Skip to content

CP5 Pump Control Panel (UDI 04033817902782)Recalls & reported problems

LivaNova Deutschland GmbH · Class 2

All models of CP5 (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04033817902782), received between Apr 19, 2023 and Sep 16, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Infusion or Flow Problem” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K112225); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem2
  • Pumping Problem2
  • Pumping Stopped2
  • Complete Loss of Power1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Insufficient Device Problem Information1
  • Insufficient Flow or Under Infusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50783874 ↗
  • MalfunctionDec 24, 2025Pumping Problem; Pumping Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23886736 ↗
  • MalfunctionNov 6, 2024Pumping Stopped; Pumping Stopped

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20619380 ↗
  • MalfunctionDec 19, 2023Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18356288 ↗
  • MalfunctionDec 12, 2023Pumping Stopped; Pumping Stopped

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CENTRIFUGAL PUMP 5 (CP5) STOPPED DURING PROCEDURE WHEN TURNING DOWN FLOW. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18309144 ↗
  • MalfunctionDec 11, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18299180 ↗
  • MalfunctionJul 20, 2023Pumping Problem; Pumping Problem

    SEE INTIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17359202 ↗
  • MalfunctionJun 28, 2023Incorrect, Inadequate or Imprecise Result or Readings; Infusion or Flow Problem; Incorrect, Inadequate or Imprecise Result or Readings; Infusion or Flow Problem

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17216958 ↗
  • MalfunctionApr 19, 2023Infusion or Flow Problem; Infusion or Flow Problem

    LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A CENTRIFUGAL PUMP 5 (CP5) GAVE A MOTOR CONTROL FAILURE MESSAGE AT SETUP. TURNING OFF AND ON THE DEVICE, THE ERROR MESSAGE DISAPPEARED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16767217 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112225. One submission often covers many models, so check the product and lot details in each record.

  • Z-0787-2022 · Dec 17, 2021 · Terminated
    Sorin Centrifugal Pump System, CP5, Drive Unit (Catalogue Item #60-01-04, UDI: 04033817900948) used in an HLM S5/C5 System configuration. A cardiopulmonary bypass speed control dev
  • Z-0401-2018 · Oct 30, 2017 · Terminated
    CP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control panel. It used with an Electrical Remote-Controlled (ERC) tubing clamp and a bubble sens
  • Z-0857-2014 · Dec 4, 2013 · Terminated
    CP5 centrifugal pump system Product Usage: The Stockert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.
  • Z-1329-2012 · Feb 23, 2012 · Terminated
    Pump control panel. The S5 Perfusion System is used to perform, control and monitor extracorporeal blood circulation during an operation for up to six (6) hours.

Other Control, pump speed, cardiopulmonary bypass devices

All Control, pump speed, cardiopulmonary bypass devices →

Sources and updates