Skip to content

TRACOE vario REF 470-08Recalls & reported problems

Tracoe medical GmbH · TRACOE vario extra Tracheostomy Tube with Adjustable Neck Flange, Low Pressure Cuff and Subglottic Suction Line · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04035324013029), received between Jul 29, 2025 and Oct 30, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2025Material Separation; Material Separation

    1) WHAT WENT WRONG WITH THE DEVICE: AN OBTURATOR IS INSTALLED IN THE LUMEN OF THE TRACHEOSTOMY CANNULA, WHICH MAKES IT EASIER TO INSERT THE CANNULA IN THE TRACHE AND TO REDUCE THE CALIPER GAP BETWEEN THE OUTER DIAMETER OF THE CANNULA AND A GUIDE. IT WAS REPORTED THAT DURING PULLING OUT THE MANDRIN, THE TIP OF THE OBTURATOR BROKE OFF AND REMAINED STUCK IN THE CANNULA. THANKS TO THE QUICK ACTION OF THE STAFF, HARM TO THE PATIENT WAS AVERTED. 2) DESCRIPTION OF HEALTH EFFECTS: NO HEALTH EFFECT OF THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23423351 ↗
  • MalfunctionJul 29, 2025Material Separation; Material Separation

    1) WHAT WENT WRONG WITH THE DEVICE: AN OBTURATOR IS INSTALLED IN THE LUMEN OF THE TRACHEOSTOMY CANNULA, WHICH MAKES IT EASIER TO INSERT THE CANNULA IN THE TRACHE AND TO REDUCE THE CALIPER GAP BETWEEN THE OUTER DIAMETER OF THE CANNULA AND A GUIDE. THE OBTURATOR IS PULLED OUT AFTER THE CANNULA HAS BEEN INSERTED INTO THE TRACHEA. WHEN PULLING IT OUT, THE TIP OF THE OBTURATOR BROKE OFF AND GOT STUCK IN THE LUMEN OF THE CANNULA. THIS SEPARATION WAS DISCOVERED IMMEDIATELY AFTER INSERTION AND AFTER IT WAS DETECTED THAT VENTILATION OF THE PATIENT DETERIORATED. AN OTHER CANNULA WAS RE-INSERTED WITHOUT ANY PROBLEM AND VENTILATION CONTINUED. 2) DESCRIPTION OF HEALTH EFFECTS: NO HARM OR DESATURATION OF THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22642467 ↗

Other Tube Tracheostomy And Tube Cuff devices

All Tube Tracheostomy And Tube Cuff devices →