Numerous models of nonsterile hemostatic forceps recallZ-1309-2025 · Aesculap Inc · initiated Jan 31, 2025
Class II Open, Classified All recalls posted in March 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Aesculap Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
FDA root cause category: Use error
What the company is doing
The recalling firm issued letters dated 1/29/2025 on 1/31/2025 via certified mail return-receipt. The letter explained the reason for the correction, identified the product impacted, provided the risk to health, and the actions required for the customer/user. The actions included: (1) Review the Urgent Medical Device Correction Notification in its entirety and ensure all users within the consignee organization are informed about the correction. Post the notification where the affected products are stored. If the consignee is a distributor, they are to forward the notification to their customers. The correction is to be extended to the end user/customer level; (2) Determine the current inventory of the affected items and comply with the following instructions: (a) Use the products identified for clamping of tissue and small vessels only; (b) Inspect the products for end-of-life indicators prior to use; (c) Visually inspect the products which should be clean and free of any signs of corrosion and cracks. Check the joint and jaw in the open position from both sides (if needed, use magnification; (d) Test the function of the device, including the jaw area, for deformation and loss of tension. An Acknowledgment Form was enclosed for completion and return via FAX or email within 2 weeks of receipt, even if the total inventory in the consignee location is 0.
Recall details
- Recall number
- Z-1309-2025
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jan 31, 2025
- Posted by the FDA
- Mar 11, 2025
- Recalling firm
- Aesculap Inc
- Quantity
- 917,965 forceps
- Product code
- HRQ
- Distribution
- Distribution was made nationwide and to PR. There was also government/military distribution. Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.
Product description
Numerous models of nonsterile hemostatic forceps: (1) REF BH100R, Carrel Artery Forceps 85MM; (2) REF BH104R, Hartmann Mosquito FCSDEL STR 100MM; (3) REF BH105R, Hartmann Mosquito FCPSDELCVD100MM; (4) REF BH108R, Micro-Halsted Forceps DEL STR125MM; (5) REF BH109R, Micro-Halsted Forceps DEL CVM 125mm; (6) REF BH114R, Hartmann Mosquito FCPSDELSTR1X2100MM; (7) REF BH115R, Hartmann Mosquito FCPSDELCVD1X2100MM; (8) REF BH118R, Micro-Halsted FCPS Del STR 1X2125MM; (9) REF BH119R, Micro-Halsted FCPS Del CVD 1X2125MM; (10) REF BH120R, Halsted-Mosquito FCPSDELSTR1X2125MM; (11) REF BH121R, Halsted-Mosquito FCPS DELCVD1X2125MM; (12) REF BH130R, Providence-Hosp FCPS Del STR140MM; (13) REF BH131R, Providence-Hosp FCPS Del CVD 140MM; (14) REF BH134R, Kelly Forceps Del STR 140MM; (15) REF BH135R, Kelly Forceps Del CVM 140MM; (16) REF BH139R, Dandy Delicate Forceps CVD140MM; (17) REF BH140R, Baby Crile Forceps Del STR 140MM; (18) REF BH141R, Baby Crile FORCEPSDELCVD140MM; (19) REF BH143R, Dandy Del FCPS LAT-CVD 1X2 140MM; (20) REF BH144R, Crile Forceps STR 140MM; (21) REF BH145R, Crile Forceps CVD 140MM; (22) REF BH150R, Baby Kocher Forceps STR 140MM; (23) REF BH151R, Kocher Forceps Del CVD 1X2140MM; (24) REF BH154R, Crile Forceps Del STR 1X2 140MM; (25) REF BH155R, Crile Forceps Del CVD 1X2 140MM; (26) REF BH160R, Leriche Delicate Forceps STR 150MM; (27) REF BH161R, Leriche Delicate Forceps CVD 150MM; (28) REF BH164R, Rankin FCPS Long Kelly Del STR160MM; (29) REF BH166R, Crile Forceps STR 160MM; (30) REF BH170R, Leriche Delicate Forceps STR 1X2150MM; (31) REF BH171R, Leriche Delicate Forceps CVD1X2150MM; (32) REF BH176R, Crile Forceps STR 1X2160MM; (33) REF BH177R, Crile Forceps CVD 1X2 160MM; (34) REF BH178R, Dixon-Lovelace Artery FCPS.1X2T.STR.160; (35) REF BH197R, Birkett Delicate Forceps CVD185MM; (36) REF BH199R, Nissen Delicate Forceps CVP 185MM; (37) REF BH200R, Adson Delicate Forceps STR185MM; (38) REF BH201R, Adson Delicate Forceps CVD 185MM; (39) REF BH202R, Halsted Delicate Forceps STR 185MM; (40) REF BH203R, Halsted Delicate Forceps CVD 185MM; (41) REF BH205R, Adson Hemostatic FCPS CVD.215MM; (42) REF BH206R, Heiss Delicate Forceps STR 200MM; (43) REF BH207R, Heiss Delicate Forceps CVD 200MM; (44) REF BH210R, Halsted-Mosquito Forceps DELSTR200MM; (45) REF BH211R, Halsted-Mosquito FCPS Del CVD 200MM; (46) REF BH212R, Halsted-Mosquito FCPSDELSTR1X2185MM; (47) REF BH213R, Halsted-Mosquito FCPSDELCVD1X2185MM; (48) REF BH214R, Halsted-Mosq.FCPSDELSTR1X2 200MM; (49) REF BH215R, Halsted-Mosq.FCPSDELCVD 1X2 200MM; (50) REF BH224R, Sarot Delicate Forceps 240MM; (51) REF BH225R, Crafoord Delicate Forceps CVD 245MM; (52) REF BH227R, Artery Forceps Crafoord MODIF.240MM; (53) REF BH228R, Bengolea Delicate Forceps STR 245MM; (54) REF BH229R, Bengolea Delicate Forceps CVD 245MM; (55) REF BH230R, Bridge Delicate Forceps STR 275MM; (56) REF BH231R, Bridge Delicate Forceps CVD 275MM; (57) REF BH234R, Sarot Delicate Forceps STR 1X2240MM; (58) REF BH316R, Pean Artery Forceps STR.165MM; (59) REF BH332R, Spencer-Wells Forceps STR130MM; (60) REF BH333R, Spencer-Wells Forceps CVD 130MM; continued
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
GTIN numbers: (1) BH100R - 04038653139064; (2) BH104R - 04038653025558; (3)BH105R - 04038653025565; (4) BH108R - 04038653025572; (5) BH109R - 04038653025589; (6) BH114R - 04038653025619; (7) BH115R - 04038653025626; (8) BH118R - 04038653025633; (9) BH119R - 04038653025640; (10) BH120R - 04038653025657; (11) BH121R - 04038653025664; (12) BH130R - 04038653025725; (13) BH131R - 04038653025732; (14) BH134R - 04038653025749; (15) BH135R - 04038653025756; (16) BH139R - 04038653025763; (17) BH140R - 04038653025770; (18) BH141R - 04038653025787; (19) BH143R - 04038653139088; (20) BH144R - 04038653025794; (21) BH145R - 04038653025800; (22) BH150R - 04038653025817; (23) BH151R - 04038653025824; (24) BH154R - 04038653025831; (25) BH155R - 04038653025848; (26) BH160R - 04038653025855; (27) BH161R - 04038653025862; (28) BH164R - 04038653025879; (29) BH166R - 04038653025893; (30) BH170R - 04038653025923; (31) BH171R - 04038653025930; (32) BH176R - 04038653025947; (33) BH177R - 04038653025954; (34) BH178R - 04038653025961; (35) BH197R - 04038653026012; (36) BH199R - 04038653026036; (37) BH200R - 04038653026043; (38) BH201R - 04038653026050; (39) BH202R - 04038653026067; (40) BH203R - 04038653026074; (41) BH205R - 04038653411542; (42) BH206R - 04038653026081; (43) BH207R - 04038653026098; (44) BH210R - 04038653026111; (45) BH211R - 04038653026128; (46) BH212R - 04038653026135; (47) BH213R - 04038653026142; (48) BH214R - 04038653026159; (49) BH215R - 04038653026166; (50) BH224R - 04038653026203; (51) BH225R - 04038653026210; (52) BH227R - 04038653026227; (53) BH228R - 04038653026234; (54) BH229R - 04038653026241; (55) BH230R - 04038653026258; (56) BH231R - 04038653026265; (57) BH234R - 04038653026272; (58) BH316R - 04038653026357; (59) BH332R - 04038653026371; (60) BH333R - 04038653026388; (61) BH334R - 04038653026395; (62) BH335R - 04038653026401; (63) BH336R - 04038653026418; (64) BH337R - 04038653026425; (65) BH339R - 04038653026449; (66) BH410R - 04038653139101; (67) BH412R - 04038653026487; (68) BH413R - 04038653026494; (69) BH414R - 04038653026500; (70) BH415R - 04038653026517; (71) BH422R - 04038653026524; (72) BH424R - 04038653026531; (73) BH425R - 04038653026548; (74) BH442R - 04038653026555; (75) BH443R - 04038653026562; (76) BH444R - 04038653026579; (77) BH445R - 04038653026586; (78) BH446R - 04038653026593; (79) BH447R - 04038653026609; (80) BH448R - 04038653026616; (81) BH449R - 04038653026623; (82) BH450R - 04038653026630; (83) BH451R - 04038653026647; (84) BH456R - 04046963068783; (85) BH470R - 04038653026654; (86) BH471R - 04038653026661; (87) BH472R - 04038653026678; (88) BH473R - 04038653026685; (89) BH474R - 04038653026692; (90) BH612R - 04038653026722; (91) BH614R - 04038653026739; (92) BH615R - 04038653026746; (93) BH618R - 04038653026753; (94) BH619R - 04038653026760; (95) BH630R - 04038653026777; (96) BH631R - 04038653026784; (97) BH642R - 04038653026791; (98) BH643R - 04038653026807; (99) BH644R - 04038653026814; (100) BH645R - 04038653026821; (101) BH646R - 04038653026838; (102) BH647R - 04038653026845; (103) BH648R - 04038653026852; (104) BH649R - 04038653026869; (105) BH650R - 04038653026876; (106) BH651R - 04038653026883; (107) BH652R - 04038653026890; (108) BH653R - 04038653026906; (109) BH654R - 04038653026913; (110) BH655R - 04038653026920; (111) BH800R - 04038653139118; (112) BH801R - 04038653139125; (113) BH804R - 04038653139132; (114) BH805R - 04038653139149; (115) BH812R - 04038653026951; (116) BH831R - 04038653026982; (117) BH833R - 04038653026999; (118) MB229R - 04046963364458; and (119) MD469 - 04046963366148.
Devices named in this recall
119 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- AESCULAP BH145RAesculap AG · HEMOSTAT1 recall5 reports
- AESCULAP BH443RAesculap AG · HEMOSTAT1 recall2 reports
- AESCULAP BH135RAesculap AG · HEMOSTAT1 recall1 report
- AESCULAP BH104RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH105RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH108RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH109RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH114RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH115RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH118RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH119RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH120RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH121RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH130RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH131RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH134RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH139RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH140RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH141RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH144RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH150RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH151RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH154RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH155RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH160RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH161RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH164RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH166RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH170RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH171RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH176RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH177RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH178RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH197RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH199RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH200RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH201RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH202RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH203RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH206RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH207RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH210RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH211RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH212RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH213RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH214RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH215RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH224RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH225RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH227RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH228RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH229RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH230RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH231RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH234RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH316RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH332RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH333RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH334RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH335RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH336RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH337RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH339RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH412RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH413RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH414RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH415RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH422RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH424RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH425RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH442RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH444RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH445RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH446RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH447RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH448RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH449RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH450RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH451RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH470RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH471RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH472RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH473RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH474RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH612RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH614RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH615RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH618RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH619RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH630RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH631RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH642RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH643RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH644RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH645RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH646RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH647RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH648RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH649RAesculap AG · HEMOSTAT1 recall0 reports
- AESCULAP BH650RAesculap AG · HEMOSTAT1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96223).
- Z-1308-2025 · Open, ClassifiedNumerous models of nonsterile hemostatic forceps: (1) REF BJ500R, Phaneuf Hysterectomy Forceps STR 210MM; (2) REF BJ501R, Phaneuf Hysterectomy Forceps ANG 205MM; (3) REF BJ511R, H…
- Z-1310-2025 · Open, ClassifiedNumerous models of nonsterile hemostatic forceps: (1) REF BH198R, Fuchsig Artery Forceps CVD.190MM; (2) REF BH304R, Pean Artery Forceps Straight 140MM; (3) REF BH813R, Artery Forc…
- Z-1311-2025 · Open, ClassifiedNumerous models of nonsterile hemostatic forceps: (1) REF FB458R, Glover ATR.CLAMP CVD 215MM; (2) REF FB459R, Glover ATR.CLAMP CVD 245MM; (3) REF FB461R, Glover ATR.CLAMP 195MM; …
- Z-1312-2025 · Open, ClassifiedNumerous models of nonsterile hemostatic forceps: (1) REF BH951R, Negus Tonsil Forceps HVY-CVD 190MM; (2) REF BH952R, Negus Tonsil Forceps X-HVY-CVD190MM; (3) REF BH957R, Schnidt …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page