TRACOE silcosoft REF 373-4.0-C48Recalls & reported problems
Tracoe medical GmbH · Tracheostomy Tube for Children · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04035324031405), received between Apr 16, 2021 and Apr 16, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2021Fluid/Blood Leak; Fluid/Blood Leak
UPON ASSESSMENT OF TRACHEOSTOMY CUFF, IT WAS NOTED THAT THE CUFF WAS DEFLATED ALTHOUGH IT WAS INFLATED WITH 2 MLS OF STERILE WATER A FEW HOURS PRIOR. THE DECISION WAS MADE TO CHANGE THE TRACHEOSTOMY TUBE. ON EXAMINATION AFTER, IT WAS NOTED THAT THE CUFF WAS BLOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11676901 ↗
Other Tube Tracheostomy And Tube Cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page