TRACOE twist REF 306-07Recalls & reported problems
Tracoe medical GmbH · Tracheostomy Tube, with Low Pressure Cuff and Subglottic Suction Line · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04035324036219), received between Nov 28, 2024 and Mar 4, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Decrease in Suction1
- Defective Device1
- Dent in Material1
- Gas/Air Leak1
- Lack of Effect1
- Material Separation1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2026Material Separation; Detachment of Device or Device Component
ON 2026-02-06 IT WAS REPORTED THAT THE 15MM CONNECTOR CAME OFF WITH THE CATHETER MOUNT ON VENTILATOR DEPENDENT PATIENT. THE INNER CANNULA GOT STUCK IN THE TRACHEOSTOMY TUBE AND COULD NOT BE EASILY REMOVED WITHOUT THE CONNECTOR. EMERGENCY TRACHEOSTOMY CHANGE WAS PERFORMED. NO HEALTH EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24511287 ↗MalfunctionJan 27, 2026Gas/Air Leak
ON 2026-01-09 AN INCREASED RISK OF ASPIRATION DUE TO A FAULTY TRACHEOSTOMY TUBE CUFF WAS REPORTED. THE FAULTY TUBE HAD BEEN INSERTED ON (B)(6) 2025 AND WAS FOUND TO BE FAULTY ON (B)(6) 2026. PRIOR TO INSERTION, THE CUFF WAS TESTED, WITH NO AIR LEAKAGE NOTED AT THAT TIME. ON (B)(6) 2026, STAFF FOUND CUFF TO BE DEFLATED. THE PATIENT EXHIBITS INVOLUNTARY MOVEMENTS, INCLUDING LIFTING BOTH HANDS TOWARD THE CHEST AND FLEXING THE NECK DOWNWARDS, WHICH OCCUR DAILY AND MAY HAVE EXERTED PRESSURE ON THE TRACHEOSTOMY SYSTEM. THE TRACHEOSTOMY TUBE WAS REMOVED AND REPLACED BY NURSING STAFF. NO HEALTH EFFECT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24187096 ↗MalfunctionNov 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ON 22-10-2025 IT WAS REPORTED THAT DURING MORNING ROUTINE CARE STAFF NOTICED THAT THE PORT WAS SLIGHTLY SPLIT. THE SUB-GLOTTIC PORT SPLIT COMPLETELY WHILE ASPIRATING. REMEDIAL ACTIONS WERE TAKEN BY HEALTHCARE FACILITY, PATIENT, OR USER SUBSEQUENT TO THE INCIDENT. THE PATIENT OF UNKNOWN AGE AND SEX REQUIRED UNPLANNED TRACHEOTOMY TUBE CHANGE. NO HEALTH EFFECT OF THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23449453 ↗MalfunctionJun 30, 2025Decrease in Suction; Dent in Material; Defective Device; Lack of Effect; Decrease in Suction; Dent in Material; Defective Device; Lack of Effect
1) WHAT WENT WRONG? WITHIN THE TIME FRAME OF (B)(6) 2025 TO (B)(6) 2025 IT WAS DETECTED BY THEMEDICAL STAFF THAT THE SUBGLOTTIC PORT HAS A KINK WHICH PREVENTS TO ASPIRE SECRETION.2) DESCRIPTION OF THE HEALTH EFFECTS: NO HEALTH EFFECT DESCRIBED. IT SEEMS THAT SUCTIONING SECRETION WAS NOT OR HARDLY POSSIBLE.
Patient problems: Insufficient Information
MAUDE report 22355912 ↗MalfunctionNov 28, 2024Mechanical Problem; Mechanical Problem
AFTER 5 DAYS OF USE THE PORT OF THE SUCTION LINE OF THE TRACOE TWIST TRACHEOSTOMY CANNULA SPLIT. UNABLE TO OBTAIN ASPIRATES SO UNPLANNED ELECTIVE CHANGE PERFORMED. NO PERCEIVED HARM TO PATIENT AS A RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20804039 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page