Skip to content

Flow Measurement Sensor Sorin 3/8" x 3/32"Recalls & reported problems

LivaNova Deutschland GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04033817901532), received between Sep 29, 2017 and Nov 20, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Inaccurate Flow Rate” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K011838); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Flow Rate1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT, DURING PRIMING, WHEN THE BLOOD FLOW RATE IS 0, THE CP5 FLOW SENSOR DISPLAY SHOWS 200ML TO 300ML AND THE FLOW IS UNSTABLE. THERE WAS NO PATIENT INVOLVEMENT. THE DISPLAY RETURNED TO NORMAL WHEN IT WAS REPLACED IT WITH A BACKUP SENSOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20732989 ↗
  • MalfunctionSep 29, 2017Inaccurate Flow Rate; Inaccurate Flow Rate

    LIVANOVA (B)(4) RECEIVED A REPORT THAT THE SCP FLOW SENSOR 3/8" DISPLAYED A FLOW RATE DEVIATION OF -0.9L/MIN WHEN COMPARED TO THE ACTUAL FLOW RATE DURING A PROCEDURE. THE SYSTEM WAS REBOOTED SEVERAL TIMES AND THE SENSOR WAS DISCONNECTED AND RE-CONNECTED, BUT THE ISSUE PERSISTED. THE SENSOR WAS REPLACED TO CONTINUE THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6903652 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K011838. One submission often covers many models, so check the product and lot details in each record.

  • Z-2439-2020 · Aug 26, 2019 · Terminated
    Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993
  • Z-2440-2020 · Aug 26, 2019 · Terminated
    Control Panel REF - 60-02-15: LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI: 04033817900986
  • Z-0167-2013 · Sep 20, 2012 · Terminated
    SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device
  • Z-0166-2013 · Sep 20, 2012 · Terminated
    SCP Pump Control Panel, Product Code: 60-02-15, Serial Numbers: 60S09121, 60S09123-60S09129, 60S09136-60S09139. The SCP is a cardiopulmonary bypass speed control device indicated

Other Control, pump speed, cardiopulmonary bypass devices

All Control, pump speed, cardiopulmonary bypass devices →

Sources and updates