Skip to content

AESCULAP BH145RRecalls & reported problems

Aesculap AG · CRILE Hemostatic Forceps, curved, 5 1/2", (140 mm), delicate, blunt, non-sterile, reusable · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04038653025800), received between Nov 3, 2021 and Dec 8, 2023 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 31, 2025Open, Classified

    The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

    Z-1309-2025 · Root cause: Use error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Integrity Problem4
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 8, 2023Fracture; Fracture

    IT WAS REPORTED TO AESCULAP INC. THAT A CRILE FORCEPS CVD 140MM (PART# BH145R) WAS USED DURING A CESAREAN PROCEDURE ON (B)(6) 2023. ACCORDING TO THE COMPLAINANT THE DEVICE HAD BROKE DURING THE PROCEDURE. THE FRAGMENT WAS ABLE TO BE RETREIVED AND AN X-RAY WAS PERFORMED ON THE PATIENT. THE ADVERSE EVENT IS FILED UNDER AIC REFERENCE (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18291816 ↗
  • MalfunctionNov 3, 2021Material Integrity Problem; Material Integrity Problem

    THIS CASE (B)(4) IS A DOUBLETTE OF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12747466 ↗
  • MalfunctionNov 3, 2021Material Integrity Problem; Material Integrity Problem

    THIS CASE (B)(4) IS A DOUBLETTE OF (B)(4) AND WILL BE CLOSED AS A DOUBLETTE. THEREFORE IT IS NO REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12747465 ↗
  • MalfunctionNov 3, 2021Material Integrity Problem; Material Integrity Problem

    THIS CASE (B)(4) WILL BE CLOSED AS A DOUBLETTE OF (B)(4). THEREFORE IT IS NO LONGER REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12747460 ↗
  • MalfunctionNov 3, 2021Material Integrity Problem; Material Integrity Problem

    THIS CASE IS NOT ASSOCIATED WITH THE LISTED MEDWATCH REPORTS BELOW. ASSOCIATED MEDWATCH-REPORTS: 9610612-2021-00726 (400533509 BH145R) 9610612-2021-00727 (400533511 BH145R) 9610612-2021-00728 (400533513 BH145R) 9610612-2021-00687 (400530950 BH145R) IT IS POSSIBLY ASSOCIATED WITH 400529799.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12747459 ↗

Other HEMOSTAT devices

All HEMOSTAT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page