AESCULAP PO004RRecalls & reported problems
Aesculap AG · METZENBAUM ADTEC MONOPOLAR TC Dissecting Scissors, complete instrument, monopolar, curved to left, working length: 12 1/4", (310 mm), diam. 5 mm, serrated (fine), blunt/blunt, rotatable, insulated, double action, consisting of PM973R, PO603R, PO958R, detachable, non-sterile, reusable · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04038653260140), received between Feb 10, 2021 and Feb 10, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Arcing” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Arcing1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2021Arcing; Arcing
IT WAS REPORTED TO AESCULAP INC. THAT METZENBAUM SERRATED SCISSORS (PART # PO004R) WERE USED DURING A UROLOGY PROCEDURE PERFORMED (B)(6) 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SURGEON USED THE SCISSORS WITH A MONOPOLAR CONNECTION TO TAKE DOWN SOME ADHESIONS AND IT ARCED NEAR THE DISTAL END OF THE INSULATION. THE SYSTEM SETTINGS WERE NOTED AS BEING 30 CUT/30 COAG. THE INSULATION PASSED THE INSULATION TEST DURING REPROCESSING PRIOR TO THE CASE. NO INJURIES TO THE PATIENT, THE ARCING HIT THE ABDOMINAL WALL AND NO BOWEL WAS BURNED. THE COMPLAINT DEVICE WAS AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. NO PATIENT HARM WAS REPORTED AS A RESULT OF THIS EVENT. REPOREDLY, THE ARCING HIT THE ABDOMINAL WALL, BUT NO BOWEL BURN OCCURRED. ALTHOUGH REQUESTED, ADDITIONAL INFORMATION HAS NOT BEEN MADE AVAILABLE. THE MALFUNCTION IS FILED UNDER AAG REFERENCE XC (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11308920 ↗
Other COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES) devices
- Olympus WA47786AOLYMPUS Winter & Ibe GmbH · Coagulator-Cutter, Endoscopic, Unipolar (And Accessories)14 reports
- n.a. 37370DLKarl Storz GmbH & Co. KG · COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES)13 reports
- Sonata® Radiofrequency Ablation Handpiece RFA-002Gynesonics, Inc. · Coagulator-Cutter, Endoscopic, Unipolar (And Accessories)12 reports
- n.a. 26183MC-SKarl Storz GmbH & Co. KG · COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES)12 reports
- n.a. 26183MD-SKarl Storz GmbH & Co. KG · COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES)7 reports
- Integra® Jarit® 600305Integra Lifesciences Production Corporation · COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES)3 reports
- Sonata® Dispersive Electrode DE-001Gynesonics, Inc. · Coagulator-Cutter, Endoscopic, Unipolar (And Accessories)3 reports
- AESCULAP PO615RAesculap AG · COAGULATOR-CUTTER, ENDOSCOPIC, UNIPOLAR (AND ACCESSORIES)2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page