Skip to content

TRACOE vario REF 470-08-PRecalls & reported problems

Tracoe medical GmbH · TRACOE vario extra Tracheostomy Tube with Adjustable Neck Flange, Low Pressure Cuff and Subglottic Suction Line, with minimally traumatic insertion system · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04035324017089), received between Dec 19, 2025 and Dec 19, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of or Failure to Bond1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    1) WHAT WENT WRONG WITH THE DEVICE: AFTER THE TRACHEAL CANNULA WAS PLACED WITHOUT COMPLICATIONS USING THE TRACOE VARIO SET, THE GUIDE CATHETER AND WIRE WERE REMOVED ACCORDING TO STANDARD PROCEDURE. IN THE PROCESS, THE CONICAL "RUBBER STOPPER" APPARENTLY DETACHED FROM THE GUIDE CATHETER AND REMAINED STUCK IN THE TRACHEAL CANNULA, MAKING VENTILATION IMPOSSIBLE. IT WAS THEN NO LONGER POSSIBLE TO THREAD THE WIRE. THE CANNULA HAD TO BE REMOVED AND A NEW TRACOE VARIO CANNULA WAS INSERTED WITHOUT WIRE GUIDANCE, WHICH WAS SUCCESSFUL. THIS RESULTED IN AN EMERGENCY SITUATION. 2) DESCRIPTION OF HEALTH EFFECTS: NO HARM TO THE PATIENT. REPORTED. THE PATIENT SURVIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23848007 ↗

Other Tube Tracheostomy And Tube Cuff devices

All Tube Tracheostomy And Tube Cuff devices →

Sources and updates