Skip to content

TRACOE twist plus REF 316-08Recalls & reported problems

Tracoe medical GmbH · TRACOE twist plus Tracheostomy Tube, with Low Pressure Cuff and Subglottic Suction Line · Class 2

All models of TRACOE twist plus (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 04035324019083), received between Jun 4, 2025 and Jul 7, 2026 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Loss of or Failure to Bond” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 6 reportsApr 2026: 5 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of or Failure to Bond13
  • Detachment of Device or Device Component2
  • Gas/Air Leak1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Aspiration/Inhalation2
  • Low Oxygen Saturation2
  • Upper Respiratory Tract Irritation2

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 7, 2026Loss of or Failure to Bond

    IT WAS REPORTED SUBGLOTTIC PORT WAS DETACHED. THE ARTICLE NUMBER AND THE LOT NUMBER WERE NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25729453 ↗
  • MalfunctionMay 19, 2026Loss of or Failure to Bond

    IT WAS REPORTED SUBGLOTTIC PORT DETACHED, SUBGLOTTIC LOAD ASPIRATION.

    Patient problems: Upper Respiratory Tract Irritation

    MAUDE report 25212395 ↗
  • MalfunctionApr 23, 2026Loss of or Failure to Bond

    SUBGLOTTIC SUCTION LINE DETACHED SUBGLOTTIC LOAD ASPIRATION TUBE EXCHANGED.

    Patient problems: Upper Respiratory Tract Irritation

    MAUDE report 24964837 ↗
  • MalfunctionApr 17, 2026Loss of or Failure to Bond

    SUBGLOTTIC PORT FELL OFF. TUBE CHANGED. NO HARM TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24910041 ↗
  • MalfunctionApr 17, 2026Loss of or Failure to Bond

    SUBGLOTTIC PORT FOUND DISCONNECTED IN BED. TUBE CHANGED. NO HARM TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24908705 ↗
  • MalfunctionApr 10, 2026Loss of or Failure to Bond

    THE SUBGLOTTIC PORT OF THE TRACHEOSTOMY BECAME DETACHED FROM THE TUBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24843090 ↗
  • MalfunctionApr 10, 2026Loss of or Failure to Bond

    SUBGLOTTIC PORT FOUND DISCONNECTED. TUBE CHANGED ON (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24838616 ↗
  • MalfunctionMar 23, 2026Loss of or Failure to Bond

    SUBGLOTTIC PORT BECAME DETACHED FROM TRACHEOSTOMY TUBE; THE PATIENT IS VERY DEPENDENT ON SGP SUCTION AND VENTILATED OVERNIGHT. THIS REQUIRED A TRACHEOSTOMY CHANGE, THIS CAUSED SIGNIFICANT UPSET TO PATIENT WHO WAS VERY DISTRESS AND UPSET USED FOR 19 DAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24660286 ↗
  • MalfunctionMar 12, 2026Loss of or Failure to Bond

    SUB-GLOTTIC PORT BECAME DETACHED FROM THE BODY OF THE TUBE WHEN PT WAS BEING ROLLED. NO PRESSURE OR PULLING OF TUBE. THIS REQUIRED AN EMERGENCY TUBE CHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24577469 ↗
  • MalfunctionMar 11, 2026Loss of or Failure to Bond

    IT WAS REPORTED ON 2026-02-23 THAT SUBGLOTTIC SUCTION PORT DETACHED FROM THE TUBE. THE PATIENT REQUIRED URGENT TRACHEOSTOMY CHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24568372 ↗
  • MalfunctionMar 5, 2026Detachment of Device or Device Component; Material Split, Cut or Torn

    ON (B)(6) 2026 PILOT BALLOON DETACHMENT ON THE TRACHEOSTOMY WAS REPORTED. EMERGENCY TUBE CHANGE WAS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24518711 ↗
  • MalfunctionMar 5, 2026Loss of or Failure to Bond

    ON (B)(6) 2026 SUBGLOTTIC SUCTION PORT DETACHMENT WAS REPORTED. PATIENT REQUIRED EMERGENCY TUBE CHANGE. NO HEALTH EFFECT REPORTED.

    Patient problems: Aspiration/Inhalation; Low Oxygen Saturation

    MAUDE report 24518441 ↗
  • MalfunctionMar 5, 2026Loss of or Failure to Bond

    IT WAS REPORTED ON 2026-02-11 THAT SUBGLOTTIC PORT BECAME DETACHED FROM TRACHEOSTOMY TUBE. THE PATIENT IS VERY DEPENDANT ON SGP SUCTION AND VENTILATED OVERNIGHT, REQUIRED AN EMERGENCY TRACHEOSTOMY CHANGE. THE PATIENT TOOK A LONG TIME TO OPTIMISE POST CUFF DEFLATION DUE TO BURDEN OF SECRETIONS, CAUSING DISCOMFORT AND DISTRESS AND DROP IN SATURATIONS.

    Patient problems: Aspiration/Inhalation; Low Oxygen Saturation

    MAUDE report 24517308 ↗
  • MalfunctionFeb 6, 2026Loss of or Failure to Bond

    SUBGLOTTIC SUCTION LINE DETACHED FROM TRACHEOSTOMY CANNULA DURING TREATMENT. THE PATIENT WAS DEPENDENT ON SGP SUCTION AND VENTILATED OVERNIGHT. THE PATIENT REQUIRED AN TRACHEOSTOMY CHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24275840 ↗
  • MalfunctionJun 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    SUBGLOTTIC SUCTION LINE DETACHED FROM TRACHEOSTOMY CANNULA DURING TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22222757 ↗
  • MalfunctionJun 4, 2025Gas/Air Leak; Gas/Air Leak

    1) WHAT WENT WRONG: CUFF LEAK WAS OBSERVED DURING USE AFTER MORE THAN 1 MONTH OF USE, WHICH IS MORE THAN THE ALLOWED DURATION OF USE OF 29 DAYS. CUFF PRESSURE WAS TESTED WITH MANOMETER AND INFLATION WAS DONE WITH MONOMETER AND SYRINGE. THE PRESSURE DECREASES IMMEDIATELY. IT IS NOT REPORTED IF A PRIORI CUFF TESTING WAS DONE AND PASSED. 2) HEALTH EFFECT: NO HEALTH EFFECT TO THE PATIENT REPORTED. THE TRACHEOSTOMY TUBE WITH THE CUFF PROBLEM WAS CHANGED THE FOLLOWING DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22139949 ↗

Other Tube tracheostomy and tube cuff devices