Skip to content

TRACOE vario REF 450-08-PRecalls & reported problems

Tracoe medical GmbH · TRACOE vario Tracheostomy Tube Spiral-Reinforced, with Adjustable Neck Flange and Low Pressure Cuff, with minimally traumatic insertion system · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04035324017027), received between Jan 27, 2026 and Jan 27, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement1
  • Use of Device Problem1

Patient problems

  • Airway Obstruction1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 27, 2026Use of Device Problem; Unintended Movement

    1) WHAT WENT WRONG WITH THE DEVICE: ON (B)(6) 2025 IT WAS REPORTED WHILE A PATIENT WAS BEING REPOSITIONED FROM A SUPINE TO A LATERAL POSITION FOR A PUNCTURE, THE TRACHEAL CANNULA BECAME TWISTED INSIDE THE SHIELD DESPITE THE LOCK BEING CORRECTLY ADJUSTED (LOCK SYMBOL CLOSED). AS A RESULT, THE TRACHEAL CANNULA WAS NO LONGER POSITIONED CORRECTLY IN THE TRACHEA AND INITIALLY LEAKED. THE PATIENT THEN HAD TO BE REINTUBATED AS AN EMERGENCY MEASURE BECAUSE VENTILATION WAS IMPOSSIBLE. A BRONCHOSCOPY SHOWED THAT THE CANNULA WAS LYING WITH THE OPENING AGAINST THE MUCOUS MEMBRANE OF THE TRACHEA AND WAS THEREFORE MISPLACED. 2) DESCRIPTION OF HEALTH EFFECTS: NO HARM WAS CAUSED TO THE PATIENT. SHORT-TERM UNDERSUPPLY OF BREATHING AIR.

    Patient problems: Airway Obstruction

    MAUDE report 24187582 ↗

Other Tube Tracheostomy And Tube Cuff devices

All Tube Tracheostomy And Tube Cuff devices →

Sources and updates