • MalfunctionJun 3, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON (B)(6) 2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25383794 ↗
  • MalfunctionJun 3, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) TECHNICIAN NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION DURING AN ON-SITE INSPECTION AND SUBMITTED PICTURES TO CQ FOR REVIEW. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY REVIEWED THE PICTURES AND RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON 05/14/2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25383793 ↗
  • MalfunctionMay 28, 2026Microbial Contamination of Device

    A CARDIOQUIP (CQ) CUSTOMER NOTIFIED CQ SERVICE THAT THE INTERNAL TUBING OF THIS DEVICE WAS SUSPECTED OF CONTAMINATION. THE CQ DIRECTOR OF OPERATIONS AND EPIDEMIOLOGY RECOMMENDED THAT THE CUSTOMER PERFORM A QUARTERLY CLEANING AND DISINFECTION PROCEDURE ACCORDING TO THE DEVICE INSTRUCTIONS FOR USE (IFU). HPC TESTING IS ALSO RECOMMENDED TO PROVIDE A QUANTITATIVE ASSESSMENT OF WATER QUALITY. NO PATIENT INVOLVEMENT WAS REPORTED. HPC TEST RESULTS OUTSIDE OF CQ SPECIFICATIONS FOR WATER QUALITY WERE RECEIVED ON (B)(6) 2026, INDICATING DEVICE CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25316423 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K102147. One submission often covers many models, so check the product and lot details in each record.

  • Z-0566-2024 · Oct 24, 2023 · Open, Classified
    CardioQuip Modular Cooler-Heater, Model: MCH-1000(m), Power: 115V-60Hz, 16A
  • Z-0438-2024 · Oct 19, 2023 · Open, Classified
    CardioQuip Modular Cooler-Heater, Model:MCH-1000(m), Power: 115V-60hz,16A
  • Z-0069-2022 · Jun 24, 2021 · Open, Classified
    CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.
  • Z-0036-2022 · May 12, 2021 · Open, Classified
    CardioQuip, Modular Cooler-Heater 1000(i), Standard Unit, Model: MCH-10RMS, CE 2797, UDI: (01) 00860000846127
  • Z-0037-2022 · May 12, 2021 · Open, Classified
    CardioQuip, Modular Cooler-Heater 1000(m),Compact Unit, Model:MCH-1000(m),CE 2797, UDI: (01)00860000846110
  • Z-0038-2022 · May 12, 2021 · Open, Classified
    CardioQuip, Modular Cooler-Heater 1000(m),Thermoelectric Cooling Lid, Model: MCH-11TEC,CE 2797, UDI: (01) 00860000846134
  • Z-0035-2022 · May 12, 2021 · Open, Classified
    CardioQuip, Modular Cooler-Heater, Model: MCH-1000(i), CE 2797, UDI: (01)00860000846103

Other Controller, Temperature, Cardiopulmonary Bypass devices

All Controller, Temperature, Cardiopulmonary Bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page