CP5 recallZ-0401-2018 · Sorin Group Deutschland GmbH · initiated Oct 30, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sorin Group Deutschland GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.

FDA root cause category: Device Design

What the company is doing

Consignees were notified of the recall via USPS trackable mail, beginning on 10/30/2017. The letter informed customers Instructions included to continue following the Instructions for use, including use of an arterial filter on the arterial line, and that a LivaNova representative will contact them to schedule an upgrade to their device.

Recall details

Recall number
Z-0401-2018
Status
Terminated
Initiated by the company
Oct 30, 2017
Recalling firm
Sorin Group Deutschland GmbH
Quantity
1,318 units
Product code
DWA
FDA submissions
K112225
Distribution
Distributed domestically to Distributed internationally to Algeria, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Jordan, Kuwait, Libya, Lithuania, Luxembourg, Malaysia, Morocco, Norway, Palestinian Territory, Panama, Peru, Poland, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, The Netherlands, Tunisia, United Kingdom, Vietnam

Product description

CP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control panel. It used with an Electrical Remote-Controlled (ERC) tubing clamp and a bubble sensor.

Code information (models, lots, serial numbers)

Model No. 60-00-60. Drive Unit: Ref #60-01-04, UDI 04033817900948. Emergency System: Ref #60-01-35, UDI 04033817900955. Pump Control Panel: Ref #60-02-60, UDI 04033817901006.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K112225 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates